EFFECT OF FEMORAL HEAD SIZE ON HIP RANGE OF MOTION TO IMPINGEMENT

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 2.



Abstract

INTRODUCTION

Femoral stem impingement can damage an acetabular liner, create polyethylene wear, and potentially lead to dislocation. To avoid component-to-component impingement, many surgeons aim to align acetabular cups based on the “Safe Zone” proposed by Lewinnek. However, a recent study indicates that the historical target values for cup inclination and anteversion defined by Lewinnek et al. may be useful but should not be considered a safe zone. The purpose of this study was to determine the effect of altering femoral head size on hip range-of-motion (ROM) to impingement.

METHODS

Ten healthy subjects were instrumented and asked to perform six motions commonly associated with hip dislocation, including picking up an object, squatting, and low-chair rising. Femur-to-pelvis relative motions were recorded throughout for flexion/extension, abduction/adduction, and internal/external rotation.

A previously reported custom, validated hip ROM three-dimensional simulator was utilized. The user imports implant models, and sets parameters for pelvic tilt, stem version, and specific motions as defined by the subjects. Acetabular cup orientations for abduction and anteversion combinations were chosen. The software was then used to compute minimum clearances or impingement between the components for any hip position.

Graphs for acetabular cup abduction vs. anteversion were generated using a tapered wedge stem with a 132º neck angle, a stem version of 15°, and a pelvic tilt of 0°. The only variable changed was femoral head size. Head sizes reviewed were 32mm, 36mm, and a Dual-Mobility liner with an effective head size of 42mm. All femoral head sizes can be used with a 50mm acetabular cup.

RESULTS

We found that the “Safe Zone” varies considerably depending on the size of the femoral head used for all subjects. A typical plot illustrating the ROM to impingement is presented in Figures 1a-1c. The area to the left of each curve represents an impingement zone for that motion, and to the right, a non-impingement zone. In all cases the non-impingement area is smaller than the Lewinnek safe zone. Motions like picking up an object and squatting had the greatest influence on impingement.

CONCLUSIONS

The true acetabular target for impingement-avoidance motion is much smaller than previously believed and identifies the need to take into account the size of the femoral head to be used. This may explain why approximately 70% of dislocations have been reported to be found even when cups were placed within the Lewinnek safe zone. Certain activities, such as picking up an object and squatting reduce the size of the safe zone. This study supports the need for better patient planning and intraoperative execution for placement of the acetabular component.