header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty



Download PDF

Abstract

Dual-energy X-ray absorptiometry (DEXA) is increasingly used to measure changes in bone mineral density (BMD) around femoral prostheses after total hip arthroplasty. We have studied the factors which affect the accuracy of these measurements. The coefficient of variation was < 2% using a hydroxyapatite phantom, 2.7% in an anthropomorphic phantom specimen, and < 1% in repeated measurements on implanted cadaver femora. The precision did not vary with different implant materials or designs. In patients we found a mean precision error of 2.7% to 3.4%. The most significant factor affecting reproducibility was rotation of the femur. We conclude that DEXA is a precise method of measurement for small changes in BMD around femoral implants, but that correct and careful positioning of patients is essential to obtain reliable results.

For access options please click here