header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

PROSPECTIVE EVALUATION OF SERUM COBALT, CHROMIUM, AND TITANIUM AFTER IMPLANTATION OF METAL-ON-METAL TOTAL HIP PROSTHESES



Abstract

Purpose: Serum cobalt and chromium levels after metal-on-metal implantations are not well known. There has been little data on the correlation with clinical and radiological surveillance. This prospective analysis followed the changes in serum cobalt, chromium, and titanium levels in order to ascertain the behaviour of the femoral implant and the bearing.

Material and methods: The series included 292 patients followed for 27–72 months who were free of kidney failure, were not given vitamin B12 supplementation, and did not have occupational exposure. The same cemented implants were used in all patients: titanium femoral stem, Metasuly heads (28 mm). Serum samples were drawn preoperatively and at three, six and twelve months then annually (systematic activity questionnaire). The limit for detection of serum cobalt and chromium was 1 nmol/l (0.05μg/l); for titanium it was 30 nmol/l (1.4m/l).

Results: The main problems encountered were two impingements and two femoral loosenings. Four patterns were identified in the time courses of serum cobalt and chromium. In decreasing order, they were: type 1, low initial level (< 50 nmol/l) then no change; type 2, high initial level (> 50 nmol/l) then decrease; type 3 low initial level then progression; type 4, high initial level then further elevation. Bilateral prostheses showed a particular pattern with elevation after the second implantation. The serum levels rapidly returned to normal after revision in the two impingement cases. The titanium levels were correlated with femoral problems which were not initially detected on the x-rays.

Discussion: Correlation between serum cobalt level and wear is difficult to establish. There was no specific pattern after dislocation. The kinetics showed certain patterns with possible prognostic significance: groups 1 (metallic silence) and 2 (breaking-in) would correspond to favourable evolution; type 3 would be difficult to interpret (missed impingement, foreign body, articular decoaptation or major change in activity level); type 4 is highly suggestive of premature wear or a biological problem (excepting bilateral implants). It is important to monitor serum cobalt and titanium simultaneously to detect an interface problem and or femoral loosening.

Correspondence should be addressed to SOFCOT, 56 rue Boissonade, 75014 Paris, France.