header advert
Results 21 - 40 of 64
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 262 - 262
1 Dec 2013
Noble P Chan N Matthies A Hart A Skinner JA
Full Access

Introduction

Recent retrieval studies and registry reports have demonstrated an alarming incidence of early failure of metal-on-metal THR. This appears to be due to fretting and corrosion at the taper junction (trunnion) between the neck and large diameter heads in metal-on-metal hip implants. It has been proposed that designs with lower bearing clearances and greater cup flexibility deform during implantation leading to increased frictional torque and micromotion at the head-neck taper junction. Small movements at the trunnion may suggest elastic deformation, but large movements may suggest slippage at the friction interface. This study was conducted using retrieved metal-on-metal components to test the hypotheses that: 1. Cup deformation through localized compression leads to increased bearing torque, and 2. Increased torques generated in large head metal-on-metal bearings cause motion of the head-neck taper junction.

Materials and Methods

Nine metal-on-metal hip implants were received from a national joint retrieval service and tested in a mechanical testing machine. The components were of three different designs (ASR, BHR, and Durom) and ranged in diameter from 42–54 mm. A custom jig was constructed to generate controlled radial compression at opposite points on the rim of an acetabular component. The jig was positioned inverted to the normal anatomical position and was angled to simulate the anatomical orientation of the cup (35° inclination, 10° anteversion). With the exception of an initial compression load of 100N, the cups were compressed at 200N intervals to a maximum of 2000N. Three trials at each cup compression load were performed. The torque developed about the trunnion axis was measured as the head articulated through a motion arc of 60° and the friction factor was calculated. Head–neck micromotion was continuously monitored using a non-displacement inductive transducer. Changes in micromotion from the 100N compression load were calculated.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 353 - 353
1 Dec 2013
Hothi H Berber R Whittaker R Skinner JA Hart A
Full Access

Metal-on-metal (MOM) total hip arthroplasty and hip resurfacing using large diameter femoral heads offer clinical advantages, however the failure rates of these hips are unacceptably high. Retrieved MOM hips have a wide range of wear rates of their bearing surfaces and there is ongoing research to identify the causes of failure.

Detailed visual inspection is the first step in the forensic examination of failed hip components, but there is no universally accepted description or process. Visible features may help explain the mechanism of failure of MOM hips.

During our analysis of 2000 MOM hip components, we have developed protocols to undertake detailed, non-destructive macroscopic and stereomicroscopic examinations of each component, using quantitative assessment to document the presence of types of damage.

We have established a systematic terminology to describe the types of damage that are observed, allowing for consistency and clarity in the vocabulary used. These include (but are not limited to):

Scratching – when there is an increase in the number and/or magnitude of scratches present on the surface, typically increasing measured roughness parameters. The severity of scratching is determined by rubbing a 0.18 mm thickness acetate gage or fingernail over the surface:

Light –visible but not detectable with gage.

Moderate – visible and detectable with gage.

Heavy – depth clearly visible and will catch a fingernail or gage.

Pitting – indentations in the surface for which the dimensions are similar in all directions.

Embedded Particles – hard particles that have become embedded in the bearing surface.

Discolouration – observed as a change in the appearance of the surface, often as colour diffraction patterns.

Polishing, gouges, etching (from corrosion), surface films, surface deposits and tribochemical reaction layers.

To assess the distribution of these types of damage on the components, each surface is considered in terms of zones defined by quadrants (cup, head, stem and taper) and sub-quadrants (cup and head only), Figure 1. Each zone is scored on a scale of 0 to 3 by determining the percentage of the surface area of the quadrant that exhibits the feature in question:

a score of 0 indicates 0%

indicates greater than 0% but less than 25%

indicates between 25% and 75% inclusive

indicates greater than 75%.

The use of zones to differentiate between the polar and equatorial regions of the cup and head surfaces can offer important information relating to the articulation of the bearing surfaces. For example in a cohort of 90 inspections, a score of 3 for light scratching was recorded in all areas of the bearing surface in 40% of cups, whilst approximately 5% had evidence of pitting, occurring near the rim.

Current research at our retrieval centre involves correlating the results of detailed inspections with a range of variables, including implant design, size and surgical position.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1011 - 1021
1 Aug 2013
Krishnan H Krishnan SP Blunn G Skinner JA Hart AJ

Following the recall of modular neck hip stems in July 2012, research into femoral modularity will intensify over the next few years. This review aims to provide surgeons with an up-to-date summary of the clinically relevant evidence. The development of femoral modularity, and a classification system, is described. The theoretical rationale for modularity is summarised and the clinical outcomes are explored. The review also examines the clinically relevant problems reported following the use of femoral stems with a modular neck.

Joint replacement registries in the United Kingdom and Australia have provided data on the failure rates of modular devices but cannot identify the mechanism of failure. This information is needed to determine whether modular neck femoral stems will be used in the future, and how we should monitor patients who already have them implanted.

Cite this article: Bone Joint J 2013;95-B:1011–21.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 250 - 253
1 Feb 2013
Jalgaonkar A Dawson-Bowling SJ Mohan AT Spiegelberg B Saifuddin A Pollock R Skinner JA Briggs TWR Aston W

Local recurrence along the biopsy track is a known complication of percutaneous needle biopsy of malignant musculoskeletal tumours. In order to completely excise the track with the tumour its identification is essential, but this becomes increasingly difficult over time. In an initial prospective study, 22 of 45 patients (48.8%) identified over a three-month period, treated by resection of a musculoskeletal tumour, had an unidentifiable biopsy site at operation, with identification statistically more difficult after 50 days. We therefore introduced the practice of marking the biopsy site with India ink. In all 55 patients undergoing this procedure, the biopsy track was identified pre-operatively (100%); this difference was statistically significant. We recommend this technique as a safe, easy and accurate means of ensuring adequate excision of the biopsy track.

Cite this article: Bone Joint J 2013;95-B:250–3.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1377 - 1381
1 Oct 2012
Jaiswal PK Bentley G Carrington RWJ Skinner JA Briggs TWR

We analysed whether a high body mass index (BMI) had a deleterious effect on outcome following autologous chondrocyte implantation (ACI) or matrix-carried autologous chondrocyte implantation (MACI) for the treatment of full-thickness chondral defects of the knee from a subset of patients enrolled in the ACI vs MACI trial at The Royal National Orthopaedic Hospital.

The mean Modified Cincinnati scores (MCS) were significantly higher (p < 0.001) post-operatively in patients who had an ideal body weight (n = 53; 20 to 24.9 kg/m2) than in overweight (n = 63; 25 to 30 kg/m2) and obese patients (n = 22; > 30 kg/m2). At a follow-up of two years, obese patients demonstrated no sustained improvement in the MCS. Patients with an ideal weight experienced significant improvements as early as six months after surgery (p = 0.007). In total, 82% of patients (31 of 38) in the ideal group had a good or excellent result, compared with 49% (22 of 45) of the overweight and 5.5% (one of 18) in the obese group (p < 0.001). There was a significant negative relationship between BMI and the MCS 24 months after surgery (r = -0.4, p = 0.001).

This study demonstrates that obese patients have worse knee function before surgery and experience no sustained benefit from ACI or MACI at two years after surgery. There was a correlation between increasing BMI and a lower MCS according to a linear regression analysis. On the basis of our findings patient selection can be more appropriately targeted.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVI | Pages 11 - 11
1 Apr 2012
Muirhead-Allwood S Sandiford NA Skinner JA Hua J Walker PS
Full Access

We present the medium to long term clinical results of 112 Uncemented custom Computer Assisted Design Computer Assisted Manufactured (CAD-CAM) total hip arthroplasties performed between 1992 and 1998 in 111 patients. Fifty three males and 58 females were included. Mean age was 46.2 years (range 24.6yrs - 62.2 yrs). Average duration of follow up was 156 months (120 – 204 months). The mean Harris Hip Score (HHS) improved from 42.4 to 90.3, mean Oxford Hip Score (OHS) improved from 43.1 to 18.2 and the mean WOMAC hip score improved from 57.0 to 11.9. There was 1 revision due to failure of the acetabular components but there were no failures of the femoral components. There were no revisions for aseptic loosening. The worst case survival in this cohort of custom femoral components at an average 13 year follow up (range 10-17 years) was 98.2% (95% Confidence interval 95% to 99%). Survival of the femoral component alone was 100%.

These results are comparable with the best medium to long term results for femoral components used in primary total hip arthroplasty (THA) with any means of fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 504 - 509
1 Apr 2012
Bentley G Biant LC Vijayan S Macmull S Skinner JA Carrington RWJ

Autologous chondrocyte implantation (ACI) and mosaicplasty are methods of treating symptomatic articular cartilage defects in the knee. This study represents the first long-term randomised comparison of the two techniques in 100 patients at a minimum follow-up of ten years. The mean age of the patients at the time of surgery was 31.3 years (16 to 49); the mean duration of symptoms pre-operatively was 7.2 years (9 months to 20 years). The lesions were large with the mean size for the ACI group being 440.9 mm2 (100 to 1050) and the mosaicplasty group being 399.6 mm2 (100 to 2000). Patients had a mean of 1.5 previous operations (0 to 4) to the articular cartilage defect. Patients were assessed using the modified Cincinnati knee score and the Stanmore-Bentley Functional Rating system. The number of patients whose repair had failed at ten years was ten of 58 (17%) in the ACI group and 23 of 42 (55%) in the mosaicplasty group (p < 0.001).

The functional outcome of those patients with a surviving graft was significantly better in patients who underwent ACI compared with mosaicplasty (p = 0.02).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 488 - 492
1 Apr 2012
Vijayan S Bartlett W Bentley G Carrington RWJ Skinner JA Pollock RC Alorjani M Briggs TWR

Matrix-induced autologous chondrocyte implantation (MACI) is an established technique used to treat osteochondral lesions in the knee. For larger osteochondral lesions (> 5 cm2) deeper than approximately 8 mm we have combined the use of two MACI membranes with impaction grafting of the subchondral bone. We report our results of 14 patients who underwent the ‘bilayer collagen membrane’ technique (BCMT) with a mean follow-up of 5.2 years (2 to 8). There were 12 men and two women with a mean age of 23.6 years (16 to 40). The mean size of the defect was 7.2 cm2 (5.2 to 12 cm2) and were located on the medial (ten) or lateral (four) femoral condyles. The mean modified Cincinnati knee score improved from 45.1 (22 to 70) pre-operatively to 82.8 (34 to 98) at the most recent review (p < 0.05). The visual analogue pain score improved from 7.3 (4 to 10) to 1.7 (0 to 6) (p < 0.05). Twelve patients were considered to have a good or excellent clinical outcome. One graft failed at six years.

The BCMT resulted in excellent functional results and durable repair of large and deep osteochondral lesions without a high incidence of graft-related complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 570 - 573
1 Apr 2012
Whittingham-Jones P Mann B Coward P Hart AJ Skinner JA

Fracture of a ceramic component in total hip replacement is a rare but potentially catastrophic complication. The incidence is likely to increase as the use of ceramics becomes more widespread. We describe such a case, which illustrates how inadequate initial management will lead to further morbidity and require additional surgery. We present the case as a warning that fracture of a ceramic component should be revised to another ceramic-on-ceramic articulation in order to minimise the risk of further catastrophic wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 425 - 430
1 Mar 2012
Picardo NE Blunn GW Shekkeris AS Meswania J Aston WJ Pollock RC Skinner JA Cannon SR Briggs TW

In skeletally immature patients, resection of bone tumours and reconstruction of the lower limb often results in leg-length discrepancy. The Stanmore non-invasive extendible endoprosthesis, which uses electromagnetic induction, allows post-operative lengthening without anaesthesia. Between 2002 and 2009, 55 children with a mean age of 11.4 years (5 to 16) underwent reconstruction with this prosthesis; ten patients (18.2%) died of disseminated disease and one child underwent amputation due to infection. We reviewed 44 patients after a mean follow-up of 41.2 months (22 to 104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to 30) and the Toronto Extremity Salvage score was 92.3% (55.2% to 99.0%). There was no local recurrence of tumour. Complications developed in 16 patients (29.1%) and ten (18.2%) underwent revision.

The mean length gained per patient was 38.6 mm (3.5 to 161.5), requiring a mean of 11.3 extensions (1 to 40), and ten component exchanges were performed in nine patients (16.4%) after attaining the maximum lengthening capacity of the implant. There were 11 patients (20%) who were skeletally mature at follow-up, ten of whom had equal leg lengths and nine had a full range of movement of the hip and knee.

This is the largest reported series using non-invasive extendible endoprostheses after excision of primary bone tumours in skeletally immature patients. The technique produces a good functional outcome, with prevention of limb-length discrepancy at skeletal maturity.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 339 - 343
1 Mar 2012
Sewell MD Hanna SA Al-Khateeb H Miles J Pollock RC Carrington RWJ Skinner JA Cannon SR Briggs TWR

Patients with skeletal dysplasia are prone to developing advanced osteoarthritis of the knee requiring total knee replacement (TKR) at a younger age than the general population. TKR in this unique group of patients is a technically demanding procedure owing to the deformity, flexion contracture, generalised hypotonia and ligamentous laxity. We retrospectively reviewed the outcome of 11 TKRs performed in eight patients with skeletal dysplasia at our institution using the Stanmore Modular Individualised Lower Extremity System (SMILES) custom-made rotating-hinge TKR. There were three men and five women with mean age of 57 years (41 to 79). Patients were followed clinically and radiologically for a mean of seven years (3 to 11.5). The mean Knee Society clinical and function scores improved from 24 (14 to 36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to 80) and 50 (22 to 74), respectively, at final follow-up. Four complications were recorded, including a patellar fracture following a fall, a tibial peri-prosthetic fracture, persistent anterior knee pain, and aseptic loosening of a femoral component requiring revision. Our results demonstrate that custom primary rotating-hinge TKR in patients with skeletal dysplasia is effective at relieving pain, with a satisfactory range of movement and improved function. It compensates for bony deformity and ligament deficiency and reduces the likelihood of corrective osteotomy. Patellofemoral joint complications are frequent and functional outcome is worse than with primary TKR in the general population.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1169 - 1177
1 Sep 2011
Underwood R Matthies A Cann P Skinner JA Hart AJ

The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component.

There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p < 0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the in vivo wear rates are discussed: these may provide an explanation as to why the ASR is more sensitive to suboptimal positioning than the BHR.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1111 - 1117
1 Aug 2011
Sewell MD Hanna SA McGrath A Aston WJS Blunn GW Pollock RC Skinner JA Cannon SR Briggs TWR

The best method of reconstruction after resection of malignant tumours of the tibial diaphysis is unknown. In the absence of any long-term studies analysing the results of intercalary endoprosthetic replacement, we present a retrospective review of 18 patients who underwent limb salvage using a tibial diaphyseal endoprosthetic replacement following excision of a malignant bone tumour. There were ten men and eight women with a mean age of 42.5 years (16 to 76). Mean follow-up was 58.5 months (20 to 141) for all patients and 69.3 months (20 to 141) for the 12 patients still alive. Cumulative patient survival was 59% (95% confidence interval (CI) 32 to 84) at five years. Implant survival was 63% (95% CI 35 to 90) at ten years. Four patients required revision to a proximal tibial replacement at a mean follow-up of 29 months (10 to 54). Complications included metastases in five patients, aseptic loosening in four, peri-prosthetic fracture in two, infection in one and local recurrence in one. The mean Musculoskeletal Tumor Society score and the mean Toronto Extremity Salvage Score were 23 (17 to 28) and 74% (53 to 91), respectively.

Although rates of complication and revision were high, custom-made tibial diaphyseal replacement following resection of malignant bone tumours enables early return to function and provides an attractive alternative to other surgical options, without apparent compromise of patient survival.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1118 - 1121
1 Aug 2011
Berber O Dawson-Bowling S Jalgaonkar A Miles J Pollock RC Skinner JA Aston WJS Briggs TWR

We describe 22 cases of bizarre parosteal osteochondromatous proliferation, or Nora’s lesion. These are surface-based osteocartilaginous lesions typically affecting the hands and feet. All patients were identified from the records of a regional bone tumour unit and were treated between 1985 and 2009. Nine lesions involved the metacarpals, seven the metatarsals, one originated from a sesamoid bone of the foot and five from long bones (radius, ulna, tibia, and femur in two). The mean age of the patients was 31.8 years (6 to 66), with 14 men and eight women. Diagnosis was based on the radiological and histological features. The initial surgical treatment was excision in 21 cases and amputation of a toe in one. The mean follow-up was for 32 months (12 to 162). Recurrence occurred in six patients (27.3%), with a mean time to recurrence of 49 months (10 to 120). Two of the eight patients with complete resection margins developed a recurrence (25.0%), compared with four of 14 with a marginal or incomplete resection (28.6%).

Given the potential surgical morbidity inherent in resection, our data suggest that there may be a role for a relatively tissue-conserving approach to the excision of these lesions.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging.

Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 572 - 579
1 May 2011
Haddad FS Thakrar RR Hart AJ Skinner JA Nargol AVF Nolan JF Gill HS Murray DW Blom AW Case CP

Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 77 - 77
1 Jan 2011
Rossi R Rose B Riley ND Jennings R Saifuddin A Skinner JA Cannon SR Briggs T Pollock R
Full Access

Introduction: Within a study group of 102 consecutive patients diagnosed with chondrosarcoma of the femur, tibia or humerus, an association with previously treated breast cancer was noted. We researched this proposed relationship.

Methods: We retrospectively reviewed the records of all patients diagnosed histologically with chondrosarcoma of the femur, tibia or humerus over a six-year period at a supra-regional bone tumour unit. We identified those patients who had previously been treated for breast cancer.

Results: There were 58 female and 44 male patients. The study group contained six females (10%, mean age 53 years) who had previously been treated for breast cancer, a higher proportion than would be expected. They were referred following identification of a solitary area of increased activity on routine screening with isotope bone scan, presumed to be a solitary bony metastasis. Most (86%) of this breast carcinoma sub-group had developed low-grade bone chondrosarcoma (Trojani grade 0.5-I) and only one case (14%) had developed high-grade chondrosarcoma (Trojani grade II–III).

Discussion: A suspicious long bone lesion on bone scan in a patient with a past medical history of breast cancer must, therefore, not be assumed to be a metastasis without further investigation; the possibility of a chondral lesion should be considered. It is important that patients receive a full multidisciplinary team investigation prior to treatment in order to obtain the correct tissue diagnosis, as the management of these conditions is often different. Our study suggests there may be a relationship between patients previously treated for breast cancer and the development of subsequent chondrosarcoma.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 76 - 76
1 Jan 2011
Rose B Tamvakopoulos G Dulay K Pollock R Flanagan A Skinner JA Briggs TWR Cannon SR
Full Access

Introduction: Low-grade fibromyxoid sarcoma (LGFMS) is a rare soft tissue neoplasm most commonly presenting in young to middle-aged adults. LGFMS is an indolent tumour with a deceptively benign histological appearance. Local recurrences are not uncommon and tumours can metastasise. The FUS-CREB3L2 gene translocation has been shown to occur commonly in cases of LGFMS. The literature suggests that the FUS-CREB3L2 fusion-gene is a specific marker for LGFMS.

Methods: We report the cytogenetic analysis of 29 LGFMS cases, and clinical outcomes of 21 patients treated surgically between 1998 and 2008 at our regional bone and soft-tissue tumour centre.

Results: The mean age was 45.4 years and mean follow-up 30.1 months. The most common tumour location was the lower limb. There were no cases of local recurrence or metastasis. Fifteen patients (52.2%) were FUS-CREB3L2 translocation-positive, suggesting either that translocation incidence in our series is lower than other studies, or that reverse-transcriptase polymerase chain reaction (PCR) is less sensitive than the literature suggests. Patients testing positive presented at a younger age (38.2 years, compared to 45.6 years), and had larger tumours than their negative counterparts (mean diameter 97.6mm, compared to 65.2mm), although there was no difference in clinical outcome.

Discussion: We conclude that PCR testing for the FUS-CREB3L2 translocation is a useful tool for confirming the diagnosis of LGFMS, but has no role in predicting short-term clinical outcome. It is not necessary to perform wide excision, and marginal margins are adequate. Longer-term follow-up is required to elucidate differences in the long-term clinical outcome between translocation-positive and negative patients.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 75 - 75
1 Jan 2011
Parratt MTR Miles J Gokaraju K Spiegelberg BGI Pollock RC Skinner JA Cannon SR Briggs TWR Blunn GW
Full Access

Introduction: Intramedullary cementless fixation of massive tumor prostheses was developed to combat the high incidence of aseptic loosening in a young population of tumor patients. Its use has also expanded to include reconstruction of failed major knee arthroplasty. Our system uses a hydroxyapatite coated titanium stem in conjunction with a hydroxyapatite collar to obtain fixation.

Methods: We present a series of 72 patients (34 females, 38 males) with a mean follow-up of six years (2 months – 14 years). The mean age at the time of surgery was 17 (5 – 77). Within this group, 51 patients had osteosarcomas, 10 had giant cell tumors, eight had Ewing’s sarcoma and one each had malignant fibrous histiocytoma, spindle cell sarcoma and failed total knee arthroplasty. Non-invasive and minimally invasive growing prostheses were used in 33 patients.

Results: Nine patients required revision of the prosthesis at a mean of 5.2 years; 5 for aseptic loosening and four for infection. Twelve patients died at a mean of 20 months (1 – 84) post-operatively. Radiological evidence demonstrated remodelling around both the stem and the collar. There was gap closure to the hydroxyapatite coated collar in cases where seating was not complete. Consistent loss of cortical bone around the mid-stem was noted in the first 12 months but remained stable beyond this time frame.

Discussion: This study highlights our experience with cementless distal femoral endosprostheses. We demonstrate good results with regard to revision rate, gap closure and osseointegration.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 77 - 77
1 Jan 2011
Jennings R Riley ND Rose B Rossi R Saifuddin A Skinner JA Cannon SR Briggs TWR Pollock R
Full Access

Introduction: Chondrosarcoma is the second most common primary malignant bone tumour. Distinguishing between grades is not necessarily straightforward and may alter the management of the disease. We evaluated the correlation between the pre-operative needle biopsy and excision biopsy histological grading of chondrosarcoma of the femur, tibia and humerus.

Methods: A consecutive retrospective series of 100 patients with a histological diagnosis of chondrosarcoma made at a supra-regional bone tumour unit was reviewed. Twenty-one patients were excluded because 20 had only excision biopsy, due to radiological confidence in the diagnosis, and one had only the pre-operative biopsy on record, thus this series included 79 available cases. The remaining patients underwent a pre-operative needle biopsy.

Results: In 11 instances, there was a discrepancy in histological grade. Therefore, there was an 86% (68 out of 79) accuracy rate for pre-operative histological grading of chondrosarcoma, based on needle biopsy. However, the accuracy of the diagnostic biopsy to distinguish low-grade from high-grade was 90% (71 out of 79).

Discussion: From this series we conclude that accurate image-guided biopsy is a very useful adjunct in determining histological grade of chondrosarcoma and the subsequent treatment plan. At present, a multidisciplinary approach, comprising experienced Orthopaedic Surgeons, Radiologists and Pathologists offers the most reliable means of accurately diagnosing and grading chondrosarcoma of long bones.