header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

A NEW APPROACH IN HIGH PRECISION BONE SURGERY: SEMI-AUTOMATIC ALIGNMENT OF AN APPLICATION-SPECIFIC INSTRUMENT GUIDE ADJUSTED BY A SMART SCREWDRIVER

International Society of Computer-Assisted Orthopaedic Surgery (CAOS), 17th Annual Scientific Meeting, Aachen, June 2017



Abstract

Robotic surgical systems reduce the cognitive workload of the surgeon by assisting in guidance and operational tasks. As a result, higher precision and a decreased surgery time are achieved, while human errors are minimised. However, most of robotic systems are expensive, bulky and limited to specific applications.

In this paper a novel semi-automatic robotic system is evaluated, that offers the high accuracies of robotic surgery while remaining small, universally applicable and easy to use. The system is composed of a universally applicable handheld device, called Smart Screwdriver (SSD) and an application specific kinematic chain serving as a tool guide. The guide mechanism is equipped with motion screws. By inserting the SSD into a screw head, the screw is identified automatically and the required number of revolutions is executed to achieve the desired pose of the tool guide.

The usability of the system was evaluated according to IEC 60601-1-6 using pedicle screw implementation as an example. The achieved positioning accuracies of the drill sleeve were comparable to those of fully automatic robotic systems with −0.54 ± 0.93 mm (max: − 2.08 mm) in medial/lateral-direction and 0.17 ± 0.51 mm (max: 1.39 mm) in cranial/caudal- direction in the pedicle isthmus. Additionally, the system is cost-effective, safe, easy to integrate in the surgical workflow and universally applicable to applications in which a static position in one or more DOF is to be adjusted.