Proximal translation of > 1 mm within the first two years of revision total hip arthroplasty correctly predicts whether or not an acetabular component is loose in 80% of cases

a case-control study with confirmed intra-operative outcomes



Download PDF

Abstract

Aims

The purpose of this study was to determine the sensitivity, specificity and predictive values of previously reported thresholds of proximal translation and sagittal rotation of cementless acetabular components used for revision total hip arthroplasty (THA) at various times during early follow-up.

Patients and Methods

Migration of cementless acetabular components was measured retrospectively in 84 patients (94 components) using Ein-Bild-Rontgen-Analyse (EBRA-Cup) in two groups of patients. In Group A, components were recorded as not being loose intra-operatively at re-revision THA (52 components/48 patients) and Group B components were recorded to be loose at re-revision (42 components/36 patients).

Results

The mean proximal translation and sagittal rotation were significantly higher in Group B than in Group A from three months onwards (p < 0.02). Proximal translation > 1.0 mm within 24 months had a positive predictive value (PPV) of 90% and a specificity of 94%, but a sensitivity of 64%. Proximal translation > 1.0 mm within the first 24 months correctly identified 76 of 94 (81%) of components to be either loose or not loose. However, ten components in Group B (24%) did not migrate proximally above 1.0 mm within the first 60 months.

Conclusion

The high PPV of EBRA-Cup measurements of proximal translation (90%) shows that this can be used in early follow-up to identify patients at risk of aseptic loosening. The absence of proximal translation within the first 60 months indicates a component is not likely to be loose at re-revision THA although it does not exclude late aseptic loosening as a cause of failure.

Cite this article: Bone Joint J 2017;99-B:465–74.