The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 938 - 942
1 Nov 1997
Cool WP Carter SR Grimer RJ Tillman RM Walker PS

We report our results in 24 children with malignant primary bone tumours of the distal femur treated with a Stanmore extendible endoprosthesis (SEER). This consists of a femoral component that can be lengthened, a constrained knee and an uncemented sliding tibial component which crosses the proximal tibial physeal plate perpendicularly.

The average age of the patients at diagnosis was ten years and the mean follow-up was 4.7 years (2.5 to 7.9). The mean growth of the affected tibia was 76% (18 to 136) and of the fibula 83% (15 to 750) of the growth of the unaffected limb.

Measurement of growth arrest lines showed that the mean growth of the proximal tibial physis on the affected side was 69% (43 to 100) of that of the normal side. The great variability in the growth of the physis cannot yet be explained.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 5 - 13
1 Jan 1996
Unwin PS Cannon SR Grimer RJ Kemp HBS Sneath RS Walker PS

We have made a retrospective study of 1001 custom-made prostheses used as replacements after surgery for bone tumours. There were 493 distal femoral, 263 proximal femoral and 245 proximal tibial prostheses. Aseptic loosening was shown to be the principal mode of failure of the implants, and 71 patients had revision for aseptic loosening of a cemented intramedullary stem.

The probability of a patient surviving aseptic loosening for 120 months was 93.8% for a proximal femoral replacement, 67.4% for a distal femoral prosthesis and 58% for a proximal tibial implant. In patients with distal femoral replacements the age of the patient at the time of operation and the percentage of bone resected were related to the risk of aseptic loosening. Young patients with distal femoral prostheses in whom a high percentage of the femur had been replaced had the poorest prognosis for survival without aseptic loosening. The percentage of bone removed had a significant effect in the proximal tibial replacement group, but the age of the patient did not. By contrast, neither the age nor the percentage of bone removed was a factor after proximal femoral replacement.

The significance of these findings is discussed in relation to mechanical factors.