header advert
Results 1 - 16 of 16
Results per page:
Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 5 - 5
1 Feb 2020
Burton W Myers C Rullkoetter P
Full Access

Introduction

Gait laboratory measurement of whole-body kinematics and ground reaction forces during a wide range of activities is frequently performed in joint replacement patient diagnosis, monitoring, and rehabilitation programs. These data are commonly processed in musculoskeletal modeling platforms such as OpenSim and Anybody to estimate muscle and joint reaction forces during activity. However, the processing required to obtain musculoskeletal estimates can be time consuming, requires significant expertise, and thus seriously limits the patient populations studied. Accordingly, the purpose of this study was to evaluate the potential of deep learning methods for estimating muscle and joint reaction forces over time given kinematic data, height, weight, and ground reaction forces for total knee replacement (TKR) patients performing activities of daily living (ADLs).

Methods

70 TKR patients were fitted with 32 reflective markers used to define anatomical landmarks for 3D motion capture. Patients were instructed to perform a range of tasks including gait, step-down and sit-to-stand. Gait was performed at a self-selected pace, step down from an 8” step height, and sit-to-stand using a chair height of 17”. Tasks were performed over a force platform while force data was collected at 2000 Hz and a 14 camera motion capture system collected at 100 Hz. The resulting data was processed in OpenSim to estimate joint reaction and muscle forces in the hip and knee using static optimization. The full set of data consisted of 135 instances from 70 patients with 63 sit-to-stands, 15 right-sided step downs, 14 left-sided step downs, and 43 gait sequences. Two classes of neural networks (NNs), a recurrent neural network (RNN) and temporal convolutional neural network (TCN), were trained to predict activity classification from joint angle, ground reaction force, and anthropometrics. The NNs were trained to predict muscle and joint reaction forces over time from the same input metrics. The 135 instances were split into 100 instances for training, 15 for validation, and 20 for testing.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 50 - 50
1 Feb 2020
Chen X Myers C Clary C Rullkoetter P
Full Access

INTRODUCTION

The magnitude of principal strain is indicative of the risks of femoral fracture,1,2 while changes in femoral strain energy density (SED) after total hip arthroplasty (THA) have been associated with bone remodeling stimulus.3 Although previous modeling studies have evaluated femoral strains in the intact and implanted femur under walking loads through successfully predicting physiological hip contact force and femoral muscle forces,1,2,3 strains during ‘high load’ activities of daily living have not typically been evaluated. Hence, the objective of this study was to compare femoral strain between the intact and the THA implanted femur under peak loads during simulated walking, stair descent, and stumbling.

METHODS

CTs of three cadaveric specimens were used to develop finite element (FE) models of intact and implanted femurs. Implanted models included a commercially-available femoral stem (DePuy Synthes, Warsaw, IN, USA). Young's moduli of the composite bony materials were interpolated from Hounsfield units using a CT phantom and established relationships.4 Peak hip contact force and femoral muscle forces during walking and stair descent were calculated using a lower extremity musculoskeletal model5 and applied to the femur FE models (Fig. 1). While maintaining the peak hip contact forces, muscle forces were further adjusted using an iterative optimization approach in FE models to reduce the femur deflection to the reported physiological range (< 5 mm).2 Femoral muscle forces during stumbling were estimated utilizing the same optimization approach with literature-reported hip contact forces as input.6 Maximum and minimum principal strains were calculated for each loading scenario. Changes in SED between intact and THA models were calculated in bony elements around the stem.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 129 - 129
1 Feb 2020
Maag C Langhorn J Rullkoetter P
Full Access

INTRODUCTION

While computational models have been used for many years to contribute to pre-clinical, design phase iterations of total knee replacement implants, the analysis time required has limited the real-time use as required for other applications, such as in patient-specific surgical alignment in the operating room. In this environment, the impact of variation in ligament balance and implant alignment on estimated joint mechanics must be available instantaneously. As neural networks (NN) have shown the ability to appropriately represent dynamic systems, the objective of this preliminary study was to evaluate deep learning to represent the joint level kinetic and kinematic results from a validated finite element lower limb model with varied surgical alignment.

METHODS

External hip and ankle boundary conditions were created for a previously-developed finite element lower limb model [1] for step down (SD), deep knee bend (DKB) and gait to best reproduce in-vivo loading conditions as measured on patients with the Innex knee (orthoload.com) (Figure1). These boundary conditions were subsequently used as inputs for the model with a current fixed-bearing total knee replacement to estimate implant-specific kinetics and kinematics during activities of daily living. Implant alignments were varied, including variation of the hip-knee-ankle angle-±3°, the frontal plane joint line −7° to +5°, internal-external femoral rotation ±3°, and the tibial posterior slope 5° and 0°. Through varying these parameters a total of 2464 simulations were completed.

A NN was created utilizing the NN toolbox in MATLAB. Sequence data inputs were produced from the alignment and the external boundary conditions for each activity cycle. Sequence outputs for the model were the 6 degree of freedom kinetics and kinematics, totaling 12 outputs. All data was normalized across the entire data set. Ten percent of the simulation runs were removed at random from the training set to be used for validation, leaving 2220 simulations for training and 244 for validation. A nine-layer bi-long short-term memory (LSTM) NN was created to take advantage of bi-LSTM layers ability to learn from past and future data. Training on the network was undertaken using an RMSprop solver until the root mean square error (RMSE) stopped reducing. Evaluation of NN quality was determined by the RMSE of the validation set.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 6 - 6
1 Feb 2020
Burton W Myers C Rullkoetter P
Full Access

Introduction

Real-time tracking of surgical tools has applications for assessment of surgical skill and OR workflow. Accordingly, efforts have been devoted to the development of low-cost systems that track the location of surgical tools in real-time without significant augmentation to the tools themselves. Deep learning methodologies have recently shown success in a multitude of computer vision tasks, including object detection, and thus show potential for the application of surgical tool tracking. The objective of the current study was to develop and evaluate a deep learning-based computer vision system using a single camera for the detection and pose estimation of multiple surgical tools routinely used in both knee and hip arthroplasty.

Methods

A computer vision system was developed for the detection and 6-DoF pose estimation of two surgical tools (mallet and broach handle) using only RGB camera frames. The deep learning approach consisted of a single convolutional neural network (CNN) for object detection and semantic key point prediction, as well as an optimization step to place prior known geometries into the local camera coordinate system. Inference on a camera frame with size of 256-by-352 took 0.3 seconds. The object detection component of the system was evaluated on a manually-annotated stream of video frames. The accuracy of the system was evaluated by comparing pose (position and orientation) estimation of a tool with the ground truth pose as determined using three retroreflective markers placed on each tool and a 14 camera motion capture system (Vicon, Centennial CO). Markers placed on the tool were transformed into the local camera coordinate system and compared to estimated location.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 45 - 45
1 Mar 2017
Myers C Laz P Shelburne K Rullkoetter P
Full Access

Introduction

Alignment of the acetabular cup and femoral components directly affects hip joint loading and potential for impingement and dislocation following total hip arthroplasty (THA) [1]. Changes to the lines of action and moment generating capabilities of the muscles as a result of component position may influence overall patient function. The objectives of this study were to assess the effect of component placement on hip joint contact forces (JCFs) and muscle forces during a high demand step down task and to identify important alignment parameters using a probabilistic approach.

Methods

Three patients following THA (2 M: 28.3±2.8 BMI; 1 F: 25.7 BMI) performed lower extremity maximum isometric strength tests and a step down task as part of a larger IRB-approved study. Patient-specific musculoskeletal models were created by scaling a model with detailed hip musculature [2] to patient segment dimensions and mass. For each model, muscle maximum isometric strengths were optimized to minimize differences between model-predicted and measured preoperative maximum isometric joint torques at the hip and knee.

Baseline simulations used patient-specific models with corresponding measured kinematics and ground reaction forces to predict hip JCFs and muscle forces using static optimization. To assess the combined effects of stem and cup position and orientation, a 1000 trial Monte Carlo simulation was performed with input variability in each degree of freedom based on the ±1 SD range in component placement relative to native geometry reported by Tsai et al. [3] (Figure 1). Maximum confidence bounds (1–99%) were predicted for the hip JCF magnitude and muscle forces for three prime muscles involved in the task (gluteus medius, gluteus minimus and psoas). HJC confidence bounds were compared to Orthoload measurements from telemetric implants from 6 patients performing the step down task. Sensitivity of hip JCF and muscle force outputs was quantified by Pearson Product-Moment correlation between the input parameter and the value of each output averaged across four points in the cycle.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 10 - 10
1 Feb 2017
Ali A Mannen E Smoger L Haas B Laz P Rullkoetter P Shelburne K
Full Access

Introduction

Patellar resurfacing affects patellofemoral (PF) kinematics, contact mechanics, and loading on the patellar bone. Patients with total knee arthroplasty (TKA) often exhibit adaptations in movement patterns that may be linked to quadriceps deficiency and the mechanics of the reconstructed knee [1]. Previous comparisons of PF kinematics between dome and anatomic resurfacing have revealed differences in patellar sagittal plane flexion [2], but further investigation of PF joint mechanics is required to understand how these differences influence performance. The purpose of this study was to compare PF mechanics between medialized dome and medialized anatomic implants using subject-specific computational models.

Methods

A high-speed stereo radiography (HSSR) system was used to capture 3D sub-mm measurement of bone and implant motion [3]. HSSR images were collected for 10 TKA patients with Attune® (DePuy Synthes, Warsaw, IN) posterior-stabilized, rotating-platform components, 5 with medialized dome and 5 with medialized anatomic patellar components (3M/7F, 62.5±6.6 years, 2.2±0.6 years post-surgery, BMI: 26.2±3.5 kg/m2), performing two activities of daily living: knee extension and lunge (Figure 1). Relative motions were tracked using Autoscoper (Brown University, Providence, RI) for implant geometries obtained from the manufacturer. A statistical shape model was used to predict the patella and track motions [4].

Subject-specific finite element models of the experiment were developed for all subjects and activities [5]. The model included implant components, patella, quadriceps, patellar tendon, and medial and lateral PF ligaments (Figure 2a). While tibiofemoral kinematics were prescribed based on experimental data, the PF joint was unconstrained. A constant 1000N quadriceps load was distributed among four muscle groups. Soft tissue attachments and pre-strain in PF ligaments were calibrated to match experimental kinematics [5]. Model outputs included PF kinematics, patellar and contact force ratios, patellar tendon angle, and moment arm.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 118 - 118
1 Feb 2017
Fitzpatrick C Clary C Rullkoetter P
Full Access

Introduction

Patellar crepitus and clunk are tendofemoral-related complications predominantly associated with posterior-stabilizing (PS) total knee arthroplasty (TKA) designs [1]. Contact between the quadriceps tendon and the femoral component can cause irritation, pain, and catching of soft-tissue within the intercondylar notch (ICN). While the incidence of tendofemoral-related pathologies has been documented for some primary TKA designs, literature describing revision TKA is sparse. Revision components require a larger boss resection to accommodate a constrained post-cam and stem/sleeve attachments, which elevates the entrance to the ICN, potentially increasing the risk of crepitus. The objective of this study was to evaluate tendofemoral contact in primary and revision TKA designs, including designs susceptible to crepitus, and newer designs which aim to address design features associated with crepitus.

Methods

Six PS TKA designs were evaluated during deep knee bend using a computational model of the Kansas knee simulator (Figure 1). Prior work has demonstrated that tendofemoral contact predictions from this model can differentiate between TKA patients with patellar crepitus and matched controls [2]. Incidence of crepitus of up to 14% has been reported in Insall-Burstein® II and PFC® Sigma® designs [3]. These designs, in addition to PFC® Sigma® TC3 (revision component), were included in the analyses. Primary and revision components of newer generation designs (NexGen®, Attune® and Attune® Revision) were also included. Designs were evaluated in a patient model with normal Insall-Salvati ratio and a modified model with patellar tendon length reduced by two standard deviations (13mm) to assess worst-case patient anatomy.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 22 - 22
1 Feb 2017
Huff D Schwartz B Fitzpatrick C Rullkoetter P
Full Access

INTRODUCTION

Within total hip replacement, articulation of the femoral head near the rim of the acetabular liner creates undesirable conditions leading to a propensity for dislocation[1], increased contact stresses[2], increased load and torque imparted on the acetabular component[3], and increased wear[4]. Propensity for rim loading is affected by prosthesis placement, as well as the kinematics and loading of the patient. The present study investigates these effects.

METHODS

CT scans from an average-sized patientwere segmented for the hemipelvis and femur of interest. DePuy Synthes implant models were aligned in a neutral position in Hypermesh. The acetabular liner was assigned deformable solid material properties, and the remainder of the model was assigned rigid properties.

Joint reaction forces and kinematics of hip flexion were taken from the public Orthoload database to represent ADLs [5]: Active flexion lying on a table, gait, bending to lift and move a load, and sit-stand. The pelvis was fully constrained, while three-degree-of-freedom (3-DOF) forces were applied to the femur. Hip flexion was kinematically-prescribed while internal-external (I-E) and adduction-abduction (Ad-Ab) DOFs were constrained.

Angles of acetabular implant positioning were based on published data by Rathod [6]. Femoral implant position was chosen based on cadaveric in vitro DePuy Synthes measurements of variation in femoral prosthesis position reported previously [7]. Acetabular and Femoral alignment angles were represented for nominal position, as well as positioning + 1σ and + 2σ from the mean in both anteversion and inclination for acetabular components, and both Varus/Valgus and Flexion (angle in sagittal plane) for the femoral component.

The analyses were automated within Matlab to execute 68 finite element analyses in Abaqus Explicit and structured in a DOE style analysis with Cup inclination, Cup version, Stem Flexion, and Stem Varus/Valgus, and Activity as variables of interest (64 runs + 4 centerpoints = 68 analyses).

From a previous study it was known that acetabular component inclination had the greatest effect on contact pressure location [7], so all data were analyzed relative to inclination, allowing other positioning variables to be represented as variation per inclination position. Results are presented as a percentage, with 0% being pole loading and 100% being rim loading, to normalize for head diameter.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 29 - 29
1 May 2016
Banks S Kefala V Cyr A Shelburne K Rullkoetter P
Full Access

“How does the knee move?” is a question of fundamental importance for treatment of knee injuries and knee replacement design. Unfortunately, we lack unambiguous and comprehensive knee function data sets and/or consensus on how healthy knees move. One can just as easily find reports stating the natural knee has a center of axial rotation in the medial compartment of the knee as in the lateral. This is due to technical and practical issues: It is extremely difficult to accurately measure knee motions during ambulatory activities and, when that can be done, very few studies have examined a range of weightbearing activities in the same study cohort. The purpose of this study is to report knee kinematics in a cohort of healthy older subjects whose motions were examined during four different movements, three of them weightbearing ambulation, using a high-speed stereoradiographic system.

Six healthy consenting subjects (age = 61 ± 5 years, body mass = 75 ± 8 kg, BMI = 27 ± 4) were observed using a high-speed stereoradiographic system while completing four tasks. Subjects were instructed to perform an unloaded, seated knee extension from high flexion to full extension; to walk at a self selected pace; to step down from a 7 inch platform; and to walk and perform a 90° direction change (pivoting). Stereoradiographic images (1080 × 1080 pixels) were acquired at 100 images/second using 40cm image intensifiers and pulsed x-ray exposures. The three-dimensional knee kinematics were measured using the XROMM software suite (xromm.org, Brown University). Post-processing of the kinematics was performed in custom Matlab programs, and included fitting spheres to the posterior condylar surfaces of each knee, and then tracking the motions of the sphere centers relative to a fixed tibial reference frame (Figure 1). The motions of these flexion-facet centers, were used to determine an average center of axial rotation (CoR) over each activity as previously reported by Banks and Hodge.

Average CoRs for all four activities were in the posterior-medial quadrant of the knee, with the CoR for open-chain knee extension being the most medial and gait the most lateral (Table 1, Figure 2). One-way ANOVA showed average CoRs are different (p « 0.001). There was considerable variation in individual CoRs, for example, with two knees showing lateral CoRs for gait and the remaining knees having medial CoRs.

It should not be surprising that natural knee motions vary with dynamic activity, yet knee kinematics often are presented as being one stereotypic, monolithic pattern of motion. Our data show that the same healthy subjects performing different dynamic activities manifest different knee motions, with open-chain knee extension having the most medial CoR and gait the most lateral. This finding is consistent with previous reports comparing stair climbing and gait in knees with various implant designs. Additional experimental data and, ultimately, validated numerical simulations should facilitate an increasingly accurate process for designing improved treatments for diseased and damaged knees.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 131 - 131
1 Jan 2016
Fitzpatrick CK Clary C Nakamura T Rullkoetter P
Full Access

Introduction

The current standard for alignment in total knee arthroplasty (TKA) is neutral mechanical axis within 3° of varus or valgus deviation [1]. This configuration has been shown to reduce wear and optimally distribute load on the polyethylene insert [2]. Two key factors (patient-specific hip-knee-ankle (HKA) angle and surgical component alignment) influence load distribution, kinematics and soft-tissue strains across the tibiofemoral (TF) joint. Improvements in wear characteristics of TKA materials have facilitated a trend for restoring the anatomic joint line [3]. While anatomic component alignment may aid in restoring more natural kinematics, the influence on joint loads and soft-tissue strains should be evaluated. The purpose of the current study was to determine the effect of varus component alignment in combination with a variety of HKA limb alignments on joint kinematics, loads and soft-tissue strain.

Methods

A dynamic three-dimensional finite element model of the lower limb of a TKA patient was developed. Detailed description of the model has been previously published [4]. The model included femur, tibia and patella bones, TF ligaments, patellar tendon, quadriceps and hamstrings, and was virtually implanted with contemporary cruciate-retaining fixed-bearing TKA components. The model was initially aligned in ideal mechanical alignment with neutral HKA limb alignment. A design-of-experiments (DOE) study was performed whereby component placement was altered from neutral to 3° and 7° varus alignment, and HKA angle was altered from neutral to ±3° and ±7° (valgus and varus) (Figure 1).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 132 - 132
1 Jan 2016
Fitzpatrick CK Nakamura T Niki Y Rullkoetter P
Full Access

Introduction

A large number of total knee arthroplasty (TKA) patients, particularly in Japan, India and the Middle East, exhibit anatomy with substantial proximal tibial torsion. Alignment of the tibial components with the standard anterior-posterior (A-P) axis of the tibia can result in excessive external rotation of the tibial components with respect to femoral component alignment. This in turn influences patellofemoral (PF) mechanics and forces required by the extensor mechanism. The purpose of the current study was to determine if a rotating-platform (RP) TKA design with an anatomic patellar component reduced compromise to the patellar tendon, quadriceps muscles and PF mechanics when compared to a fixed-bearing (FB) design with a standard dome-shaped patellar component.

Methods

A dynamic three-dimensional finite element model of the knee joint was developed and used to simulate a deep knee bend in a patient with excessive external tibial torsion (Figure 1). Detailed description of the model has been previously published [1]. The model included femur, tibia and patellar bones, TKA components, patellar ligament, quadriceps muscles, PF ligaments, and nine primary ligaments spanning the TF joint. The model was virtually implanted with two contemporary TKA designs; a FB design with domed patella, and a RP design with anatomic patella. The FB design was implanted in two different alignment conditions; alignment to the tibial A-P axis, and optimal alignment for bone coverage. Four different loading conditions (varying internal-external (I-E) torque and A-P force) were applied to the model to simulate physiological loads during a deep knee bend. Quadriceps muscle force, patellar tendon force, and PF and TF joint forces were compared between designs.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 54 - 54
1 Jul 2014
Fitzpatrick C Vigneron L Kannan S Shah S (Cheryl) Liu X De Boodt S Rullkoetter P
Full Access

Summary Statement

Computational models are the primary tools for efficient design-phase exploration of knee replacement concepts before in vitro testing. To improve design-phase efficiency, a subject-specific computational platform was developed that allows designers to assess devices in realistic conditions by directly integrating subject-specific experimental data in these models.

Introduction

Early in the design-phase of new implant design, numerous in vitro tests would be desirable to assess the influence of design parameters or component alignment on the performance of the device. However, cadaveric testing of knee replacement devices is a costly and time-consuming procedure, requiring manufacture of parts, preparation of cadaveric specimens, and personnel to carry of the experiments. Validated computational models are ideally suited for pre-clinical, high-volume design evaluation. Initial development of these models requires substantial time and expertise; once developed, however, computational simulations may be applied for comparative evaluation of devices in an extremely efficient manner [Baldwin et al. 2012]. Still, computational models are complementary of experimental testing and for this reason, computational models tuned with subject-specific experimental data, e.g. soft tissue parameters, could bring even more efficiency in the design phase. The objective of the current study was to develop a platform of tools that easily allows for subject-specific knee simulations. The system integrates with commercially available medical imaging and finite element software to allow for direct, efficient comparison of designs and surgical alignment under a host of different boundary conditions.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 411 - 411
1 Nov 2011
Baldwin M Clary C Maletsky L Rullkoetter P
Full Access

Design phase evaluation of potential implant designs requires verified computational and experimental models. Computational models are important where parametric evaluation of geometric features experimentally is both cost and time-prohibitive due to the need to manufacture complex parts, and provide information not easily measured experimentally, such as internal stresses/strains in the implant or bone.

However, before implementation into the design process, a thorough verification/validation is required. In this study, a finite element model of the Kansas knee simulator (KKS) was developed and a systematic verification of predicted joint kinematics was performed by comparison with experimental measurements, including evaluating the patellofemoral joint first in isolation, followed by whole joint kinematic comparisons.

Four unmatched, healthy cadaver knees (average age 63 yrs) were mounted in the KKS to reproduce a simulated gait and deep knee bend activity in their natural and implanted states. Finite element models of the KKS assembly and the four cadaver specimens in their natural and implanted states were created. Isolated patellofem-oral kinematics were initially verified during simulated deep knee bend. Average RMS differences between predicted and experimental natural patellar kinematics were less than 3.1° and 1.7 mm for rotations and translations, respectively, while differences in implanted kinematics were less than 2.1° and 1.6 mm between 10 and 110° femoral flexion. Similar agreement was found with the subsequent whole joint simulations.

Deep knee bend tibiofemoral internal-external (IE) and varus-valgus (VV) rotations had average RMS differences from experimental measurements of 1.5 ± 0.4° and 0.9 ± 0.5°, respectively. Anterior-posterior (AP), inferior-superior (IS) and medial-lateral translations matched within 1.8 ±0.8 mm, 1.2 ±0.7 mm, and 0.6 ±0.1 mm, respectively.

The experimental and verified computational tools can be used in harmony for pre-clinical assessment of implant designs; the computational model allows rapid screening of implant geometry or alignment issues and provides additional insight into joint mechanics such as implant stresses or bone strains, while the experimental simulator can subsequently be utilized to assess in cadavera only the most promising designs or features identified.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 465 - 465
1 Nov 2011
Petrella A Armstrong J Patel V Laz P Rullkoetter P
Full Access

Cross-shear has been shown to increase ultra-high molecular weight polyethylene (UHMWPE) wear in pin-on-disk, total knee, total hip, and spinal disc replacement testing. Computer modelling of implant wear holds promise for improving efficiency in the development of new implant designs, but it is desirable to accurately account for the effects of cross-shear in the computational simulation. Several studies have sought to propose a quantitative metric for cross-shear in multidirectional sliding and to correlate average cross-shear intensity with apparent wear rate measured in experiments. The apparent wear rate accounts for the total volume loss from all points on the UHMWPE surface. In principle, if the cross-shear metric correlates with experimental wear rates, it is then possible to predict an estimated wear rate for any arbitrary set of kinematic inputs. UHMWPE wear may then be simulated numerically with some form of Archard’s law.

One limitation of the above approach is that counter-face kinematics are homogenized by the use of a spatially and temporally averaged apparent wear rate. In a sliding contact interface of a joint implant in vivo, the intensity of cross-shear wear may vary with time and location on the surface. To address this variation we have proposed a novel cross-shear metric (x*) and developed a modified form of Archard’s law that is capable to differentiate between unidirectional and multidirectional sliding wear. The wear model and x* have been implemented in an explicit finite element framework (ABAQUS) that is capable of quantifying wear from any number of wear surfaces (e.g., front side, backside, post) with completely general geometry and loading conditions.

Preliminary validation of x* and the wear model have been performed by comparison with data from the open literature. Cross-shear metric x* is easy to compute, exhibits invariance to the choice of kinematic reference frame, and is able to reliably distinguish between similarly shaped sliding paths of different lengths – all improvements compared to cross-shear metrics described elsewhere. The wear model that incorporates x* has shown good agreement with pin-on-disk and cervical disc replacement wear results previously reported. Ongoing research focuses on demonstrating similar validity of the model for cross-shear wear in hip and knee replacements.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 460 - 460
1 Nov 2011
Clary C Pal S Rullkoetter P
Full Access

Musculoskeletal models of the lower limb lend insight into muscle forces and joint mechanics during dynamic activities. However, traditional musculoskeletal modeling is based on rigid body assumptions, and frequently represents the knee as a hinge joint, neglecting the complex interactions between the patella, femur, and tibia. Implementation of the musculoskeletal modeling framework in an explicit finite element environment allows joint contact to be easily incorporated, as well as representation of any structure as rigid or fully deformable in order to evaluate, for example, implant stresses or bone strain. Prediction of these values is particularly valuable when evaluating implant mechanics after total knee replacement.

A finite element, musculoskeletal model of an implanted right lower limb was constructed, including thirteen muscles crossing the knee joint. A Hill-type muscle model was developed to allow muscle activation within the explicit FE framework. Muscle forces were predicted by optimization of muscle activation patterns during flexion-extension and chair-rise activities. The effect of muscle path representation was investigated using two approaches: lines of action directly between the origin and insertion sites of the muscles, and lines of action along the centroid of the muscle bodies. Incorporating anatomic muscle paths into the model reduced the predicted peak quadriceps force during the chair-rise activity by 46%, and reduced the peak tibio-femoral contact pressure by 14%. In addition, bone strain was predicted during the activity for the implanted patella, and showed peak bone strain at the edge of the implant near the inferior pole.

The muscle-activated models demonstrated the advantages of an explicit finite element framework, and allow rapid, rigid body simulation in addition to the full contact, deformable analyses when greater resolution is required.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 413 - 414
1 Nov 2011
Dennis D Kim R Johnson D Springer B Fehring T Rullkoetter P Laz P Baldwin M
Full Access

Introduction: Patellar crepitus (PC) has been reported in 13% of cruciatesubstituting total knee arthroplasty (TKA) patients resulting from synovial tissue impingement within the femoral component intercondylar box (IB). Patient factors, component design, and technical errors have been implicated in PC. We compared primary TKA patients with PC requiring surgery against matched controls to identify significant variables.

Methods: The databases of 2 institutions were reviewed to identify patients requiring surgery for PC. A control group matched for age, sex, and BMI was identified.

Patient charts and radiographs were reviewed. Statistical analysis was performed.

Significant variables associated with patient anatomy, implant size and alignment were subsequently investigated in a computational model to evaluate tendofemoral contact.

Results: Between 2002 and 2008, over 4000 primary TKAs were performed using the Press Fit Condylar Sigma (DePuy, Warsaw, Indiana) TKA. Of these, 59 knees developed PC requiring surgery. The mean time to presentation was 10.9 months. The incidence of PC correlated with greater number of previous surgeries (1.18 vs. 0.44, p= 0.002), decreased patellar button size (35.7 vs. 37.1mm, p=0.003), shorter patellar tendon length (54.5 vs. 57.9mm, p=0.01), and increase in posterior femoral condylar offset (1.27mm vs. 0.17mm, p=0.022). Using a patellar component of 32 or 35mm significantly increased the risk of PC compared to the use of a 38 or 41mm component (p< 0.01, RR=1.61, OR 2.63). Modeling results demonstrated decreased patellar tendon length created increased tendofemoral contact near the IB, while larger buttons increased separation between the tendon and the box edge.

Conclusion: Shortened patellar tendon length and use of smaller patellar components may expose the quadriceps tendon to increased irritation as it traverses across the femoral component IB. Increasing posterior femoral offset may increase quadriceps tendon tension, further risking synovial tissue impingement within the IB.