Trail IA Nuttall D Stanley JK

We undertook a radiological analysis of 186 standard Souter implants to determine survivorship and to analyse the pattern of failure in those needing revision. The implants had been inserted as a primary procedure in patients with rheumatoid arthritis of the elbow at our hospital over the last 12 years.

Taking revision as an endpoint, the survivorship after 12 years was 87%. If, however, revision and loosening, defined as the Hindex value equivalent to demarcation of 1 mm around the whole implant, are also included, the survivorship falls to 80%. Of the 24 implants revised, 18 (75%) were for problems with the humeral component, three (12.5%) with the ulnar component and three (12.5%) for instability.

Loosening of the humeral component occurred when the implant extended into the humerus, with the tip moving anteriorly on to the anterior humeral cortex. Our study indicates that loosening can be predicted by the rate of change in this angle of extension of the prosthesis.