Bone & Joint Research
Vol. 7, Issue 7 | Pages 485 - 493
1 Jul 2018
Numata Y Kaneuji A Kerboull L Takahashi E Ichiseki T Fukui K Tsujioka J Kawahara N

Objective

Cement thickness of at least 2 mm is generally associated with more favorable results for the femoral component in cemented hip arthroplasty. However, French-designed stems have shown favorable outcomes even with thin cement mantle. The biomechanical behaviors of a French stem, Charnley-Marcel-Kerboull (CMK) and cement were researched in this study.

Methods

Six polished CMK stems were implanted into a composite femur, and one million times dynamic loading tests were performed. Stem subsidence and the compressive force at the bone-cement interface were measured. Tantalum ball (ball) migration in the cement was analyzed by micro CT


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 342 - 348
1 Mar 2010
El Masri F Kerboull L Kerboull M Courpied JP Hamadouche M

We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component.

At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 304 - 309
1 Mar 2009
Kerboull L Hamadouche M Kerboull M

We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30.

There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 4 | Pages 658 - 663
1 Jul 1990
Sedel L Kerboull L Christel P Meunier A Witvoet J

Total hip replacement using an alumina head and socket and a titanium alloy stem is evaluated in a series of patients under 50 years of age. Between April 1977 and December 1986, 86 such replacements were performed in 75 patients, but mainly because patients had difficulty travelling from Africa, only 71 hips were followed up adequately; of these, 56 were primary procedures and 15 revisions. Survivorship analysis showed that 98% of the prostheses were retained for 10 years. On clinical and radiological examination 51 of the 71 hips were stable and acceptable, 15 had radiological changes on the acetabular side, and one on the femoral side; four other cases had clinical and radiological changes suggesting impending failure, possibly because fixation of the socket was inadequate. There were no differences between the results of the primary procedures and those of revisions. In these young patients, the results seem better with alumina-on-alumina hips than with other varieties, possibly because of their remarkably low wear.