header advert
Results 1 - 2 of 2
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 824 - 827
1 Jun 2011
Wanner S Gstöttner M Meirer R Hausdorfer J Fille M Stöckl B

Biofilm-associated infections in wounds or on implants are difficult to treat. Eradication of the bacteria is nearly always impossible, despite the use of specific antibiotics. The bactericidal effects of high-energy extracorporeal shock waves on Staphylococcus aureus have been reported, but the effect of low-energy shock waves on staphylococci and staphylococcal biofilms has not been investigated. In this study, biofilms grown on stainless steel washers were examined by electron microscopy. We tested ten experimental groups with Staph. aureus-coated washers and eight groups with Staph. epidermidis.

The biofilm-cultured washers were exposed to low-energy shock waves at 0.16 mJ/mm2 for 500 impulses. The washers were then treated with cefuroxime, rifampicin and fosfomycin, both alone and in combination. All tests were carried out in triplicate. Viable cells were counted to determine the bactericidal effect.

The control groups of Staph. aureus and Staph. epidermidis revealed a cell count of 6 × 108 colony-forming units/ml. Complete eradication was achieved using the combination of antibiotic therapy (single antibiotic in Staph. aureus, a combination in Staph. epidermidis) and shock wave application (p < 0.01).

We conclude that shock waves combined with antibiotics could be tested in an in vitro model of infection.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 741 - 744
1 May 2005
Beaulé PE Krismer M Mayrhofer P Wanner S Le Duff M Mattesich M Stoeckl B Amstutz HC Biedermann R

Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse - femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers.

The accuracy of the method was found to be ± 1.6 mm for the x-direction and ± 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively.

EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.