header advert
Results 1 - 2 of 2
Results per page:
The Bone & Joint Journal
Vol. 101-B, Issue 1 | Pages 47 - 54
1 Jan 2019
Clough T Bodo K Majeed H Davenport J Karski M

Aims

We report the long-term clinical and radiological outcomes of a consecutive series of 200 total ankle arthroplasties (TAAs, 184 patients) at a single centre using the Scandinavian Total Ankle Replacement (STAR) implants.

Patients and Methods

Between November 1993 and February 2000, 200 consecutive STAR prostheses were implanted in 184 patients by a single surgeon. Demographic and clinical data were collected prospectively and the last available status was recorded for further survival analysis. All surviving patients underwent regular clinical and radiological review. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot scoring system. The principal endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan–Meier survival curves were generated with 95% confidence intervals and the rate of failure calculated for each year.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 2 | Pages 259 - 265
1 Mar 1993
Dall D Learmonth I Solomon M Miles A Davenport J

We report the results of a 4- to 17-year clinical and radiological follow-up of 264 Charnley first-generation stems in comparison with those of 402 second- and subsequent-generation stems. The incidence of fracture was 4.1% in first-generation stems and 0.5% in second- and subsequent-generation stems. The incidence of stem loosening requiring or likely to require revision was 3.1% in first-generation and 11.4% in second-generation stems. We believe that the increased loosening rate in second- and subsequent-generation stems is due to their larger cross-sectional area, which produces an increase in flexural stiffness.