The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1468 - 1474
1 Nov 2005
Steffen RT Smith SR Urban JPG McLardy-Smith P Beard DJ Gill HS Murray DW

We inserted an electrode up the femoral neck into the femoral head of ten patients undergoing a metal-on-metal hip resurfacing arthroplasty through a posterior surgical approach and measured the oxygen concentration during the operation. In every patient the blood flow was compromised during surgery, but the extent varied. In three patients, the oxygen concentration was zero at the end of the procedure. The surgical approach caused a mean 60% drop (p < 0.005) in oxygen concentration while component insertion led to a further 20% drop (p < 0.04). The oxygen concentration did not improve significantly on wound closure. This study demonstrates that during hip resurfacing arthroplasty, patients experience some compromise to their femoral head blood supply and some have complete disruption.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1493 - 1497
1 Nov 2005
Price AJ Short A Kellett C Beard D Gill H Pandit H Dodd CAF Murray DW

Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique.

In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks post-insertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year.

The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1488 - 1492
1 Nov 2005
Price AJ Dodd CAF Svard UGC Murray DW

We present a comparison of the results of the Oxford unicompartmental knee arthroplasty in patients younger and older than 60 years of age. The ten-year all-cause survival of the < 60 years of age group (52) was 91% (95% confidence interval (CI) 12), while in the ≥ 60 years of age group (512), the figure was 96% (95% CI 3). For the younger group, the mean Hospital for Special Surgery score at ten-year follow-up (n = 21) was 94 of 100, compared with a mean of 86 of 100 for the older group (n = 135). The results show that the Oxford unicompartmental arthroplasty can achieve ten-year results that are comparable to total knee arthroplasty in patients < 60 years of age. We conclude that for patients aged over 50, age should not be considered a contraindication for this procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 940 - 945
1 Jul 2005
Pandit H Ward T Hollinghurst D Beard DJ Gill HS Thomas NP Murray DW

Abnormal sagittal kinematics after total knee replacement (TKR) can adversely affect functional outcome. Two important determinants of knee kinematics are component geometry and the presence or absence of a posterior-stabilising mechanism (cam-post). We investigated the influence of these variables by comparing the kinematics of a TKR with a polyradial femur with a single radius design, both with and without a cam-post mechanism.

We assessed 55 patients, subdivided into four groups, who had undergone a TKR one year earlier by using an established fluoroscopy protocol in order to examine their kinematics in vivo. The kinematic profile was obtained by measuring the patellar tendon angle through the functional knee flexion range (0° to 90°) and the results compared with 14 normal knees. All designs of TKR had abnormal sagittal kinematics compared with the normal knee. There was a significant (p < 0.05) difference between those of the two TKRs near to full extension. The presence of the cam-post mechanism did not influence the kinematics for either TKR design. These differences suggest that surface geometry is a stronger determinant of kinematics than the presence or absence of a cam-post mechanism for these two designs. This may be because the cam-post mechanism is ineffective.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 921 - 927
1 Jul 2005
Glyn-Jones S Gill HS Beard DJ McLardy-Smith P Murray DW

Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability.

Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°, sd 0.42°) and internally rotate (mean 1.28°, sd 0.99°). The CPS-Plus showed no significant valgus rotation (mean 0.2°, sd 0.42°) or internal rotation (mean −0.03°, sd 0.75°). A wider, more rectangular cross-section improves rotational stability and may have a better long-term outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 320 - 323
1 Mar 2005
Little CP Ruiz AL Harding IJ McLardy-Smith P Gundle R Murray DW Athanasou NA

We present the histological findings of bone retrieved from beneath the femoral components of failed metal-on-metal hip resurfacing arthroplasties. Of a total of 377 patients who underwent resurfacing arthroplasty, 13 required revision; for fracture of the femoral neck in eight, loosening of a component in three and for other reasons in two. None of these cases had shown histological evidence of osteonecrosis in the femoral bone at the time of the initial implantation.

Bone from the remnant of the femoral head showed changes of osteonecrosis in all but one case at revision. In two cases of fracture which occurred within a week of implantation, the changes were compatible with early necrosis of the edge of the fracture. In the remaining six fractures, there were changes of established osteonecrosis. In all but one of the non-fracture cases, patchy osteonecrosis was seen.

We conclude that histological evidence of osteonecrosis is a common finding in failed resurfaced hips. Given that osteonecrosis is extensive in resurfaced femoral heads which fail by fracture, it is likely to play a role in the causation of these fractures.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 2 | Pages 172 - 176
1 Mar 2004
Glyn-Jones S Gill HS McLardy-Smith P Murray DW

The Birmingham hip resurfacing (BHR) arthroplasty is a metal-on-metal prosthesis for which no medium- or long-term results have been published. Despite this, it is increasing in popularity as an alternative to stemmed prostheses for younger patients. Since the fixation of the socket is conventional, the major concern is long-term failure of the femoral component. This can be predicted by the use of roentgen stereophotogrammetric analysis (RSA). We have therefore undertaken such a study of the BHR femoral component over a period of two years.

Twenty patients (22 hips) underwent a standard BHR procedure. Migration of the femoral component was measured by RSA at intervals of three, six, 12 and 24 months. At 24 months the total three-dimensional migration of the head was 0.2 mm. This was not statistically significant. Previous studies have shown that implants which loosen quickly have rapid early migration. Our results therefore suggest that the BHR femoral component is an inherently stable device which is likely to perform well in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 1 | Pages 62 - 67
1 Jan 2003
Price AJ Rees JL Beard D Juszczak E Carter S White S de Steiger R Dodd CAF Gibbons M McLardy-Smith P Goodfellow JW Murray DW

Before proceeding to longer-term studies, we have studied the early clinical results of a new mobile-bearing total knee prosthesis in comparison with an established fixed-bearing device. Patients requiring bilateral knee replacement consented to have their operations under one anaesthetic using one of each prosthesis. They also agreed to accept the random choice of knee (right or left) and to remain ignorant as to which side had which implant. Outcomes were measured using the American Knee Society Score (AKSS), the Oxford Knee Score (OKS), and determination of the range of movement and pain scores before and at one year after operation.

Preoperatively, there was no systematic difference between the right and left knees. One patient died in the perioperative period and one mobile-bearing prosthesis required early revision for dislocation of the meniscal component.

At one year the mean AKSS, OKS and pain scores for the new device were slightly better (p < 0.025) than those for the fixed-bearing device. There was no difference in the range of movement.

We believe that this is the first controlled, blinded trial to compare early function of a new knee prosthesis with that of a standard implant. It demonstrates a small but significant clinical advantage for the mobile-bearing design.


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 5 | Pages 653 - 657
1 Jul 2002
Robinson BJ Rees JL Price AJ Beard DJ Murray DW Smith PM Dodd CAF

When the Oxford unicompartmental meniscal bearing arthroplasty is used in the lateral compartment of the knee, 10% of the bearings dislocate. A radiological review was carried out to establish if dislocation was related to surgical technique.

The postoperative radiographs of 46 lateral unicompartmental arthroplasties were analysed. Five variables which related to the position and alignment of the components were measured. Dislocations occurred in six knees.

Only one of the five variables, the proximal tibial varus angle, had a statistically significant relationship to dislocation. This variable quantifies the height of the lateral joint line. The mean proximal tibial varus angle for knees the bearings of which had dislocated was 9° and for those which had not it was 5°. In both groups it was greater than would be expected in the normal knee (3°).

Our study suggests that a high proximal tibial varus angle is associated with dislocation. The surgical technique should be modified to account for this, with care being taken to avoid damage to or over-distraction of the lateral soft tissues.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 8 | Pages 1098 - 1103
1 Nov 2001
Beard DJ Murray DW Gill HS Price AJ Rees JL Alfaro-Adrián J Dodd CAF

We have assessed the effectiveness of reconstruction of the anterior cruciate ligament (ACL) in reducing functional tibial translation (TT).

The gait of 11 ACL-deficient patients was studied using Vicon equipment before and after surgery. Measurements of the angle between the patellar tendon and the long axis of the tibia were obtained in order to calculate TT in the sagittal plane relative to the uninjured limb during standing and walking.

Before surgery, patients did not show abnormal TT on the injured side, but after surgery significant anterior TT was found in the operated limb for every parameter of gait.

Abnormal anterior TT occurring during activity does not seem to be reduced by reconstruction; rather, it increases. It may be that the increased translation results from relaxation of excess contraction of the hamstring muscles, since compensatory muscle activity no longer is required in a reconstructed knee. The reduction of TT may not be an appropriate objective in surgery on the ACL.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 7 | Pages 1034 - 1036
1 Sep 2001
Rees JL Price AJ Lynskey TG Svärd UCG Dodd CAF Murray DW

Satisfactory selection criteria are essential for the successful outcome of unicompartmental knee arthroplasty (UCA). We report the frequency of revision of the Oxford medial unicompartmental arthroplasty in knees previously treated for anteromedial osteoarthritis by high tibial osteotomy (HTO). The combined results from three sources were used to allow statistical analysis of this uncommon subgroup.

In the combined series of 631 knees (507 patients) which had medial unicompartmental replacement, 613 were primary procedures and 18 were for a failed HTO. The mean follow-up times of the two groups were similar (5.8 years and 5.4 years, respectively). At review, 19 (3.1%) of the primary procedures and five (27.8%) of those undertaken for a failed HTO had been revised to total knee replacement. Survival analysis revealed the ten-year cumulative survivals to be 96% and 66%, respectively. The log-rank comparison of these survivals revealed a highly significant difference (p < 0.0001).

We recommend that the Oxford UCA should not be used in knees which have previously undergone an HTO.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 3 | Pages 377 - 383
1 Apr 2001
Portinaro NM Murray DW Benson MKD

The anatomy and development of the growing acetabulum are not clearly understood. We dissected and studied histologically two acetabula from the pelvis of a three-month-old infant. Relative rates of growth at the different growth plates were assessed by comparing the height of the proliferative layer with that of the hypertrophic layer.

The three bones which form the acetabulum are surrounded by growth plates on all sides except medially. These face towards the centre of the triradiate cartilage, the limbs of the triradiate cartilage and the articular surface and each may be divided into four distinct areas according to the orientation of its cell columns which reflect the direction of growth.

Growth was particularly rapid at the ischial growth plates directed towards the centre and the articular cartilage, and on both sides of the anterior limb of the triradiate cartilage. These findings may explain the mechanism by which the acetabulum changes orientation and inclination with growth.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 7 | Pages 996 - 1000
1 Sep 2000
Weale AE Murray DW Baines J Newman JH

Failure of a unicompartmental knee replacement (UKR) may be caused by progressive osteoarthritis of the knee and/or failure of the prosthesis. Limb alignment can influence both of these factors. We have examined the fate of the other compartments and measured changes in leg alignment after UKR.

A total of 50 UKRs was carried out on 45 carefully selected patients between 1989 and 1992. At operation, deliberate attempts were made to avoid overcorrection of the deformity. Four patients died, one patient was lost to follow-up and two knees were revised before review which was at a minimum of five years. Standard long-leg weight-bearing anteroposterior views of the knee and skyline views of the patellofemoral joint were taken before and at eight months and five years after operation. The radiographs of the remaining 43 knees were reviewed twice by blind and randomised assessment to measure the progression of osteoarthritis within the joints. Overcorrection of the deformity in the coronal plane was avoided in all but two knees. Only one showed evidence of progression of osteoarthritis within the patellofemoral joint, and this was only identified in one of the four assessments. Deterioration in the state of the opposite tibiofemoral compartment was not seen. Varus deformity tended to recur. Recurrent varus of 2° was observed between eight months and five years after operation. There was no correlation between the postoperative tibiofemoral angle and the extent of recurrent varus recorded at five years. Changes in alignment may be indicative of minor polyethylene wear or of subsidence of the tibial component.

The incidence of progressive osteoarthritis within the knee was very low after UKR. Patients should be carefully selected and overcorrection of the deformity be avoided.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 6 | Pages 892 - 900
1 Aug 2000
Neale SD Fujikawa Y Sabokbar A Gundle R Murray DW Graves SE Howie DW Athanasou NA

Mononuclear osteoclast precursors are present in the wear-particle-associated macrophage infiltrate found in the membrane surrounding loose implants. These cells are capable of differentiating into osteoclastic bone-resorbing cells when co-cultured with the rat osteoblast-like cell line, UMR 106, in the presence of 1,25(OH)2 vitamin D3. In order to develop an in vitro model of osteoclast differentiation which more closely parallels the cellular microenvironment at the bone-implant interface in situ, we determined whether osteoblast-like human bone-derived cells were capable of supporting the differentiation of osteoclasts from arthroplasty-derived cells and analysed the humoral conditions required for this to occur.

Long-term co-culture of arthroplasty-derived cells and human trabecular-bone-derived cells (HBDCs) resulted in the formation of numerous tartrate-resistant-acid-phosphatase (TRAP) and vitronectin-receptor (VNR)-positive multinucleated cells capable of extensive resorption of lacunar bone. The addition of 1,25(OH)2 vitamin D3 was not required for the formation of osteoclasts and bone resorption. During the formation there was release of substantial levels of M-CSF and PGE2. Exogenous PGE2 (10−8 to 10−6M) was found to stimulate strongly the resorption of osteoclastic bone. Our study has shown that HBDCs are capable of supporting the formation of osteoclasts from mononuclear phagocyte precursors present in the periprosthetic tissues surrounding a loose implant. The release of M-CSF and PGE2 by activated cells at the bone-implant interface may be important for the formation of osteoclasts at sites of pathological bone resorption associated with aseptic loosening.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 5 | Pages 790 - 795
1 Sep 1999
Weale AE Murray DW Newman JH Ackroyd CE

Patella infera may occur after reconstruction of the anterior cruciate ligament (ACL), high tibial osteotomy and total knee replacement (TKR). Restriction of movement of the knee and pain may result. Our aim was to compare the incidence and to assess the effects of patella infera after TKR and unicompartmental knee replacement (UKR).

We reviewed radiographs of the knees of 84 patients who had had either TKR or UKR as part of a randomised, controlled trial. The length of the patellar tendon was measured on serial radiographs taken before, at eight months and at five years after operation.

There was no significant change in the length of the patellar tendon after UKR, but a significant reduction was observed after TKR. Five years after the operation, the shortening of the tendon had increased to a mean of 3.5 mm. Of the knees with TKR reviewed at five years, 34% developed patella infera, defined as 10% or more of shortening, compared with 5% of those with UKR. Shortening was greatest in those knees which had required a lateral release; in this subgroup the mean shortening was 7.2 mm. Shortening correlated with restriction of movement and pain in the knee.

Our study has shown that patella infera develops in most patients after TKR with lateral release, and in approximately 25% of patients after TKR without this additional procedure. Patella infera rarely occurs after UKR. It is associated with restriction of movement and pain in the knee. It may be an effect of the more extensive exposure required to perform TKR and may, in part, explain the better clinical results of UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 5 | Pages 783 - 789
1 Sep 1999
Weale AE Murray DW Crawford R Psychoyios V Bonomo A Howell G O’Connor J Goodfellow JW

We determined the outcome of 56 ‘Oxford’ unicompartmental replacements performed for anteromedial osteoarthritis of the knee between 1982 and 1987. Of these, 24 were in patients who had died without revision, one was lost to follow-up and two had been revised. Of the remaining 29 knees, 26 were examined clinically and radiologically, two were only examined clinically and one patient was contacted by telephone. The mean age of the patients was 80.3 years.

At a mean follow-up of 11.4 years (10 to 14) the measurements of the knee score, range of movement and degree of deformity were not significantly different from those made one to two years after operation, except that the range of flexion had improved. Comparison of fluoroscopically-controlled radiographs at a similar interval of time showed no change in the appearance of the lateral compartments. The retained articular cartilage continued to function for ten or more years which suggests that anteromedial osteoarthritis may be considered as a focal disorder of the knee. This justifies continued efforts to develop methods of treatment which preserve intact joint structures.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 4 | Pages 742 - 742
1 Jul 1999
MURRAY DW


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 1 | Pages 130 - 134
1 Jan 1999
Alfaro-Adrián J Gill HS Murray DW

Studies using roentgen stereophotogrammetric analysis (RSA) have shown that the femoral components of cemented total hip replacements (THR) migrate distally relative to the bone, but it is not clear whether this occurs at the cement-implant or the cement-bone interface or within the cement mantle. Our aim was to determine where this migration occurred, since this has important implications for the way in which implants function and fail.

Using RSA we compared for two years the migration of the tip of the stem with that of the cement restrictor for two different designs of THR, the Exeter and Charnley Elite. We have assumed that if the cement restrictor migrates, then at least part of the cement mantle also migrates.

Our results have shown that the Exeter migrates distally three times faster than the Charnley Elite and at different interfaces. With the Exeter migration was at the cement-implant interface whereas with the Charnley Elite there was migration at both the cement-bone and the cement-implant interfaces.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 990 - 993
1 Nov 1998
Crawford R Sabokbar A Wulke A Murray DW Athanasou NA

We present a case in which the growth of an intraosseous cyst arising from the proximal tibiofibular joint appeared to have been increased by polyethylene wear particles from a medial unicompartmental knee replacement. Histological examination of the cyst wall showed a histiocytic response associated with numerous polyethylene wear particles. This case demonstrates that there is a direct communication between the joint cavity and the cyst. Such communication is probably through openings in the articular cartilage large enough to allow the passage of these particles.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 983 - 989
1 Nov 1998
Murray DW Goodfellow JW O’Connor JJ

Retrieval studies have shown that the use of fully congruent meniscal bearings reduces wear in knee replacements. We report the outcome of 143 knees with anteromedial osteoarthritis and normal anterior cruciate ligaments treated by unicompartmental arthroplasty using fully congruous mobile polyethylene bearings. At review, 34 knees were in patients who had died and 109 were in those who were still living. The mean elapsed time since operation was 7.6 years (maximum 13.8). We established the status of all but one knee.

There had been five revision operations giving a cumulative prosthetic survival rate at ten years (33 knees at risk) of 98% (95% CI 93% to 100%). Considering the knee lost to follow-up as a failure, the ‘worst-case’ survival rate was 97%. No failures were due to polyethylene wear or aseptic loosening of the tibial component. One bearing which dislocated at four years was reduced by closed manipulation.

The ten-year survival rate is the best of those reported for unicompartmental arthroplasty and not significantly different from the best rates for total knee replacement.