OUR EXPERIENCE WITH A RECALLED HIP RESURFACING IMPLANT AT A MINIMUM FIVE-YEAR FOLLOW-UP

Canadian Orthopaedic Association (COA) and Canadian Orthopaedic Research Society (CORS) Annual Meeting, June 2016; PART 2.



Abstract

Hip resurfacing offers an attractive alternative to conventional total hip arthroplasty in young active patients. It is particularly advantageous for bone preservation for future revisions. Articular Surface Replacement (ASR) is a hip resurfacing prosthesis manufactured by DePuy Orthopaedics Inc. (Warsaw, IN). The manufacturer voluntarily recalled the ASR system in 2010 after an increasing number of product failures. The present study aimed to determine the long-term results in a large cohort of patients who received the ASR prosthesis.

Between February 2004 and August 2010, 592 consecutive hip resurfacings using the ASR (DePuy Orthopaedics Inc., Warsaw, IN) resurfacing implant were performed in 496 patients (391 males and 105 females). The mean age of the patients at the time of the surgery was 54 (range: 25 to 74) years. Osteoarthritis was the most common diagnosis in 575 hips (97.1%). The remaining patients (2.9%) developed secondary degenerative disease from ankylosing spondylitis, avascular necrosis, developmental hip dysplasia, and rheumatoid arthritis. Clinical and radiographic information was available for all patients at the last follow up. Cobalt (Co) and chromium (Cr) levels were measured in 265 patients (298 hips) by inductively coupled plasma-mass spectrometry (ICP-MS).

The average follow up of the study was 8.6 years (range: 5.2 to 11.6 years). The mean Harris hip and UCLA scores significantly improved from 44 and 2 pre-operatively to 85.3 and 7.1 respectively. The median Co and Cr ion level was 3.81 microgram per liter and 2.15 microgram per liter respectively. Twenty-seven patients (5.4%) were found to have blood levels of both Co and Cr ions that were greater than 7 microgram per liter. Fifty-four patients (9.1%) were revised to a total hip arthroplasty. Kaplan-Meier survival analysis showed a survival rate of 87.1% at 8.6 years with revision for any cause and 87.9% if infection is removed. A significantly higher survival rate was observed for the male patients (90.2%, p <0.0001) and for the patients with ASRs with femoral heads diameters larger than 52 mm (90.1%, p=0.0003).

This study confirms that patient selection criteria are of great importance to the overall survivorship of hip resurfacing arthroplasty. Improved clinical results have been reconfirmed with the use of larger diameter femoral heads.


Email: