Intraoperative pressure sensors improve soft-tissue balance but not clinical outcomes in total knee arthroplasty: a multicentre randomized controlled trial



Download PDF

Abstract

Aims

Intraoperative pressure sensors allow surgeons to quantify soft-tissue balance during total knee arthroplasty (TKA). The aim of this study was to determine whether using sensors to achieve soft-tissue balance was more effective than manual balancing in improving outcomes in TKA.

Methods

A multicentre randomized trial compared the outcomes of sensor balancing (SB) with manual balancing (MB) in 250 patients (285 TKAs). The primary outcome measure was the mean difference in the four Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4) in the two groups, comparing the preoperative and two-year scores. Secondary outcomes included intraoperative balance data, additional patient-reported outcome measures (PROMs), and functional measures.

Results

There was no significant difference in ΔKOOS4 between the two groups at two years (mean difference 0.4 points (95% confidence interval (CI) -4.6 to 5.4); p = 0.869), and multiple regression found that SB was not associated with a significant ΔKOOS4 (0.2-point increase (95% CI -5.1 to 4.6); p = 0.924). There were no significant differences between groups in other PROMs. Six-minute walking distance was significantly increased in the SB group (mean difference 29 metres; p = 0.015). Four-times as many TKAs were unbalanced in the MB group (36.8% MB vs 9.4% SB; p < 0.001). Irrespective of group assignment, no differences were found in any PROM when increasing ICPD thresholds defined balance.

Conclusion

Despite improved quantitative soft-tissue balance, the use of sensors intraoperatively did not differentially improve the clinical or functional outcomes two years after TKA. These results question whether a more precisely balanced TKA that is guided by sensor data, and often achieved by more balancing interventions, will ultimately have a significant effect on clinical outcomes.

Cite this article: Bone Joint J 2022;104-B(5):604–612.

Take home message

Although it is clear that sensors significantly improve intraoperative soft-tissue balance, this study found their use was not more effective than manual techniques in improving the clinical outcomes of total knee arthropalsty (TKA) two years after surgery.

These results question whether a more precisely balanced TKA that is guided by sensor data, and often achieved by more balancing interventions, will ultimately have a significant effect on clinical outcomes.

Introduction

Although clearly a multifactorial issue, a commonly cited reason for dissatisfaction following total knee arthroplasty (TKA) is failure to restore the soft-tissue balance.1-4 Instability following TKA is also a leading cause of early revision, and ensuring more precise soft-tissue balance may reduce revision rates and improve outcomes.5,6 While computer-assisted navigation can quantify the angular positioning of implants and align the limb precisely,7 historically no such instruments have been available to quantify the soft-tissue balance.

In order to address this issue, intraoperative pressure sensors were introduced for use during TKA. They have been demonstrated to reliably and accurately quantify soft-tissue balance,8-10 and are used to direct ligamentous releases and bony readjustments.11 While several studies have reported improvements in outcomes with the use of sensors, others have shown no significant correlation.12-18 Despite their ability to improve soft-tissue balance, it is not clear whether sensors improve early clinical and functional outcomes in TKA. Furthermore, despite precise balance being considered an important determinant of outcome, it has not been established that it, in fact, improves outcomes compared with TKAs that are not balanced.

The aim of this study, therefore, was to determine whether the use of sensors to attain soft-tissue balance in TKA improved outcomes compared with manual balancing techniques. Secondary aims were to determine if attaining precise balancing, regardless of the use of sensors, improved outcomes. The primary null hypothesis was that in patients undergoing primary TKA, achieving soft-tissue balance with the use of sensors would not lead to a greater improvement in the aggregated mean Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4)19 up to two years after surgery compared with those in whom manual balancing techniques were used. Secondary null hypotheses were that other patient-reported outcome measures (PROMs), functional measures, and quantitative measures of soft-tissue balance would not differentially improve with use of sensors, and that achieving balance, regardless of method, would not improve PROMs. If it can be shown that using sensors to achieve balance improves clinical outcomes in patients undergoing TKA, their routine use may be justified.

Methods

A multicentre, randomized controlled trial (RCT) was undertaken comparing the outcomes of sensor balancing (SB) in patients undergoing TKA, compared with manual balancing (MB). Patients, assessors, and statisticians were blinded to group allocation. Ethical approval was obtained from South Eastern Sydney Local Health District (Approval No. HREC/18/POWH/320). The study was prospectively registered in the Australian New Zealand Clinical Trials Registry (ACTRN#12618000817246p). The protocol was published prior to the start of the trial.20

Eight surgeons undertook operations at 11 sites in public and private Australian hospitals between October 2018 and March 2019. They had been in practice for between eight and 13 years, and each performed between 50 and 300 TKAs per annum. At the beginning of the trial, five surgeons had used pressure sensors for at least one year. The remaining three had used them intermittently in previous TKAs, but then intensively for 15 to 30 cases prior to the start of the study, experience which has been shown to mitigate any learning curve.21

A pragmatic approach was used for the inclusion of patients: all those aged between 20 and 85 years who were scheduled for primary TKA to treat primary osteoarthritis, inflammatory arthritis, or post-traumatic arthritis were eligible. Those undergoing unilateral or bilateral procedures, those with extra-articular deformity from previous fracture or osteotomy, and those with severe stiffness were also included, but those who required constrained prostheses due to significant ligament deficiencies, and those undergoing TKA for acute fracture or tumour, were excluded.

A total of 250 patients underwent 285 TKAs. At two years, 227 patients (90.8%) had complete follow-up. Figure 1 shows the details of the groups at each timepoint. There were 124 patients (141 TKAs) in the MB group and 126 (144 TKAs) in the SB group who received the allocated intervention and whose data were analyzed. There were no significant differences in the baseline or surgical characteristics between groups (Table I).

Fig. 1 
          Flow diagram showing recruitment and allocation of patients. LTFU, lost to follow-up; PROMs, patient-reported outcome measures; TKA, total knee arthroplasty.

Fig. 1

Flow diagram showing recruitment and allocation of patients. LTFU, lost to follow-up; PROMs, patient-reported outcome measures; TKA, total knee arthroplasty.

Table I.

Details of the patients in the study.

Variable Manual balance Sensor balance p-value
Total, n 141 144
Mean age, yrs (SD) 69.2 (8.1) 68.9 (8.7) 0.787*
Mean BMI, kg/m2 (SD) 31.4 (5.4) 31.7 (7.6) 0.664*
Sex, n (%)
Female 83 (58.9) 85 (59.0) > 0.999
Male 53 (37.6) 59 (41.0)
Side, n (%) 0.781
Left 64 (45.4) 62 (43.1)
Right 77 (54.6) 82 (56.9)
Preoperative alignment, n (%) 0.746
Neutral 30 (21.3) 26 (18.1)
Varus 79 (56.0) 86 (59.7)
Valgus 28 (19.9) 27 (18.8)
Operating time, mins (SD) 74.5 (17.6) 75.8 (18.6) 0.544*
Alignment strategy, n (%) 0.575
MA 116 (82.3) 123 (85.4)
KA 25 (17.7) 21 (14.6)
Alignment referencing method, n (%) 0.230
IM femur + tibia 44 (31.2) 43 (29.9)
IM femur/EM tibia 21 (14.9) 20 (13.9)
Patient-specific instrumentation 21 (14.9) 35 (24.3)
Computer-assisted 46 (32.6) 42 (29.2)
Robotic-assisted 9 (6.4) 4 (2.8)
Femoral fixation, n (%) > 0.999
Cemented 139 (98.6) 142 (98.6)
Non-cemented 2 (1.4) 2 (1.4)
Tibial fixation, n (%)
Cemented 141 (100) 144 (100) N/A
Patellar implant, n (%) 0.270
Oval 118 (83.7) 126 (87.5)
Round 6 (4.3) 1 (0.7)
Inset 9 (6.4) 10 (6.9)
Not resurfaced 8 (5.7) 7 (4.9)
Prosthetic stability, n (%) 0.572
Posterior-stabilized 124 (87.9) 130 (90.3)
Cruciate-retaining 17 (12.1) 14 (9.7)
  1. *

    Independent-samples t-test.

  1. Chi-squared test.

  1. EM, extramedullary guide; IM, intramedullary guide; KA, kinematic alignment; MA, mechanical alignment; N/A, not applicable; SD, standard deviation.

After the eligible patients were identified and consent was obtained, they were allocated randomly (1:1) to the MB or SB group using a remote, centralized telephone service. Surgeons were notified of the allocation only after the intraoperative soft-tissue balance was documented, and prior to any balancing procedures. Allocation was stratified according to surgeon and the age and sex of the patient. For patients undergoing bilateral TKA or sequential unilateral TKA on different days, both knees were allocated to the same treatment arm, as randomization was done at the level of the patient, not the knee.

Surgeons used sensor data to balance the knee in the SB group, and in the MB group, they used their usual balancing method, including manual laxity assessments in varying knee positions and spacer block techniques. They remained blinded to the sensor pressures at all times. Pressures were recorded by non-surgical staff both during and after balancing the knee, with the computer screen turned away from the surgical field.

All TKAs were performed using the Legion/Genesis II system (Smith & Nephew, USA). In order to maintain the pragmatism of the study, no restrictions were placed on the technique, including the use of alignment instruments (conventional instruments, patient-specific guides, computer-assisted navigation, or robotic-assisted surgery); alignment strategy (mechanical (MA) or kinematic alignment (KA)); the stability of the prosthesis (cruciate-retaining or posterior-stabilized); the method of fixation (cemented or cementless); and the decision to resurface the patella.

The measurements of compartmental pressures and tibiofemoral congruence were standardized using the following protocol: the arthrotomy was temporarily clipped and the patella reduced; the heel was then supported with one hand while the other hand cradled the posterior aspect of the distal femur; excessive varus, valgus, or compressive forces on the knee were avoided. Compartmental pressures in both groups were recorded once per patient at three positions of flexion of the knee (10°, 45°, and 90°), before and after balancing. In a previous study, we reported high reliability of the sensor measurements between observers at 10°, with reliability decreasing slightly at higher angles of flexion.8 “Balance” in this study was defined as an absolute intercompartmental pressure difference (ICPD) of ≤ 15 psi with a compartmental pressure of between 5 and 40 psi in at least two of three positions of the knee.11 Tibiofemoral congruence was recorded as the angular match between the femoral component and the sensor, with “optimal rotation” defined as ≤ 5° difference at two of three positions of the knee.

In the SB group, balancing was performed as described by Roche et al.22 If an absolute pressure in any one compartment was ≥ 60 psi, or if there was an ICPD ≥ 40 psi, the bone was re-cut. Otherwise, for ICPDs of between 16 and 40 psi, soft-tissue releases were performed. The aim was to achieve a final ICPD of ≤ 15 psi, with individual compartmental pressures of between 5 and 40 psi at all angles of flexion, values that have been associated with improved outcomes.12-16

In the MB group, balancing was performed using the same techniques, but decisions about when to perform balancing and which interventions to use were at the surgeon’s discretion.

The primary outcome measure was the difference in the aggregated mean of the four subscales of the KOOS (ΔKOOS4) between the preoperative values and those two years after surgery. These four subscales (pain, symptoms, function in daily living, and knee-related quality of life) are considered the most specific to recovery after TKA.19 The fifth subscale, function in sport/recreation, has a significant floor effect in these patients and was therefore not included. The KOOS4 scores range from 0 (worst) to 100 (best).

In order to gain a more holistic view, several secondary outcomes were assessed. Additional PROMs, recorded at six months and one and two years postoperatively, included all five subscales of KOOS;19,23 the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC);24 the Forgotten Joint Score (FJS-12);25 and, as a measure of overall health status, the EuroQol five-dimension five-level questionnaire (EQ-5D-5L).26,27

Functional outcome measurements were recorded at six months postoperatively by research physiotherapists at each site who were blinded to the patients’ allocations. Tests included the Timed Up and Go test (TUG),28 Six-Minute Walk Test (6MWT),29 and active range of motion (ROM) of the knee.

Intraoperatively, the proportions of unbalanced knees between the groups were compared after grading them as “mildly unbalanced” (ICPD > 15 psi, or individual compartmental pressure outside 5 to 40 psi in two or more positions of flexion); “moderately unbalanced” (ICPD > 40 psi in at least one position); or “severely unbalanced” (ICPD > 60 psi in at least one position).

Quantitative ICPDs and balancing procedures (soft-tissue releases and bone recuts) were compared between the groups. Finally, differences in clinical outcomes were compared based on the degree of imbalance of the TKA (mildly, moderately, or severely unbalanced), regardless of group assignment.

Adverse events were recorded at all timepoints.

The study was powered for the secondary analysis (balanced versus unbalanced knees, assuming a ratio of 3:1), knowing this would over-power the study for the primary analysis (intention-to-treat, with 1:1 allocation). This ratio was based on previous research, which showed that nearly all patients who underwent sensor-guided balancing were truly balanced, while fewer than half of those who were considered balanced with manual assessment were.30 A sample size of 200 (assuming 150 patients balanced and 50 unbalanced) would provide 90% power to detect an eight-point (minimum clinically important) ΔKOOS4, with a standard deviation of 15 and a level of significance of 0.05.19 This sample size would provide 96% power for the primary analysis. Assuming a 10% loss to follow-up, a sample size of 222 patients was set.

Statistical analysis

Descriptive statistics were used to characterize differences in baseline demographics and intraoperative measurements. Chi-squared tests and independent-samples t-tests were used for comparison of differences between groups. Intention-to-treat analysis was used when analyzing the results. Linear regression models were used to adjust for different baseline and surgical covariates when considering the effect of group assignment (MB or SB) on PROM outcome variables. Covariates included age, sex, BMI, the side of the surgery, surgeon, alignment technique, alignment strategy (KA or MA), and operating time. Significance was set at p < 0.05.

Results

The difference in the mean ΔKOOS4 between the preoperative and two years postoperative values was 44.9 points (-15 to 86) in the MB group and 44.5 points (-10 to 93) in the SB group. The mean difference in the change in KOOS4 score between the groups was not significant (0.4 points (95% confidence interval (CI) -4.6 to 5.4); p = 0.869, independent-samples t-test). Multiple regression of the effect of group allocation on the primary outcome measure, adjusting for baseline and surgical covariates showed that the use of sensors was not associated with a statistically significant or clinically important ΔKOOS4 (0.2-point increase (95% CI -5.1 to 4.6); R2 6.7%; p = 0.924). The assumptions for linear regression were satisfied.

There were no differences between MB and SB groups when comparing preoperative, postoperative, or change of score at six months, one year, and two years for KOOS subscales, FJS-12, WOMAC, and EQ visual analogue scale for general health (Table II and Figure 2). Furthermore, when assessing the effect of group allocation and adjusting for baseline and surgical covariates, no significant differences in FJS-12, WOMAC, and EQ5D-5L scores were found.

Table II.

Mean Knee Osteoarthritis Outcome Scores and other patient-reported outcome measures for the manually balanced and sensor-balanced groups.

Outcome Mean score (SD) Mean between-group difference at 2 yrs (95% CI) p-value*
Preoperative 6 mths 1 yr 2 yrs
KOOS 4
MB 38.0 (14.3) 73.4 (16.9) 80.0 (15.7) 82.1 (15.5) 0.4 (-4.6 to 5.4) 0.869
SB 38.1 (15.0) 75.1 (14.6) 80.8 (14.6) 82.7 (15.1)
KOOS symptoms
MB 42.5 (20.4) 70.5 (18.1) 78.7 (15.7) 81.9 (14.4) 2.4 (-3.3 to 8.1) 0.402
SB 43.6 (18.1) 73.0 (16.8) 79.6 (15.1) 81.7 (16.4)
KOOS pain
MB 39.3 (15.9) 78.8 (17.4) 84.7 (14.5) 86.6 (15.3) 1.7 (-3.8 to 7.1) 0.548
SB 39.6 (18.2) 79.5 (15.6) 85.7 (16.0) 86.4 (14.9)
KOOS ADL
MB 43.1 (17.2) 78.9 (17.6) 83.5 (18.1) 85.0 (15.6) -0.6 (-6.0 to 4.8) 0.836
SB 43.3 (20.0) 80.9 (14.1) 85.3 (15.5) 86.7 (14.8)
KOOS QoL
MB 24.8 (16.2) 65.1 (23.7) 73.1 (22.5) 75.1 (23.0) 0.0 (-6.5 to 6.5) 0.993
SB 24.5 (14.0) 67.3 (20.6) 72.6 (19.5) 76.1 (20.9)
KOOS sports
MB 15.1 (18.2) 50.2 (24.0) 57.4 (28.3) 54.6 (29.2) -3.6 (-11.1 to 3.9) 0.344
SB 14.7 (16.1) 52.5 (25.1) 58.8 (25.4) 59.0 (27.0)
FJS-12
MB 12.8 (15.5) 47.6 (28.4) 57.9 (29.6) 63.6 (29.0) 2.2 (-5.4 to 9.8) 0.563
SB 11.8 (14.5) 47.6 (26.0) 53.9 (30.5) 60.0 (29.2)
EQ-VAS
MB 71.1 (19.0) 82.1 (12.6) 81.8 (13.7) 80.1 (15.9) -2.5 (-7.3 to 2.4) 0.325
SB 69.5 (19.4) 80.8 (13.6) 82.0 (14.8) 81.3 (14.8)
WOMAC total
MB 54.6 (15.7) 20.7 (16.3) 15.7 (15.6) 14.0 (14.4) 0.1 (-4.9 to 5.1) 0.971
SB 54.5 (18.1) 18.8 (13.1) 14.2 (14.1) 13.0 (13.8)
  1. *

    Independent-samples t-test.

  1. ADL, activities of daily living; CI, confidence interval; EQ-VAS, EuroQol visual analogue scale; FJS-12, Forgotten Joint Score; KOOS, Knee injury and Osteoarthritis Outcome Score; MB, manual balance; PROMs, patient-reported outcome measures; QoL, quality of life; SB, sensor balance; SD, standard deviation; WOMAC, Western Ontario and McMaster Universities Osteoarthritis Index.

Fig. 2 
          Change in the aggregated mean of four subscales of the Knee injury and Osteoarthritis Outcome score (KOOS4) over two years comparing the manually and sensor balanced groups.

Fig. 2

Change in the aggregated mean of four subscales of the Knee injury and Osteoarthritis Outcome score (KOOS4) over two years comparing the manually and sensor balanced groups.

Six months postoperatively, those in the SB group had a significantly longer mean walking distance for the 6MWT (366 metres MB group, 395 metres SB group; mean difference 29 m; p = 0.015, independent-samples t-test). There were no significant differences in the mean TUG test or active ROM of the knee between groups (Table III).

Table III.

Measures of functional outcome in the manually and sensor balanced groups, six months postoperatively.

Outcome Manual balance Sensor balance p-value*
Mean Timed Up and Go test, secs (SD) 10.0 (2.4) 9.7 (4.5) 0.580
Mean Six Minute Walk Test, metres (SD) 366.2 (89.3) 395.3 (96.3) 0.015
Mean active ROM, ° (SD)
Extension 8.7 (5.5) 8.7 (6.1) 0.868
Flexion 114.1 (10.7) 113.8 (12.1) 0.832
Arc of motion 105.3 (12.9) 105.3 (15.07) 0.960
  1. *

    Independent-samples t-test.

  1. ROM, range of motion; SD, standard deviation.

As indicated in Table IV, four times as many TKAs in the MB group were “mildly unbalanced” compared with the SB group (50 vs 13; p < 0.001, chi-squared test). The mean ICPDs were significantly higher in the MB group at all angles of flexion, with mean ICPDs outside the balanced range at 10° and 45° (Table V and Figure 3). The use of sensors increased the rate of medial soft-tissue releases, all soft-tissue releases, and bone recuts (Table VI).

Table IV.

Proportion of patients with an unbalanced total knee arthroplasty in the manual balance and sensor balance groups in three positions of knee flexion (10°, 45°, and 90°).

Definition Manual balance, n (%) Sensor balance, n (%) p-value
Mildly unbalanced in 2 of 3 knee positions (ICPD ≥ 15 psi) 50 (36.8) 13 (9.4) < 0.001*
Moderately unbalanced (ICPD ≥ 40 psi) 15 (11.0) 1 (0.7) < 0.001
Severely unbalanced (ICPD ≥ 60 psi) 4 (2.9) 0 (0.0) 0.125
  1. *

    Chi-squared test.

  1. Fisher’s exact test.

  1. ICPD, intercompartmental pressure difference; psi, pounds per square inch.

Table V.

Intercompartmental pressure differences between the manual balance and sensor balance groups at 10°, 45° and 90° of knee flexion.

Knee flexion Mean ICPD, psi (SD) p-value*
Manual balance Sensor balance
10° 19.9 (17.3) 7.6 (6.8) < 0.001
45° 15.7 (13.9) 8.5 (9.1) < 0.001
90° 12.2 (8.8) 8.4 (6.8) < 0.001
  1. *

    Independent-samples t-test.

  1. Mean ICPDs ≥ 15 psi were defined as “unbalanced”.

  1. ICPD, intercompartmental pressure difference; psi, pounds per square inch; SD, standard deviation.

Fig. 3 
          Box plots of intercompartmental pressure differences (ICPD; pounds per square inch) comparing manually balanced and sensor balanced groups at a) 10°, b) 45°, and c) 90° of knee flexion.

Fig. 3

Box plots of intercompartmental pressure differences (ICPD; pounds per square inch) comparing manually balanced and sensor balanced groups at a) 10°, b) 45°, and c) 90° of knee flexion.

Table VI.

Difference in proportions of balancing procedures between the manual balance and sensor balance groups.

Balancing procedure Manual balance (%) Sensor balance (%) p-value*
Medial soft-tissue release 10 25.4 0.013
Lateral soft-tissue release 19.3 18.3 0.595
Any soft-tissue release 28.6 40.1 0.045
Any bone recut 8.6 21.1 0.004
  1. *

    Chi-squared test.

Regardless of group, TKAs with any degree of imbalance (mild, moderate, or severe) showed no significant differences in PROMs at six months or one or two years when compared with balanced knees.

Six TKAs (4.8%) the MB group and three (2.4%) in the SB group were treated with a manipulation under anaesthesia (MUA) (Table VII). One patient in each group underwent a soft-tissue release postoperatively for iliotibial band pain. One patient in the MB group had a debridement with implant retention for deep infection. There were six unrelated deaths during the study period, three in each group.

Table VII.

Differences between manually and sensor-balanced groups in rates of adverse events.

Adverse event Manual balance, n (%) Sensor balance, n (%)
Manipulation under anaesthesia 6 (4.8) 3 (2.4)
Tendon release 1 (0.8) 1 (0.8)
Superficial SSI requiring oral antibiotics 4 (3.2) 3 (2.4)
Superficial SSI requiring parenteral antibiotics 1 (0.8) 1 (0.8)
SSI requiring debridement and liner exchange 1 (0.8) 0 (0)
Death (unrelated cause) 3 (2.4) 3 (2.4)
  1. SSI, surgical site infection.

Discussion

The null hypothesis could not be rejected on the basis of these results, in that the use of sensors intraoperatively did not lead to a greater improvement in KOOS4 scores when compared with manual balancing at two years postoperatively. Furthermore, none of the other knee-specific or general health outcomes improved with the use of sensors when compared with MB. Sensors did, however, reduce the proportion of unbalanced knees four-fold, improve the quantitative balance of the knee by increasing the rates of balancing procedures that were undertaken, and increase the six-minute walking distance.

One secondary outcome of specific note is that there were no differences in any outcomes when quantitatively balanced knees were compared with knees with mild, moderate, or severe imbalance. This challenges the long-standing belief that precisely balanced TKAs, with or without ligament releases, result in better outcomes. This finding is of particular importance with the growing adoption of KA, a technique that results in better soft-tissue balance than MA, while also reducing the requirements for ligament releases.31,32

There are several key features that support the strength of these findings compared with the literature to date.12-14,18,30,33 First, they are likely to be more generalizable because they are from a multicentre trial that included numerous surgeons, a variety of operating techniques, and broad inclusion criteria. Secondly, it is the first study to report two-year outcomes, the postoperative time at which maximal improvement occurs, particularly when considering joint awareness.34 Thirdly, regression was used to control for the effects of potential confounding factors in the analysis of outcomes. Fourthly, functional outcomes were analyzed as another important and practical measure of surgical success. Lastly, this is the first study to examine whether precise soft-tissue balance, independent of the use of sensors, contributes to improved clinical outcomes.

Two previous observational studies comparing MB to SB showed conflicting clinical results at six months after surgery.13,17 In a non-randomized trial, Golladay et al14 found that patients with a quantifiably balanced knee at six months had significantly higher satisfaction and better mean FJS-12 scores. A recent RCT by Wood et al,18 which compared MB with SB in 152 TKAs, found no differences in the PROMs (Knee Society Scores35 and Oxford Knee Scores36), between the groups at one year postoperatively. Previous cohort studies suggested a correlation between intraoperative sensor measurements and clinical outcomes.12,16 In the current study, no differences were found in any outcome score when incremental ICPD boundaries were applied to define balance.

Several previous studies have also found that SB improves soft-tissue balance by increasing the rates of balancing procedures, which are used.30,37 Similar rates of imbalance were reported by Wood et al18 (MB group 36%, SB group 5.3%) to our findings (MB 36.8%, SB 9.4%). This provides further support for the view that sensors reduce soft-tissue imbalance, but they do so by increasing the numbers of balancing interventions required. Whether the use of robotic technologies also reduces the rates of imbalance by reducing the rates of ligament release through virtual pre-resection positioning of the components remains to be seen. Another important metric is the reduced rates of MUA that have been documented with the use of sensors in TKA.13,38 In the current study, although twice as many patients in the MB group required a MUA, the difference was not statistically significant.

The SB group had a longer mean 6MWT by 29 metres, which is within the reported minimum clinically important difference (MCID) threshold of between 14 and 30 m.39 This test has been shown to be an excellent predictor of capacity for longer, more functional walking after TKA.29 This may be a spurious statistical finding, however, as other functional measures were not improved, and the degree of soft-tissue imbalance did not influence the outcome of this test when analyzed separately.

The study had limitations. First, the surgeons were high-volume arthroplasty surgeons with significant experience in balancing techniques including KA and gap balancing, and this may not be representative of routine orthopaedic practice. Second, it is possible that the current definitions of soft-tissue balance do not correlate with constitutional joint laxities, which vary in different angles of knee flexion.40 Thirdly, only one prosthetic system was used, and clinical outcomes may vary with other systems and levels of implant stability. Fourth, the broad inclusion criteria and variation between the surgeons’ operating techniques increase the study’s pragmatism, and therefore, potentially, the generalizability of the findings, but they may reduce the overall sensitivity of the results when making direct comparisons between the groups. Fifth, there were small numbers of unbalanced knees, and because the study was not powered to tease out the secondary endpoint of patient outcomes in balanced versus unbalanced TKAs, the true effect of precise balance theoretically could have been underestimated. Finally, potentially deleterious effects of increased rates of ligament release and corresponding damage to the soft-tissue envelope cannot be ignored when interpreting these findings. Further research should investigate whether kinematically aligned TKAs may have better outcomes than the ligament lengthening often required in more neutrally aligned TKAs.

Although it is clear that sensors significantly improve intraoperative soft-tissue balance, we found that their use was not more effective than manual techniques in improving the clinical outcomes of TKA two years postoperatively. As a corollary, these results question whether a more precisely balanced TKA that is guided by sensor data, and often achieved by an increased rate of balancing interventions, will ultimately have a significant effect on clinical outcomes. Future research is required to examine the long-term benefit from precise soft-tissue balance, including the evaluation of prosthetic survival and joint biomechanics.


Correspondence should be sent to Samuel J. MacDessi. E-mail:

References

1. Ali A , Sundberg M , Robertsson O , et al. Dissatisfied patients after total knee arthroplasty: a registry study involving 114 patients with 8-13 years of followup . Acta Orthop . 2014 ; 85 ( 3 ): 229 233 . Crossref PubMed Google Scholar

2. Choi YJ , Ra HJ . Patient satisfaction after total knee arthroplasty . Knee Surg Relat Res . 2016 ; 28 ( 1 ): 1 15 . Crossref Google Scholar

3. Williams DP , O’Brien S , Doran E , et al. Early postoperative predictors of satisfaction following total knee arthroplasty . Knee . 2013 ; 20 ( 6 ): 442 446 . Crossref PubMed Google Scholar

4. Dunbar MJ , Richardson G , Robertsson O . I can't get no satisfaction after my total knee replacement: rhymes and reasons . Bone Joint J . 2013 ; 95-B ( 11 Suppl A ): 148 152 . Crossref PubMed Google Scholar

5. Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) . Hip, Knee & Shoulder Arthroplasty Annual Report 2021, AOA, Adelaide, South Australia . 2021 . https://aoanjrr.sahmri.com/documents/10180/712282/Hip%2C+Knee+%26+Shoulder+Arthroplasty/bb011aed-ca6c-2c5e-f1e1-39b4150bc693 ( date last accessed 10 March 2022 ). Google Scholar

6. No authors listed . NJR 16th Annual Report . National Joint Registry . 2019 . https://reports.njrcentre.org.uk/portals/0/pdfdownloads/njr%2016th%20annual%20report%202019.pdf ( date last accessed 10 March 2022 ). Google Scholar

7. Hetaimish BM , Khan MM , Simunovic N , Al-Harbi HH , Bhandari M , Zalzal PK . Meta-analysis of navigation vs conventional total knee arthroplasty . J Arthroplasty . 2012 ; 27 ( 6 ): 1177 1182 . Crossref PubMed Google Scholar

8. Thompson K , Griffiths-Jones W , Frendin L , et al. Interobserver agreement of sensor-derived compartmental pressure measurements in computer-assisted total knee arthroplasty . Knee . 2020 ; 27 ( 3 ): 717 722 . Crossref PubMed Google Scholar

9. MacDessi SJ , Wood JA , Diwan AD , Harris IA , SENSOR BALANCE Study Group . Surgeon-defined assessment is a poor predictor of knee balance in total knee arthroplasty: a prospective, multicenter study . Knee Surg Sports Traumatol Arthrosc . 2021 ; 29 ( 2 ): 498 506 . Crossref PubMed Google Scholar

10. MacDessi SJ , Gharaibeh MA , Harris IA . How accurately can soft tissue balance be determined in total knee arthroplasty? J Arthroplasty . 2019 ; 34 ( 2 ): 290 294 . Crossref PubMed Google Scholar

11. Gustke K . Use of smart trials for soft-tissue balancing in total knee replacement surgery . J Bone Joint Surg Br . 2012 ; 94-B ( 11 Suppl A ): 147 150 . Crossref PubMed Google Scholar

12. Gustke KA , Golladay GJ , Roche MW , Jerry GJ , Elson LC , Anderson CR . Increased satisfaction after total knee replacement using sensor-guided technology . Bone Joint J . 2014 ; 96-B ( 10 ): 1333 1338 . Crossref PubMed Google Scholar

13. Chow JC , Breslauer L . The use of intraoperative sensors significantly increases the patient-reported rate of improvement in primary total knee arthroplasty . Orthopedics . 2017 ; 40 ( 4 ): e648 e651 . Crossref PubMed Google Scholar

14. Golladay GJ , Bradbury TL , Gordon AC , et al. Are patients more satisfied with a balanced total knee arthroplasty . J Arthroplasty . 2019 ; 34 ( 7S ): S195 S200 . Crossref PubMed Google Scholar

15. Gustke KA , Golladay GJ , Roche MW , Elson LC , Anderson CR . Primary TKA patients with quantifiably balanced soft-tissue achieve significant clinical gains sooner than unbalanced patients . Adv Orthop . 2014 ; 2014 : 628695 . Crossref PubMed Google Scholar

16. Meneghini RM , Ziemba-Davis MM , Lovro LR , Ireland PH , Damer BM . Can intraoperative sensors determine the "target" ligament balance? Early outcomes in total knee arthroplasty . J Arthroplasty . 2016 ; 31 ( 10 ): 2181 2187 . Crossref PubMed Google Scholar

17. Song SJ , Kang SG , Lee YJ , Kim KI , Park CH . An intraoperative load sensor did not improve the early postoperative results of posterior-stabilized TKA for osteoarthritis with varus deformities . Knee Surg Sports Traumatol Arthrosc . 2019 ; 27 ( 5 ): 1671 1679 . Crossref Google Scholar

18. Wood TJ , Winemaker MJ , Williams DS , Petruccelli DT , Tushinski DM , de Beer JV . Randomized controlled trial of sensor-guided knee balancing compared to standard balancing technique in total knee arthroplasty . J Arthroplasty . 2021 ; 36 ( 3 ): 953 957 . Crossref PubMed Google Scholar

19. Roos EM . The 2012 User’s guide to: Knee injury and Osteoarthritis Outcome Score (KOOS) . 2012 . http://www.koos.nu/index.html ( date last accessed 29 April 2021 ). Google Scholar

20. MacDessi SJ , Bhimani A , Burns AWR , et al. Does soft tissue balancing using intraoperative pressure sensors improve clinical outcomes in total knee arthroplasty? A protocol of a multicentre randomised controlled trial . BMJ Open . 2019 ; 9 ( 5 ): e027812 . Crossref PubMed Google Scholar

21. Gharaibeh MA , Chen DB , MacDessi SJ . Soft tissue balancing in total knee arthroplasty using sensor-guided assessment: is there a learning curve . ANZ J Surg . 2018 ; 88 ( 5 ): 497 501 . Crossref PubMed Google Scholar

22. Roche M , Elson L , Anderson C . Dynamic soft tissue balancing in total knee arthroplasty . Orthop Clin North Am . 2014 ; 45 ( 2 ): 157 165 . Crossref PubMed Google Scholar

23. Roos EM , Lohmander LS . The Knee injury and Osteoarthritis Outcome Score (KOOS): from joint injury to osteoarthritis . Health Qual Life Outcomes . 2003 ; 1 : 64 . Crossref PubMed Google Scholar

24. Roos EM , Klässbo M , Lohmander LS . WOMAC Osteoarthritis Index: Reliability, validity, and responsiveness in patients with arthroscopically assessed osteoarthritis . Scand J Rheumatol . 1999 ; 28 ( 4 ): 210 215 . Crossref Google Scholar

25. Hamilton DF , Loth FL , Giesinger JM , et al. Validation of the English language Forgotten Joint Score-12 as an outcome measure for total hip and knee arthroplasty in a British population . Bone Joint J . 2017 ; 99-B ( 2 ): 218 224 . Crossref PubMed Google Scholar

26. EuroQol Group . EuroQol--a new facility for the measurement of health-related quality of life . Health Policy . 1990 ; 16 ( 3 ): 199 208 . Crossref PubMed Google Scholar

27. Herdman M , Gudex C , Lloyd A , et al. Development and preliminary testing of the new five-level version of EQ-5D (EQ-5D-5L) ). Qual Life Res . 2011 ; 20 ( 10 ): 1727 1736 . Crossref PubMed Google Scholar

28. Podsiadlo D , Richardson S . The timed "Up & Go": a test of basic functional mobility for frail elderly persons . J Am Geriatr Soc . 1991 ; 39 ( 2 ): 142 148 . Crossref Google Scholar

29. Ko V , Naylor JM , Harris IA , Crosbie J , Yeo AET . The six-minute walk test is an excellent predictor of functional ambulation after total knee arthroplasty . BMC Musculoskelet Disord . 2013 ; 14 : 145 . Crossref PubMed Google Scholar

30. Cho K-J , Seon J-K , Jang W-Y , Park C-G , Song E-K . Objective quantification of ligament balancing using VERASENSE in measured resection and modified gap balance total knee arthroplasty . BMC Musculoskelet Disord . 2018 ; 19 ( 1 ): 266 . Crossref PubMed Google Scholar

31. MacDessi SJ , Griffiths-Jones W , Chen DB , et al. Restoring the constitutional alignment with a restrictive kinematic protocol improves quantitative soft-tissue balance in total knee arthroplasty: a randomized controlled trial . Bone Joint J . 2020 ; 102-B ( 1 ): 117 124 . Crossref PubMed Google Scholar

32. Chang JS , Kayani B , Wallace C , Haddad FS . Functional alignment achieves soft tissue balance in total knee arthroplasty as measured with quantitative sensor-guided technology . Bone Joint J . 2021 ; 103-B ( 3 ): 507 514 . Crossref Google Scholar

33. Gustke KA , Golladay GJ , Roche MW , Elson LC , Anderson CR . A new method for defining balance: promising short-term clinical outcomes of sensor-guided TKA . J Arthroplasty . 2014 ; 29 ( 5 ): 955 960 . Crossref PubMed Google Scholar

34. Giesinger K , Hamilton DF , Jost B , Holzner B , Giesinger JM . Comparative responsiveness of outcome measures for total knee arthroplasty . Osteoarthr Cartil . 2014 ; 22 ( 2 ): 184 189 . Crossref PubMed Google Scholar

35. Noble PC , Scuderi GR , Brekke AC . Development of a new knee society scoring system . Clin Orthop Relat Res . 2012 ; 470 ( 1 ): 20 32 . Crossref PubMed Google Scholar

36. Dawson J , Fitzpatrick R , Murray D , Carr A . Questionnaire on the perceptions of patients about total knee replacement . J Bone Joint Surg Br . 1998 ; 80-B ( 1 ): 63 69 . Crossref PubMed Google Scholar

37. Gustke KA , Golladay GJ , Roche MW , Elson LC , Anderson CR . A targeted approach to ligament balancing using kinetic sensors . J Arthroplasty . 2017 ; 32 ( 7 ): 2127 2132 . Crossref PubMed Google Scholar

38. Geller JA , Lakra A , Murtaugh T . The use of electronic sensor device to augment ligament balancing leads to a lower rate of arthrofibrosis after total knee arthroplasty . J Arthroplasty . 2017 ; 32 ( 5 ): 1502 1504 . Crossref PubMed Google Scholar

39. Bohannon RW , Crouch R . Minimal clinically important difference for change in 6-minute walk test distance of adults with pathology: a systematic review . J Eval Clin Pract . 2017 ; 23 ( 2 ): 377 381 . Crossref PubMed Google Scholar

40. Park S-J , Seon J-K , Park J-K , Song E-K . Effect of PCL on flexion-extension gaps and femoral component decision in TKA . Orthopedics . 2009 ; 32 ( 10 Suppl ): 22 25 . Crossref PubMed Google Scholar

Author contributions

S. J. MacDessi: Conceptualization, Methodology, Funding acquisition, Supervision, Investigation, Formal analysis, Validation, Writing – original draft, Writing – review & editing.

J. A. Wood: Funding acquisition, Project administration, Data curation, Validation, Visualization, Writing – review & editing.

A. Diwan: Supervision, Writing – review & editing.

I. A. Harris: Supervision, Formal analysis, Writing – review & editing.

SENSOR BALANCE Study Group

The following authors contributed to the investigation and writing of this study:

Aziz Bhimani (MBBS, FRACS, FAOrthA), Consulting Orthopaedic Surgeon, Wollongong Public Hospital, Wollongong, Australia, Wollongong Private Hospital, Wollongong, Australia.

Alexander Burns (MBBS, FRACS, FAOrthA), Consulting Orthopaedic Surgeon, Canberra Private Hospital, Canberra, Australia; Calvary John James Private Hospital, Canberra, Australia.

Darren Chen (MBBS, FRACS, FAOrthA), Consulting Orthopaedic Surgeon, St George Private Hospital, Sydney, Australia Canterbury Public Hospital, Sydney, Australia.

Anthony Leong (MBBS, FRACS, FAOrthA), Consulting Orthopaedic Surgeon, Wollongong Public Hospital, Wollongong, Australia; Wollongong Private Hospital, Wollongong, Australia; Kareena Private Hospital, Sydney, Australia.

Jonathan Mulford (MBBS, FRACS, FAOrthA), Consulting Orthopaedic Surgeon, Calvary St Vincent’s Hospital, Launceston, Australia.

Richard Walker (MBBS, FRACS, FAOrthA), Consulting Orthopaedic Surgeon, Sydney Southwest Private Hospital, Sydney, Australia; Fairfield Public Hospital, Sydney, Australia.

Robert Molnar (MBBS, FRACS, FAOrthA), Consulting Orthopaedic Surgeon, St George Private Hospital, Sydney, Australia; Sutherland Hospital, Sydney, Australia.

Funding statement

The authors disclose receipt of the following financial or material support for the research, authorship, and/or publication of this article: a research grant was received from the Ramsay Hospital Research Foundation (Australia) to assist with administration of the project.

ICMJE COI statement

S. J. Macdessi reports a grant from Smith & Nephew, unrelated to this study.

Ethical review statement

Ethical approval was obtained from South Eastern Sydney Local Health District (Approval No, HREC/18/POWH/320). The protocol was published prior to the start of the trial.

Open access funding

There was no external funding for the open access fee.

Open access statement

This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND 4.0) licence, which permits the copying and redistribution of the work only, and provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc-nd/4.0/

Trial registration number

Australian New Zealand Clinical Trials Registry (ACTRN#12618000817246p)

Twitter

Follow S. J. MacDessi @samuelmacdessi

This article was primary edited by J. Scott.