ACCURACY OF MEASURING PLANAR ANTEVERSION IN ANTEROPOSTERIOR RADIOGRAPH

The International Society for Technology in Arthroplasty (ISTA), 30th Annual Congress, Seoul, South Korea, September 2017. Part 1 of 2.



Abstract

Introduction

The correct anteversion of the acetabular cup is critical to achieve optimal outcome after total hip arthroplasty. While number of method has been described to measure the anteversion in plane anteroposterior and lateral radiograph, it is still controversial which method provides best anteversion measurement. While many of the previous studies used CT scan to validate the anteversion measured in plane anteroposterior radiograph, this may cause potential bias as the anteversion measured in CT scan reflects true anteversion while the anteversion measurement methods in plane radiograph are design to measure the planar anteversion. Thus, in the current study, we tried to find the optimal anteversion measurement method free from the previously described bias.

Material and method

Custom made cup model was developed which enables change in anteversion and inclination. Simple radiograph was taken with the cup in 10° to 70° degree of inclination at 10° increments and for each inclination angle, anteversion was corrected from 0° to 30° at 5° increments. The radiograph was taken with the beam directed at the center of the cup (mimicking hip centered anteroposterior radiograph) and at 9cm medial to the cup (mimicking pelvis anteroposterior radiograph). The measurements were done by two orthopaedic surgeons using methods described by 1) Pradhan et al, 2) Lewinak et al, 3) Widmer et al, and 4) Liaw et al. For each measurements, the anteversion were compared with the actual anteversion.

Result

Interoverver correlation (kappa value) were high in all measurements ranging 0.988 to 0.998. Regardless of how the radiograph was taken, Pradhan method was the most accurate measurement method showing difference of 2.17° ± 1.69° and −2.5° ± 1.93° compare to the actual anteversion respectively for hip centered radiograph and pelvis anteroposterior radiograph. The Widmer method showed the least accuracy (pelvis AP : −6.75° ± 4.62°, hip centered AP : −14.84° ± 4.36°). However, when the anteversion were measured in the safe zone with the inclination in 30° to 50° Liaw's method in hip centered radiograph showed the highest accuracy (1.63° ± 1.4°).

Conclusion

The study indicates that the Pradhan's method may provide the most accurate anteversion measurement. However, with the hip in 30° to 50° inclination, Liaw's method measured from hip centred radiograph will provide most accurate anteversion measurement.