header advert
Results 1 - 2 of 2
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1468 - 1474
1 Nov 2012
Hill JC Archbold HAP Diamond OJ Orr JF Jaramaz B Beverland DE

Restoration of leg length and offset is an important goal in total hip replacement. This paper reports a calliper-based technique to help achieve these goals by restoring the location of the centre of the femoral head. This was validated first by using a co-ordinate measuring machine to see how closely the calliper technique could record and restore the centre of the femoral head when simulating hip replacement on Sawbone femur, and secondly by using CT in patients undergoing hip replacement.

Results from the co-ordinate measuring machine showed that the centre of the femoral head was predicted by the calliper to within 4.3 mm for offset (mean 1.6 (95% confidence interval (CI) 0.4 to 2.8)) and 2.4 mm for vertical height (mean -0.6 (95% CI -1.4 to 0.2)). The CT scans showed that offset and vertical height were restored to within 8 mm (mean -1 (95% CI -2.1 to 0.6)) and -14 mm (mean 4 (95% CI 1.8 to 4.3)), respectively.

Accurate assessment and restoration of the centre of the femoral head is feasible with a calliper. It is quick, inexpensive, simple to use and can be applied to any design of femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 308 - 313
1 Mar 2008
Murtha PE Hafez MA Jaramaz B DiGioia AM

Three-dimensional surface models of the normal hemipelvis derived from volumetric CT data on 42 patients were used to determine the radius, depth and orientation of the native acetabulum. A sphere fitted to the lunate surface and a plane matched to the acetabular rim were used to calculate the radius, depth and anatomical orientation of the acetabulum. For the 22 females the mean acetabular abduction, anteversion, radius and normalised depth were 57.1° (50.7° to 66.8°), 24.1° (14.0° to 33.3°), 25 mm (21.7 to 30.3) and 0.79 mm (0.56 to 1.04), respectively. The same parameters for the 20 males were 55.5° (47.7° to 65.9°), 19.3° (8.5° to 32.3°), 26.7 mm (24.5 to 28.7) and 0.85 mm (0.65 to 0.99), respectively.

The orientation of the native acetabulum did not match the safe zone for acetabular component placement described by Lewinnek. During total hip replacement surgeons should be aware that the average abduction angle of the native acetabulum exceeds that of the safe zone angle. If the concept of the safe zone angle is followed, abduction of the acetabular component should be less than the abduction of the native acetabulum by approximately 10°.