header advert
Results 1 - 2 of 2
Results per page:
Bone & Joint Research
Vol. 11, Issue 4 | Pages 229 - 238
11 Apr 2022
Jaeger S Eissler M Schwarze M Schonhoff M Kretzer JP Bitsch RG

Aims

One of the main causes of tibial revision surgery for total knee arthroplasty is aseptic loosening. Therefore, stable fixation between the tibial component and the cement, and between the tibial component and the bone, is essential. A factor that could influence the implant stability is the implant design, with its different variations. In an existing implant system, the tibial component was modified by adding cement pockets. The aim of this experimental in vitro study was to investigate whether additional cement pockets on the underside of the tibial component could improve implant stability. The relative motion between implant and bone, the maximum pull-out force, the tibial cement mantle, and a possible path from the bone marrow to the metal-cement interface were determined.

Methods

A tibial component with (group S: Attune S+) and without (group A: Attune) additional cement pockets was implanted in 15 fresh-frozen human leg pairs. The relative motion was determined under dynamic loading (extension-flexion 20° to 50°, load-level 1,200 to 2,100 N) with subsequent determination of the maximum pull-out force. In addition, the cement mantle was analyzed radiologically for possible defects, the tibia base cement adhesion, and preoperative bone mineral density (BMD).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1356 - 1361
1 Oct 2012
Streit MR Walker T Bruckner T Merle C Kretzer JP Clarius M Aldinger PR Gotterbarm T

The Oxford mobile-bearing unicompartmental knee replacement (UKR) is an effective and safe treatment for osteoarthritis of the medial compartment. The results in the lateral compartment have been disappointing due to a high early rate of dislocation of the bearing. A series using a newly designed domed tibial component is reported.

The first 50 consecutive domed lateral Oxford UKRs in 50 patients with a mean follow-up of three years (2.0 to 4.3) were included. Clinical scores were obtained prospectively and Kaplan-Meier survival analysis was performed for different endpoints. Radiological variables related to the position and alignment of the components were measured.

One patient died and none was lost to follow-up. The cumulative incidence of dislocation was 6.2% (95% confidence interval (CI) 2.0 to 17.9) at three years. Survival using revision for any reason and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85 to 99) at three years, respectively. Outcome scores, visual analogue scale for pain and maximum knee flexion showed a significant improvement (p < 0.001). The mean Oxford knee score was 43 (sd 5.3), the mean Objective American Knee Society score was 91 (sd 13.9) and the mean Functional American Knee Society score was 90 (sd 17.5). The mean maximum flexion was 127° (90° to 145°). Significant elevation of the lateral joint line as measured by the proximal tibial varus angle (p = 0.04) was evident in the dislocation group when compared with the non-dislocation group.

Clinical results are excellent and short-term survival has improved when compared with earlier series. The risk of dislocation remains higher using a mobile-bearing UKR in the lateral compartment when compared with the medial compartment. Patients should be informed about this complication. To avoid dislocations, care must be taken not to elevate the lateral joint line.