The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 1 | Pages 113 - 115
1 Jan 1991
Willett K

In the light of EEC proposals on the avoidance of damage to hearing caused by noise, a study was undertaken to determine the risk posed by powered orthopaedic instruments. The noise levels from a number of air-powered and electric tools were measured and analysed and found to exceed the recommended levels. The predicted daily personal noise exposure was calculated and the potential for hearing damage confirmed. Twenty-seven senior orthopaedic staff were then assessed by audiometry; evidence of noise-induced hearing loss was found in half the subjects. The increasing use of powered instruments in elective orthopaedics and fracture fixation may present a significant cumulative risk to the hearing of orthopaedic surgeons and theatre personnel. The use of ear defenders should be promoted, and manufacturers should be encouraged to develop instruments with lower noise emission levels.


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 4 | Pages 607 - 610
1 Aug 1988
Willett K Simmons C Bentley G

A prospective study of 120 consecutive total hip replacements showed that deep suction drains produced maximal drainage volumes in the first 24 hours. Their continued presence resulted in minimal further drainage, did not reduce the likelihood of haematoma formation and led in some cases to the spread of skin organisms into the wound.