Abstract
Aims
The aim of this study was to report the results of custom-made endoprostheses with extracortical plates plus or minus a short, intramedullary stem aimed at preserving the physis after resection of bone sarcomas in children.
Patients and Methods
Between 2007 and 2017, 18 children aged less than 16 years old who underwent resection of bone sarcomas, leaving ≤ 5 cm of bone from the physis, and reconstruction with a custom-made endoprosthesis were reviewed. Median follow-up was 67 months (interquartile range 45 to 91). The tumours were located in the femur in 11 patients, proximal humerus in six, and proximal tibia in one.
Results
The five-year overall survival rate was 78%. No patient developed local recurrence. The five-year implant survival rate was 79%. In all, 11 patients (61%) developed a complication. Seven patients (39%) required further surgery to treat the complications. Implant failures occurred in three patients (17%) including one patient with aseptic loosening and two patients with implant or periprosthetic fracture. The preserved physis continued to grow at mean 3.3 cm (0 to 14). The mean Musculoskeletal Society score was 88% (67% to 97%).
Conclusion
Custom-made endoprostheses that aim to preserve the physis are a safe and effective option for preserving physeal growth, limb length, and joint function with an acceptable rate of complications.