The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 1 | Pages 51 - 53
1 Jan 1999
Stöckl B Sandow M Krismer M Biedermann R Wimmer C Frischhut B

We carried out 71 primary total hip arthroplasties using porous-coated, hemispherical press-fit Duraloc ‘100 Series’ cups in 68 consecutive patients; 61 were combined with the cementless Spotorno stem and ten with the cemented Lubinus SP II stem. Under-reaming of 2 mm achieved a press-fit. Of the 71 hips, 69 (97.1%) were followed up after a mean of 2.4 years. Migration analysis was performed by the Ein Bild Röntgen Analyse method, with an accuracy of 1 mm.

The mean total migration after 24 months was 1.13 mm. Using the definition of loosening as a total migration of 1 mm, it follows that 30 out of 63 cups (48%) were loose at 24 months.