Vol. 99-B, Issue 6 | Pages 708 - 713
1 Jun 2017
Rushton PRP Siddique I Crawford R Birch N Gibson MJ Hutton MJ

The MAGnetic Expansion Control (MAGEC) system is used increasingly in the management of early-onset scoliosis. Good results have been published, but there have been recent reports identifying implant failures that may be associated with significant metallosis surrounding the implants. This article aims to present the current knowledge regarding the performance of this implant, and the potential implications and strategies that may be employed to identify and limit any problems.

We urge surgeons to apply caution to patient and construct selection; engage in prospective patient registration using a spine registry; ensure close clinical monitoring until growth has ceased; and send all explanted MAGEC rods for independent analysis.

The MAGEC system may be a good instrumentation system for the treatment of early-onset scoliosis. However, it is innovative and like all new technology, especially when deployed in a paediatric population, robust systems to assess long-term outcome are required to ensure that patient safety is maintained.

Cite this article: Bone Joint J 2017;99-B:708–13.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 5 | Pages 830 - 834
1 Sep 1999
Khaw FM Worthy SA Gibson MJ Gholkar A

We studied MR images of the spine in a consecutive series of 100 patients with acute compression of the spinal cord due to metastases. All patients had documented neurological deficit and histologically proven carcinoma. MRI was used to localise bony metastatic involvement and soft-tissue impingement of the cord. A systematic method of documenting metastatic involvement is described.

A total of 43 patients had compression at multiple levels; 160 vertebral levels were studied. In 120 vertebrae (75%), anterior, lateral and posterior bony elements were involved. Soft-tissue impingement of the spinal cord often involved more than one quadrant of its circumference. In 69 vertebrae (43%) there was concomitant anterior and posterior compression. Isolated involvement of a vertebral body was observed in only six vertebrae (3.8%).

We have shown that in most cases of acute compression of the spinal cord due to metastases there is coexisting involvement of both anterior and posterior structures.