header advert
Results 1 - 4 of 4
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_14 | Pages 66 - 66
1 Nov 2018
Summer B Schwarzenlander K Reyna AP Thomas P Kretzer P Vas A Grupp T
Full Access

Total knee arthroplasty is a well-established treatment for degenerative joint disease, on the other hand metal ion release of cobalt or chromium and particle formation can trigger intolerance reactions. Biotribological examinations can help to assess the metal ion release in different settings. The purpose of this study was the evaluation of inter-laboratory differences in the metal ion concentration analysis. Samples were generated in a 3+1 station knee wear simulator (EndoLab GmbH, Thansau, Germany) with a medium size Columbus Knee System with or without AS multilayer coating. The wear simulation was performed under highly demanding activity (HDA) profile and samples were taken after 0.5, 5.0, 5.5. and 8.0 million cycles. The samples were blinded and sent to three different laboratories and the content of chromium, cobalt, molybdenum, nickel, and zirconium was assessed by inductively coupled plasma mass spectrometry (ICP-MS). The AS multilayer coating clearly reduced the release of chromium, cobalt and molybdenum. Mean levels were: Chromium 9329.78µg/l ± 985.44 vs 503.75µg/l ± 54.19, cobalt 10419.00µg/l ± 15.517.53 vs 2.60µg/l ± 1.35, molybdenum 2496.33µg/l ± 102.62 vs 2.46µg/l ± 2.31. Interestingly we found especially for nickel and zirconium big inter-laboratory differences in the metal assessment. There were up to 10-fold higher values in comparison of one laboratory to another. The data demonstrate that results of metal ion assessment should be evaluated by interlaboratory comparison and should be critically interpreted.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 48 - 48
1 Apr 2018
Reyna ALP Fritz B Schwiesau J Summer B Thomas P Grupp TM
Full Access

Total knee arthroplasty is a well established treatment for degenerative joint disease with good clinical results. However, complications may occur due to a biological response to polyethylene wear particles, leading to osteolysis and aseptic loosening, as well as local and systemic hypersensitivity reactions triggered by metal ions and particles such as chromium, cobalt and molybdenum. Moreover, there is an increasing demand on the performance of these implants, as this treatment is also performed in heavier, younger and middle-aged adults who have a significant physical activity and higher life expectancy. The purpose of the following study was to compare the wear characteristics and performance of a zirconium nitride (ZrN) coated knee implant, designed for patients with metal ion hypersensitivity, against the clinically established cobalt-chromium (CoCr) version under a high demanding activities wear simulation.

Medium size AS Columbus® DD (Aesculap AG, Tuttlingen, Germany) femoral and tibial components with a ZrN surface were tested in comparison with the cobalt-chromium version Columbus® DD. For both groups, ultra-high-molecular weight polyethylene (UHMWPE) gliding surfaces (size T3, high 10 mm) were used. Wear simulation was performed on a load controlled 4 station knee wear simulator (EndoLab GmbH, Thansau, Germany) capable of reproducing loads and movement of daily activities measured in vivo (Bergmann et al, 2014) on 8 patients and normalized to a patient weight of 100 kg (Schwiesau et al, 2014). The load profiles were applied for 5 million cycles in a combination of 40% stairs up, 40% stairs down, 10% level walking, 8% chair raising and 2% deep squatting. Test serum was changed every 0.5 million cycles and all the components were cleaned and analyzed according to ISO 14243-2:2009(E). The gliding surfaces were evaluated for gravimetric wear and wear patterns, femur components analyzed for scratches and the test medium analyzed for metal ion concentration (cobalt, chromium, molybdenum and zirconium) using ICP-MS according to ISO 17294-2.

The present study showed a wear rate reduction for the ZrN group (1.01 ± 0.29 mg/million) in comparison with the CoCr group (2.40 ± 1.18 mg/million cycles). The articulation surface of the ZrN coated femurs remained polished after the testing period, whereas the uncoated femurs showed wear scratches. Furthermore, the metal ion release from the ZrN coated implants was reduced orders of magnitude in comparison with the CoCr implants through the entire test. These results demonstrate the efficiency of ZrN coated knee implants to reduce wear as well as to prevent metal ion release in the knee joint.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 66 - 67
1 Jan 2011
Kwon Y Thomas P Summer B McLardy-Smith P Ostlere S Gundle R Whitwell D Gibbons C Athanasou N Gill H Murray D
Full Access

Introduction: Symptomatic abnormal soft-tissue masses relating to the hip joint, such as those described as pseudotumours, are being increasingly reported following metal-on-metal hip resurfacing arthroplasty (MoMHRA). These were found to be locally destructive, requiring revision surgery in a high proportion (75%) of patients. Lymphocyte infiltrations seen in pseudotumours were similar to aseptic lymphocyte vascular associated lesion (ALVAL), which is thought to represent a T-lymphocyte-mediated delayed type hypersensitivity. Therefore, a delayed hypersensitivity reaction to nickel (Ni), chromium (Cr) or cobalt (Co) has been suggested to play a role in pseudotumour aetiology. In patients with bilateral MoMHRA who presented with symptoms on one side, subsequent scans have demonstrated pseudotumours both on the symptomatic and asymptomatic side. Thus, there are concerns that there may be an appreciable number of asymptomatic pseudotumours that surgeons are unaware of and these may eventually become symptomatic.

Aim: The aims of this study were:

to determine the prevalence of asymptomatic pseudotumours after MoMHRA; and

to measure Co and Cr ion levels as well as lymphocyte proliferation responses to Ni, Co and Cr (the principal elements in the CoCr alloy used in MoMHRA) in MoMHRA patients with and without asymptomatic pseudotumours.

Methods: A total of 201 MoMHRA implanted hips in 158 patients (97 male, 61 female) with a mean age of 56 years (range 33–73 years) were evaluated. The mean follow-up was 61 months (range 13–88 months). Resurfacing devices implanted included 128 Birmingham Hip Resurfacing, 66 Conserve Plus and seven ReCap. The control groups included additional 20 patients, 10 male and 10 female (a mean age 68 years, range 57–80 years) with metal-on-polyethylene total hip arthroplasty and a further 22 age-matched patients (a mean age 55 years) without any metal implants. Ultrasound was used as the initial imaging modality and MRI was used to assess the extent of the identified masses. Patients with a soft-tissue mass had ultrasound-guided aspiration or core biopsy performed. Venous blood samples were collected in all patients for serum cobalt and chromium ion levels analysis using Inductively-Coupled Plasma Mass Spectrometer and lymphocyte transformation tests (LTT). The Oxford Hip Score (OHS) was used to measure the functional outcomes of patients. Acetabular component abduction angle was measured from standardised anteroposterior pelvis radiographs.

Results: Prevalence – Pseudotumours were found in 7 patients (6 female and 1 male). The overall prevalence of asymptomatic pseudotumours was 4%, with a relatively very high (30%) prevalence in females with bilateral implants. Histological examinations showed extensive necrosis of connective tissue, in which there were scattered aggregates of metal particles and a diffuse lymphocyte infiltrate.

Metal Ion Levels – The presence of pseudotumour was associated with significantly higher median serum cobalt levels (9.2mg/L vs. 1.9mg/L, p< 0.001), chromium levels (12.0mg/L vs. 2.1mg/L, p< 0.001), hip aspirate cobalt levels (1182 mg/L vs. 86.2mg/L, p=0.003), and aspirate chromium levels (883mg/L vs. 114.8mg/ L, p=0.006), as well as with inferior functional scores (OHS 41 vs. 47 p< 0.001). There was no significant difference in acetabular cup inclination angle (p=0.51). Lymphocyte Reactivity: A higher incidence and level of enhanced lymphocyte reactivity to Ni (p=0.001), but not to Co or Cr (the principal elements in the CoCr alloy used in metal-on-metal hip resurfacing implants), was found in patients with MoMHRA compared to the patients without MoM implants. However, lymphocyte reactivity to Co, Cr and Ni did not significantly differ in patients with pseudotumours compared to those patients without pseudotumours.

Conclusion: The prevalence of asymptomatic pseudotumours in females was high, especially in females with bilateral MoMHRA implants (30%). The patients with ‘asymptomatic’ pseudotumours were in fact mildly symptomatic. Lymphocyte reactivity to Co, Cr and Ni did not differ in patients with pseudotumour compared to those patients without pseudotumours, suggesting that systemic hypersensitivity type IV reactions, mediated by lymphocyte reactivity to these metals, is not the dominant mechanism in pathogenesis of the soft tissue pseudotumours. Furthermore, pseudotumours were not detected in those patients who had normal levels of cobalt and chromium ions. This suggests that pseudotumours do not occur if MoM articulations are well functioning. Therefore, pseudotumours are likely to be a biological consequence of the large amount of metal debris generated in vivo due to excessive wear.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_IV | Pages 523 - 523
1 Oct 2010
Kwon Y Gill H Mclardy-Smith P Monk P Murray D Ostlere S Summer B Thomas P
Full Access

Recently, a series of locally destructive soft tissue pseudotumour has been reported in patients following metal-on-metal hip resurfacing arthroplasty (MoMHRA), requiring revision surgery in a high percentage of patients. Based on the histological evidence of lymphocytic infiltration, a delayed hypersensitivity reaction to nickel (Ni), chromium (Cr) or cobalt (Co) has been suggested to play a role in its aetiology. The aim of this study was to investigate the incidence and level of hypersensitivity reaction to metals in patients with pseudotumour.

Materials and Methods: 25 patients were investigated in this Ethics approved study:

Group 1: MoMHRA patients with pseudotumours, detected on the ultrasound and confirmed with MRI (n=6, 5 F:1 M, mean age 53 years);

Group 2: MoMHRA patients without pseudotumours (n=13, 7 F:6 M, mean age 55 years); and

Group 3: age-matched control subjects without metal implants (n=6, 4 F:2 M, mean age 54 years).

Lymphocyte transformation tests (LTT) were used to measure lymphocyte proliferation responses to metals. Peripheral blood mononuclear cells were isolated from heparinized blood samples using standard Ficoll–Hypaque® (Pharmacia). The PBMC were cultured at a cell density of 106 cells/mL. Culture was set up in the presence of either:

medium alone;

nickel chloride (Sigma; 10-4M-10-6M);

cobalt chloride (10-4M-10-6M); and

chromium chloride (10-4M-10-6M).

After 5 days of culture, cells were pulsed with [3H]-thymidine and proliferation was assessed by scintillation counting. The stimulation index (SI) was calculated by the ratio of mean counts per minute of stimulated to unstimulated cultures. A SI value of greater than 2.0 was interpreted as a positive result.

Results: A clinical history of metal allergy was reported in 2/6 in Group 1, 2/13 in Group 2, and none in Group 3. In pseudotumour group, the incidence of reactivity to Ni, Co and Cr was 60%, 17% and 0%, respectively. Within Group 2, the reactivity to Ni, Co and Cr was 69%, 8% and 15%, respectively. One control subject had reactivity to Ni. Inter-group comparisons of mean SI values (Kruskal-Wallis non-parametric analysis of variance) showed no significant differences (p> 0.05).

Discussion: The incidence of enhanced lymphocyte response to metals in patients with MoMHRA was more common than the control group. However, in comparison with non-pseudotumour patients, there was no significant difference in the incidence or the level of lymphocyte reactivity in patients with pseudotumour. We conclude that patients with MoMHRA have an enhanced lymphocyte response to metal ions, reflecting exposure and immune reactivity. However, patients with pseudotumours have a similar proliferative response to those without pseudotumours, which suggests that type IV hypersensitivity may not be the cause of the pseudotumours.