LOW WEAR OF HIGHLY CROSS-LINKED POLYETHYLENE IN TOTAL HIP REPLACEMENT AT A MEAN 11-YEAR FOLLOW-UP

European Orthopaedic Research Society (EORS) 24th Annual Meeting, 14–16 September 2016. Part 2.



Abstract

Highly cross-linked polyethylene (HXLPE) is now a common used bearing surface in total hip arthroplasty. Current studies report superior wear rates with the use of HXLPE in total hip arthroplasty. However, there are few studies to support its long term use. The aim of this study is to measure the long term wear of HXLPE and evaluate patient satisfaction at more than 10 years follow up.

44 total hip arthroplasties were performed through a direct lateral approach by a single surgeon. All patients received the same uncemented acetabular component, mean liner thickness was 6.91mm (SD= 0.68). 16 of the femur components were cemented. Outcomes analysed include wear rates, osteolysis, revision rates, SF12 and Oxford hip scores. Wear rate was calculated using computer software (Polyware®) using edge detection software.

Mean age at surgery was 58.9 years (SD= 11.67). The mean follow up was 11.3 years (SD= 1.19). There was no evidence of osteolysis and none had undergone revision surgery. Mean two dimensional wear was 0.38mm (SD= 0.25) and mean wear rate per year was 0.03mm (SD= 0.02, range 0.009 to 0.078). Oxford hip score at last follow up indicated satisfactory joint function (mean= 42 SD= 6.2).

Our results support the use of highly cross-linked polyethylene in primary total hip replacements. The absence of osteolysis and need for revision surgery over a mean of 11.3 years is very encouraging.