header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

THE USE OF A NOVEL HEXAPOD FRAME SYSTEM FOR ACUTE AND CHRONIC TIBIAL AND FEMORAL DEFORMITY CORRECTION

European Orthopaedic Research Society (EORS) 2016, 24th Annual Meeting, 14–16 September 2016. Part 1.



Abstract

The TL Hex (Orthofix) is a relatively new hexapod frame system that we have been using at our institution since August 2013 to treat acute fractures and correct tibial and femoral deformity. We report our initial experience of 48 completed treatments with this novel system in 46 patients and discuss illustrative cases.

For acute fracture, 30 patients (24 male, 7 female) required framing with a mean age of 43 years (range 19–80). One patient underwent bilateral framing. The tibia was involved in all cases. In two cases, the femur also required framing. Open fractures occurred in 13 cases (43.3%).

For elective limb reconstruction, 16 patients (14 male, two female) required framing with a mean age of 35 years (range 16–67). One patient underwent bilateral framing. The tibia was involved in all but one case, which affected the femur. Surgical indications included congenital deformity in four cases, malunion in eight cases, non-union in three cases and chronic infection in two cases.

For acute fractures, the mean frame time was 164 days (range 63–560) and all but one fracture achieved union. Complications included pin, wire or strut failure requiring adjustment (three patients) and pin site infection (six patients). Three patients are being considered for residual deformity correction or treatment of non-union.

In the elective limb reconstruction group, mean frame time was 220 days (range 140–462). All treatments successfully achieved deformity correction and bone union. Complications included two pin site infections. There was no evidence of recurrence of infection in the two osteomyelitis cases.

In conclusion, the TL Hex frame system appears to be a safe and reliable tool for limb reconstruction. We have observed acceptable frame times, low complication rates and almost 100% bony union.