header advert
Results 1 - 14 of 14
Results per page:
The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 352 - 358
1 Apr 2024
Wilson JM Trousdale RT Bedard NA Lewallen DG Berry DJ Abdel MP

Aims

Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct.

Methods

We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 598 - 603
1 May 2022
Siljander MP Gausden EB Wooster BM Karczewski D Sierra RJ Trousdale RT Abdel MP

Aims

The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes.

Methods

We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1103 - 1110
1 Jun 2021
Tetreault MW Hines JT Berry DJ Pagnano MW Trousdale RT Abdel MP

Aims

This study aimed to determine outcomes of isolated tibial insert exchange (ITIE) during revision total knee arthroplasty (TKA).

Methods

From 1985 to 2016, 270 ITIEs were performed at one institution for instability (55%, n = 148), polyethylene wear (39%, n = 105), insert fracture/dissociation (5%, n = 14), or stiffness (1%, n = 3). Patients with component loosening, implant malposition, infection, and extensor mechanism problems were excluded.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 57 - 61
1 Jun 2019
Chalmers BP Mangold DG Hanssen AD Pagnano MW Trousdale RT Abdel MP

Aims

Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR.

Patients and Methods

Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 923 - 924
1 Jul 2018
Abdel MP Tibbo ME Stuart MJ Trousdale RT Hanssen AD Pagnano MW


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 925 - 929
1 Jul 2018
Abdel MP Tibbo ME Stuart MJ Trousdale RT Hanssen AD Pagnano MW

Aims

It has been suggested that mobile-bearing total knee arthroplasty (TKA) might lead to better outcomes by accommodating some femorotibial rotational mismatch, thereby reducing contact stresses and polyethylene wear. The aim of this study was to determine whether there is a difference between fixed- and mobile-bearing versions of a contemporary TKA with respect to durability, range of movement (ROM) and function, ten years postoperatively.

Patients and Methods

A total of 240 patients who were enrolled in this randomized controlled trial (RCT) underwent a primary cemented TKA with one of three tibial components (all-polyethylene fixed-bearing, modular metal-backed fixed-bearing and mobile-bearing). Patients were reviewed at a median follow-up of ten years (IQR 9.2 to 10.4).


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1192 - 1196
1 Sep 2015
Amanatullah DF Siman H Pallante GD Haber DB Sierra RJ Trousdale RT

When fracture of an extensively porous-coated femoral component occurs, its removal at revision total hip arthroplasty (THA) may require a femoral osteotomy and the use of a trephine. The remaining cortical bone after using the trephine may develop thermally induced necrosis. A retrospective review identified 11 fractured, well-fixed, uncemented, extensively porous-coated femoral components requiring removal using a trephine with a minimum of two years of follow-up.

The mean time to failure was 4.6 years (1.7 to 9.1, standard deviation (sd) 2.3). These were revised using a larger extensively porous coated component, fluted tapered modular component, a proximally coated modular component, or a proximal femoral replacement. The mean clinical follow-up after revision THA was 4.9 years (2 to 22, sd 3.1). The mean diameter of the femoral component increased from 12.7 mm (sd 1.9) to 16.2 mm (sd 3.4; p >  0.001). Two revision components had radiographic evidence of subsidence that remained radiographically stable at final follow-up. The most common post-operative complication was instability affecting six patients (54.5%) on at least one occasion.

A total of four patients (36.4%) required further revision: three for instability and one for fracture of the revision component. There was no statistically significant difference in the mean Harris hip score before implant fracture (82.4; sd 18.3) and after trephine removal and revision THA (81.2; sd 14.8, p = 0.918).

These findings suggest that removal of a fractured, well-fixed, uncemented, extensively porous-coated femoral component using a trephine does not compromise subsequent fixation at revision THA and the patient’s pre-operative level of function can be restored. However, the loss of proximal bone stock before revision may be associated with a high rate of dislocation post-operatively.

Cite this article: Bone Joint J 2015;97-B:1192–6.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 312 - 317
1 Mar 2015
Amanatullah DF Howard JL Siman H Trousdale RT Mabry TM Berry DJ

Revision total hip arthroplasty (THA) is challenging when there is severe loss of bone in the proximal femur. The purpose of this study was to evaluate the clinical and radiographic outcomes of revision THA in patients with severe proximal femoral bone loss treated with a fluted, tapered, modular femoral component. Between January 1998 and December 2004, 92 revision THAs were performed in 92 patients using a single fluted, tapered, modular femoral stem design. Pre-operative diagnoses included aseptic loosening, infection and peri-prosthetic fracture. Bone loss was categorised pre-operatively as Paprosky types III-IV, or Vancouver B3 in patients with a peri-prosthetic fracture. The mean clinical follow-up was 6.4 years (2 to 12). A total of 47 patients had peri-operative complications, 27 of whom required further surgery. However, most of these further operations involved retention of a well-fixed femoral stem, and 88/92 femoral components (97%) remained in situ. Of the four components requiring revision, three were revised for infection and were well fixed at the time of revision; only one (1%) was revised for aseptic loosening. The most common complications were post-operative instability (17 hips, 19%) and intra-operative femoral fracture during insertion of the stem (11 hips, 12%). Diaphyseal stress shielding was noted in 20 hips (22%). There were no fractures of the femoral component. At the final follow-up 78% of patients had minimal or no pain.

Revision THA in patients with extensive proximal femoral bone loss using the Link MP fluted, tapered, modular stem led to a high rate of osseointegration of the stem at mid-term follow-up.

Cite this article: Bone Joint J 2015; 97-B:312–17.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 29 - 34
1 Jan 2015
Novais EN Potter GD Clohisy JC Millis MB Kim YJ Trousdale RT Carry PM Sierra RJ

Obesity is a risk factor for complications following many orthopaedic procedures. The purpose of this study was to investigate whether obesity was an independent risk factor increasing the rate of complications following periacetabular osteotomy (PAO) and to determine whether radiographic correction after PAO was affected by obesity.

We retrospectively collected demographic, clinical and radiographic data on 280 patients (231 women; 82.5% and 49 men; 17.5%) who were followed for a mean of 48 months (12 to 60) after PAO. A total of 65 patients (23.2%) were obese (body mass index (BMI) > 30 kg/m2). Univariate and multivariate analysis demonstrated that BMI was an independent risk factor associated with the severity of the complications. The average probability of a patient developing a major complication was 22% (95% confidence interval (CI) 11.78 to 38.21) for an obese patient compared with 3% (95% CI 1.39 to 6.58) for a non-obese patient The odds of a patient developing a major complication were 11 times higher (95% CI 4.71 to 17.60, p <  0.0001) for an obese compared with a non-obese patient.

Following PAO surgery, there was no difference in radiographic correction between obese and non-obese patients. PAO procedures in obese patients correct the deformity effectively but are associated with an increased rate of complications.

Cite this article: Bone Joint J 2015;97-B:29–34.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1455 - 1458
1 Nov 2014
Amanatullah DF Rachala SR Trousdale RT Sierra RJ

Dysplasia of the hip, hypotonia, osteopenia, ligamentous laxity, and mental retardation increase the complexity of performing and managing patients with Down syndrome who require total hip replacement (THR). We identified 14 patients (six males, eight females, 21 hips) with Down syndrome and degenerative disease of the hip who underwent THR, with a minimum follow-up of two years from 1969 to 2009. In seven patients, bilateral THRs were performed while the rest had unilateral THRs. The mean clinical follow-up was 5.8 years (standard deviation (sd) 4.7; 2 to 17). The mean Harris hip score was 37.9 points (sd 7.8) pre-operatively and increased to 89.2 (sd 12.3) at final follow-up (p = 1x10-9). No patient suffered a post-operative dislocation. In three patients, four hips had revision THR for aseptic loosening at a mean follow-up of 7.7 years (sd 6.3; 3 to 17). This rate of revision THR was higher than expected. Our patients with Down syndrome benefitted clinically from THR at mid-term follow-up.

Cite this article: Bone Joint J 2014;96-B:1455–8.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 21 - 25
1 Nov 2013
Perry KI Trousdale RT Sierra RJ

The treatment of hip dysplasia should be customised for patients individually based on radiographic findings, patient age, and the patient’s overall articular cartilage status. In many patients, restoration of hip anatomy as close to normal as possible with a PAO is the treatment of choice.

Cite this article: Bone Joint J 2013;95-B, Supple A:21–5.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1702 - 1702
1 Dec 2005
SIERRA RJ TROUSDALE RT CABANELA ME


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 21 - 24
1 Jan 2005
Sierra RJ Trousdale RT Cabanela ME

We surveyed 343 young women with 420 total hip arthroplasties (THAs) regarding pregnancy and childbirth after THA. The mean age at surgery was 35 years (18 to 45). The mean length of follow-up after the initial arthroplasty was 16 years (6 to 27).

Of these 343 women, 47 (13.7%) had a successful pregnancy after their primary THA. The first baby after a THA was delivered vaginally in 30 patients and by Caesarean section in 17. Of the 343 patients, 138 underwent a revision. For the entire series, the risk of revision at five years was 5%, at ten years 24%, and at 20 years 50%. After adjusting for age at surgical intervention, the risk of revision was not significantly associated with childbirth. Of the 47 patients who had a successful pregnancy, 28 (60%) noted an increase in pain in the hip during pregnancy and ten of these patients had persistent pain after their pregnancy. Seven patients complained of pain in the groin in the replaced hip after childbirth. At the time of this survey, five of these patients (70%) had had revision THA.

Childbirth is not affected by the presence of a THA. Pregnancy after THA is not associated with decreased survival of the prosthesis. Pain in the hip is common during pregnancy in these patients. Pain in the groin which persists after delivery commonly leads to revision of the THA.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 4 | Pages 532 - 534
1 May 2000
Crockarell JR Trousdale RT Guyton JL

The anterior centre-edge (VCA) angle quantifies the anterior cover of the femoral head, and angles of less than 20° are considered abnormal. We have measured the VCA angles in hips without osteoarthritic changes. We took bilateral false-profile radiographs of nine female and 30 male cadavers without signs of osteoarthritis. The mean age at the time of death was 72 years (46 to 92). The mean VCA angle was 32.8° (17.7 to 53.6). The SD was 7.9°.

Our findings suggest that the threshold of abnormality of the VCA angle may be slightly lower than previously thought. This information may be useful in counselling patients with asymptomatic acetabular dysplasia.