header advert
Results 1 - 2 of 2
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1567 - 1573
1 Dec 2007
Kolling C Simmen BR Labek G Goldhahn J

Since the introduction of the first National Arthroplasty Register in Sweden in 1975, many other countries have tried to adopt the successful Scandinavian system. However, not all have overcome the political and practical difficulties of establishing a working register. We have surveyed the current registries to establish the key factors required for an effective database. We have received detailed information from 15 arthroplasty registers worldwide. The legal conditions under which they operate together with the methods of collection and handling of the data differ widely, but the fulfilment of certain criteria is necessary achieve a high degree of completeness of the data to ensure the provision of statistically relevant information.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 426 - 432
1 Mar 2005
Mueller CA Eingartner C Schreitmueller E Rupp S Goldhahn J Schuler F Weise K Pfister U Suedkamp NP

The treatment of fractures of the proximal tibia is complex and makes great demands on the implants used. Our study aimed to identify what levels of primary stability could be achieved with various forms of osteosynthesis in the treatment of diaphyseal fractures of the proximal tibia. Pairs of human tibiae were investigated. An unstable fracture was simulated by creating a defect at the metaphyseal-diaphyseal junction. Six implants were tested in a uniaxial testing device (Instron) using the quasi-static and displacement-controlled modes and the force-displacement curve was recorded. The movements of each fragment and of the implant were recorded video-optically (MacReflex, Qualysis). Axial deviations were evaluated at 300 N.

The results show that the nailing systems tolerated the highest forces. The lowest axial deviations in varus and valgus were also found for the nailing systems; the highest axial deviations were recorded for the buttress plate and the less invasive stabilising system (LISS). In terms of rotational displacement the LISS was better than the buttress plate.

In summary, it was found that higher loads were better tolerated by centrally placed load carriers than by eccentrically placed ones. In the case of the latter, it appears advantageous to use additive procedures for medial buttressing in the early phase.