The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 385 - 389
1 Mar 2014
Attal R Maestri V Doshi HK Onder U Smekal V Blauth M Schmoelz W

Using human cadaver specimens, we investigated the role of supplementary fibular plating in the treatment of distal tibial fractures using an intramedullary nail. Fibular plating is thought to improve stability in these situations, but has been reported to have increased soft-tissue complications and to impair union of the fracture. We proposed that multidirectional locking screws provide adequate stability, making additional fibular plating unnecessary. A distal tibiofibular osteotomy model performed on matched fresh-frozen lower limb specimens was stabilised with reamed nails using conventional biplanar distal locking (CDL) or multidirectional distal locking (MDL) options with and without fibular plating. Rotational stiffness was assessed under a constant axial force of 150 N and a superimposed torque of ± 5 Nm. Total movement, and neutral zone and fracture gap movement were analysed.

In the CDL group, fibular plating improved stiffness at the tibial fracture site, albeit to a small degree (p = 0.013). In the MDL group additional fibular plating did not increase the stiffness. The MDL nail without fibular plating was significantly more stable than the CDL nail with an additional fibular plate (p = 0.008).

These findings suggest that additional fibular plating does not improve stability if a multidirectional distal locking intramedullary nail is used, and is therefore unnecessary if not needed to aid reduction.

Cite this article: Bone Joint J 2014;96-B:385–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 973 - 976
1 Jul 2009
Kralinger F Unger S Wambacher M Smekal V Schmoelz W

The medial periosteal hinge plays a key role in fractures of the head of the humerus, offering mechanical support during and after reduction and maintaining perfusion of the head by the vessels in the posteromedial periosteum. We have investigated the biomechanical properties of the medial periosteum in fractures of the proximal humerus using a standard model in 20 fresh-frozen cadaver specimens comparable in age, gender and bone mineral density. After creating the fracture, we displaced the humeral head medial or lateral to the shaft with controlled force until complete disruption of the posteromedial periosteum was recorded. As the quality of periosteum might be affected by age and bone quality, the results were correlated with the age and the local bone mineral density of the specimens measured with quantitative CT.

Periosteal rupture started at a mean displacement of 2.96 mm (sd 2.92) with a mean load of 100.9 N (sd 47.1). The mean maximum load of 111.4 N (sd 42.5) was reached at a mean displacement of 4.9 mm (sd 4.2). The periosteum was completely ruptured at a mean displacement of 34.4 mm (sd 11.1). There was no significant difference in the mean distance to complete rupture for medial (mean 35.8 mm (sd 13.8)) or lateral (mean 33.0 mm (sd 8.2)) displacement (p = 0.589).

The mean bone mineral density was 0.111 g/cm3 (sd 0.035). A statistically significant but low correlation between bone mineral density and the maximum load uptake (r = 0.475, p = 0.034) was observed.

This study showed that the posteromedial hinge is a mechanical structure capable of providing support for percutaneous reduction and stabilisation of a fracture by ligamentotaxis. Periosteal rupture started at a mean of about 3 mm and was completed by a mean displacement of just under 35 mm. The microvascular situation of the rupturing periosteum cannot be investigated with the current model.