header advert
Results 1 - 20 of 25
Results per page:
The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 74 - 81
1 May 2024
Callary SA Broekhuis D Barends J Ramasamy B Nelissen RGHH Solomon LB Kaptein BL

Aims

The aim of this study was to compare the biomechanical models of two frequently used techniques for reconstructing severe acetabular defects with pelvic discontinuity in revision total hip arthroplasty (THA) – the Trabecular Metal Acetabular Revision System (TMARS) and custom triflange acetabular components (CTACs) – using virtual modelling.

Methods

Pre- and postoperative CT scans from ten patients who underwent revision with the TMARS for a Paprosky IIIB acetabular defect with pelvic discontinuity were retrospectively collated. Computer models of a CTAC implant were designed from the preoperative CT scans of these patients. Computer models of the TMARS reconstruction were segmented from postoperative CT scans using a semi-automated method. The amount of bone removed, the implant-bone apposition that was achieved, and the restoration of the centre of rotation of the hip were compared between all the actual TMARS and the virtual CTAC implants.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 136 - 143
1 Feb 2024
van der Lelij TJN Marang-van de Mheen PJ Kaptein BL Koster LA Ljung P Nelissen RGHH Toksvig-Larsen S

Aims

The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface.

Methods

Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1168 - 1176
1 Nov 2023
Yüksel Y Koster LA Kaptein BL Nelissen RGHH den Hollander P

Aims

Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes.

Methods

A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 977 - 984
1 Sep 2023
Kamp T Gademan MGJ van Zon SKR Nelissen RGHH Vliet Vlieland TPM Stevens M Brouwer S

Aims

For the increasing number of working-age patients undergoing total hip or total knee arthroplasty (THA/TKA), return to work (RTW) after surgery is crucial. We investigated the association between occupational class and time to RTW after THA or TKA.

Methods

Data from the prospective multicentre Longitudinal Leiden Orthopaedics Outcomes of Osteoarthritis Study were used. Questionnaires were completed preoperatively and six and 12 months postoperatively. Time to RTW was defined as days from surgery until RTW (full or partial). Occupational class was preoperatively assessed and categorized into four categories according to the International Standard Classification of Occupations 2008 (blue-/white-collar, high-/low-skilled). Cox regression analyses were conducted separately for THA and TKA patients. Low-skilled blue-collar work was used as the reference category.


Aims

The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs).

Methods

A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 148 - 157
1 Feb 2023
Koster LA Rassir R Kaptein BL Sierevelt IN Schager M Nelissen RGHH Nolte PA

Aims

The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs.

Methods

A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 221 - 226
1 Feb 2022
Edwards NM Varnum C Nelissen RGHH Overgaard S Pedersen AB

Aims

The aim of this study was to examine whether socioeconomic status (SES) is associated with a higher risk of infections following total hip arthroplasty (THA) at 30 and 90 days.

Methods

We obtained individual-based information on SES markers (cohabitation, education, income, and savings) on 103,901 THA patients from Danish health registries between 1 January 1995 and 31 December 2017. The primary outcome measure was any hospital-treated infection (i.e. all infections). The secondary outcomes were further specified to specific hospital-treated infections (pneumonia, urinary tract infection, and periprosthetic joint infection). The primary timepoint was within 90 days. In addition, the outcomes were further evaluated within 30 days. We calculated the cumulative incidence, and used the pseudo-observation method and generalized linear regression to estimate adjusted risk ratios (RRs) with 95% confidence intervals (CIs) for each marker.


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 855 - 863
1 May 2021
Koster LA Meinardi JE Kaptein BL Van der Linden - Van der Zwaag E Nelissen RGHH

Aims

The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS).

Methods

A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively.


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1158 - 1166
14 Sep 2020
Kaptein BL den Hollander P Thomassen B Fiocco M Nelissen RGHH

Aims

The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs).

Methods

A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1016 - 1024
1 Aug 2020
Hasan S van Hamersveld KT Marang-van de Mheen PJ Kaptein BL Nelissen RGHH Toksvig-Larsen S

Aims

Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using radiostereometric analysis (RSA) known for its ability to predict aseptic loosening.

Methods

A total of 72 patients were randomized to either cementless 3D-printed or a cemented cruciate retaining TKA. RSA and clinical scores were evaluated at baseline and postoperatively at three, 12, and 24 months. A mixed model was used to analyze the repeated measurements.


The Bone & Joint Journal
Vol. 100-B, Issue 3 | Pages 309 - 317
1 Mar 2018
Kolk A Henseler JF Overes FJ Nagels J Nelissen RGHH

Aims

Since long-term outcome of teres major tendon transfer surgery for irreparable posterosuperior rotator cuff (RC) tears is largely unknown, the primary aim of this study was to evaluate the long-term outcome of the teres major transfer. We also aimed to report on the results of a cohort of patients with a similar indication for surgery that underwent a latissimus dorsi tendon transfer.

Patients and Methods

In this prospective cohort study, we reported on the long-term results of 20 consecutive patients with a teres major tendon transfer for irreparable massive posterosuperior RC tears. Additionally, we reported on the results of the latissimus dorsi tendon transfer (n = 19). The mean age was 60 years (47 to 77). Outcomes included the Constant score (CS), and pain at rest and during movement using the Visual Analogue Scale (VAS).


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1333 - 1341
1 Oct 2016
van der Voort P Valstar ER Kaptein BL Fiocco M van der Heide HJL Nelissen RGHH

Aims

The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos R) bone cement is no longer commercially available and was superseded by Refobacin bone cement R and Palacos R + G in 2005. However, the performance of these newly introduced bone cements have not been tested in a phased evidence-based manner, including roentgen stereophotogrammetric analysis (RSA).

Patients and Methods

In this blinded, randomised, clinical RSA study, the migration of the Stanmore femoral component was compared between Refobacin bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties. The primary outcome measure was femoral component migration measured using RSA and secondary outcomes were Harris hip score (HHS), Hip disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D) and Short Form 36 (SF-36).


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 391 - 397
1 Mar 2015
van Embden D Stollenwerck GANL Koster LA Kaptein BL Nelissen RGHH Schipper IB

The aim of this study was to quantify the stability of fracture-implant complex in fractures after fixation. A total of 15 patients with an undisplaced fracture of the femoral neck, treated with either a dynamic hip screw or three cannulated hip screws, and 16 patients with an AO31-A2 trochanteric fracture treated with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric analysis was used at six weeks, four months and 12 months post-operatively to evaluate shortening and rotation.

Migration could be assessed in ten patients with a fracture of the femoral neck and seven with a trochanteric fracture. By four months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1) had occurred in the fracture of the femoral neck group and 5.0 mm (-0.13 to 12.9) in the trochanteric fracture group. A wide range of rotation occurred in both types of fracture. Right-sided trochanteric fractures seem more rotationally stable than left-sided fractures.

This prospective study shows that migration at the fracture site occurs continuously during the first four post-operative months, after which stabilisation occurs. This information may allow the early recognition of patients at risk of failure of fixation.

Cite this article: Bone Joint J 2015;97-B:391–7.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1520 - 1524
1 Nov 2014
van der Zwaal P Pijls BG Thomassen BJW Lindenburg R Nelissen RGHH van de Sande MAJ

The purpose of this study was to evaluate the natural history of rheumatoid disease of the shoulder over an eight-year period. Our hypothesis was that progression of the disease is associated with a decrease in function with time.

A total of 22 patients (44 shoulders; 17 women, 5 men, (mean age 63)) with rheumatoid arthritis were followed for eight years. All shoulders were assessed using the Constant score, anteroposterior radiographs (Larsen score, Upward-Migration-Index (UMI)) and ultrasound (US). At final follow-up, the Short Form-36, disabilities of the arm, shoulder and hand (DASH) Score, erythrocyte sedimentation rate and use of anti-rheumatic medication were determined.

The mean Constant score was 72 points (50 to 88) at baseline and 69 points (25 to 100) at final follow-up. Radiological evaluation showed progressive destruction of the peri-articular structures with time. This progression of joint and rotator cuff destruction was significantly associated with the Constant score. However, at baseline only the extent of rotator cuff disease and the UMI could predict the Constant score at final follow-up.

A plain anteroposterior radiograph of the shoulder is sufficient to assess any progression of rheumatoid disease and to predict functional outcome in the long term by using the UMI as an indicator of rotator cuff degeneration.

Cite this article: Bone Joint J 2014;96-B:1520–4.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 765 - 771
1 Jun 2014
Thomassen BJW den Hollander PHC Kaptijn HH Nelissen RGHH Pilot P

We hypothesised there was no clinical value in using an autologous blood transfusion (ABT) drain in either primary total hip (THR) or total knee replacement (TKR) in terms of limiting allogeneic blood transfusions when a modern restrictive blood management regime was followed. A total of 575 patients (65.2% men), with a mean age of 68.9 years (36 to 94) were randomised in this three-arm study to no drainage (group A), or to wound drainage with an ABT drain for either six hours (group B) or 24 hours (group C). The primary outcome was the number of patients receiving allogeneic blood transfusion. Secondary outcomes were post-operative haemoglobin (Hb) levels, length of hospital stay and adverse events.

This study identified only 41 transfused patients, with no significant difference in distribution between the three groups (p = 0.857). The mean pre-operative haemoglobin (Hb) value in the transfused group was 12.8 g/dL (9.8 to 15.5) versus 14.3 g/dL (10.6 to 18.0) in the non-transfused group (p < 0.001, 95% confidence interval: 1.08 to 1.86). Post-operatively, the median of re-transfused shed blood in patients with a THR was 280 mL (Interquartile range (IQR) 150 to 400) and in TKR patients 500 mL (IQR 350 to 650) (p <  0.001). ABT drains had no effect on the proportion of transfused patients in primary THR and TKR. The secondary outcomes were also comparable between groups.

Cite this article: Bone Joint J 2014;96-B:765–71.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1656 - 1661
1 Dec 2013
Kraal T van der Heide HJL van Poppel BJ Fiocco M Nelissen RGHH Doets HC

Little is known about the long-term outcome of mobile-bearing total ankle replacement (TAR) in the treatment of end-stage arthritis of the ankle, and in particular for patients with inflammatory joint disease. The aim of this study was to assess the minimum ten-year outcome of TAR in this group of patients.

We prospectively followed 76 patients (93 TARs) who underwent surgery between 1988 and 1999. No patients were lost to follow-up. At latest follow-up at a mean of 14.8 years (10.7 to 22.8), 30 patients (39 TARs) had died and the original TAR remained in situ in 28 patients (31 TARs). The cumulative incidence of failure at 15 years was 20% (95% confidence interval (CI) 11 to 28). The mean American Orthopaedic Foot and Ankle Society (AOFAS) ankle–hindfoot score of the surviving patients at latest follow-up was 80.4 (95% CI 72 to 88). In total, 21 patients (23 TARs) underwent subsequent surgery: three implant exchanges, three bearing exchanges and 17 arthrodeses. Neither design of TAR described in this study, the LCS and the Buechel–Pappas, remains currently available. However, based both on this study and on other reports, we believe that TAR using current mobile-bearing designs for patients with end-stage arthritis of the ankle due to inflammatory joint disease remains justified.

Cite this article: Bone Joint J 2013;95-B:1656–61.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1626 - 1631
1 Dec 2013
van der Weegen W Brakel K Horn RJ Hoekstra HJ Sijbesma T Pilot P Nelissen RGHH

The aim of this study was to establish the natural course of unrevised asymptomatic pseudotumours after metal-on-metal (MoM) hip resurfacing during a six- to 12-month follow-up period. We used repeated metal artefact reduction sequence (MARS)-magnetic resonance imaging (MRI), serum metal ion analysis and clinical examination to study 14 unrevised hips (mean patient age 52.7 years, 46 to 68, 5 female, 7 male) with a pseudotumour and 23 hips (mean patient age 52.8 years, 38 to 69, 7 female, 16 male) without a pseudotumour. The mean post-operative time to the first MARS-MRI scan was 4.3 years (2.2 to 8.3), and mean time between the first and second MARS-MRI scan was eight months (6 to 12). At the second MRI scan, the grade of severity of the pseudotumour had not changed in 35 hips. One new pseudotumour (Anderson C2 score, moderate) was observed, and one pseudotumour was downgraded from C2 (moderate) to C1 (mild). In general, the characteristics of the pseudotumours hardly changed.

Repeated MARS-MRI scans within one year in patients with asymptomatic pseudotumours after MoM hip resurfacing showed little or no variation. In 23 patients without pseudotumour, one new asymptomatic pseudotumour was detected.

This is the first longitudinal study on the natural history of pseudotumours using MARS-MRI scans in hip resurfacing, and mirrors recent results for 28 mm diameter MoM total hip replacement.

Cite this article: Bone Joint J 2013;95-B:1626–31.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1209 - 1216
1 Sep 2013
van der Voort P Pijls BG Nouta KA Valstar ER Jacobs WCH Nelissen RGHH

Mobile-bearing (MB) total knee replacement (TKR) was introduced to reduce the risk of aseptic loosening and wear of polyethylene inserts. However, no consistent clinical advantages of mobile- over fixed-bearing (FB) TKR have been found. In this study we evaluated whether mobile bearings have an advantage over fixed bearings with regard to revision rates and clinical outcome scores. Furthermore, we determined which modifying variables affected the outcome.

A systematic search of the literature was conducted to collect clinical trials comparing MB and FB in primary TKR. The primary outcomes were revision rates for any reason, aseptic loosening and wear. Secondary outcomes included range of movement, Knee Society score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score and radiological parameters. Meta-regression techniques were used to explore factors modifying the observed effect.

Our search yielded 1827 publications, of which 41 studies met our inclusion criteria, comprising over 6000 TKRs. Meta-analyses showed no clinically relevant differences in terms of revision rates, clinical outcome scores or patient-reported outcome measures between MB and FB TKRs. It appears that theoretical assumptions of superiority of MB over FB TKR are not borne out in clinical practice.

Cite this article: Bone Joint J 2013;95-B:1209–16.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 718 - 718
1 May 2013
Pijls BG van der Heide HJL Nelissen RGHH

We welcome letters to the Editor concerning articles that have recently been published. Such letters will be subject to the usual stages of selection and editing; where appropriate the authors of the original article will be offered the opportunity to reply.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 523 - 529
1 Apr 2013
Henseler JF Nagels J van der Zwaal P Nelissen RGHH

Surgical repair of posterosuperior rotator cuff tears has a poorer outcome and a higher rate of failure compared with repairs of supraspinatus tears. In this prospective cohort study 28 consecutive patients with an irreparable posterosuperior rotator cuff tear after failed conservative or surgical treatment underwent teres major tendon transfer. Their mean age was 60 years (48 to 71) and the mean follow-up was 25 months (12 to 80). The mean active abduction improved from 79° (0° to 150°) pre-operatively to 105° (20° to 180°) post-operatively (p = 0.011). The mean active external rotation in 90° abduction improved from 25° (0° to 70°) pre-operatively to 55° (0° to 90°) post-operatively (p < 0.001). The mean Constant score improved from 43 (18 to 78) pre-operatively to 65 (30 to 86) post-operatively (p < 0.001). The median post-operative VAS (0 to 100) for pain decreased from 63 (0 to 96) pre-operatively to 5 (0 to 56) post-operatively (p < 0.001).

In conclusion, teres major transfer effectively restores function and relieves pain in patients with irreparable posterosuperior rotator cuff tears and leads to an overall clinical improvement in a relatively young and active patient group with limited treatment options.

Cite this article: Bone Joint J 2013;95-B:523–9.