header advert
Results 1 - 14 of 14
Results per page:
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_2 | Pages 119 - 119
2 Jan 2024
Arthur L Min X Tu S Campi S Mellon S Murray D
Full Access

Tibial periprosthetic fracture is an important complication of the Oxford Unicompartmental Knee Replacement (OUKR). Primary fixation of cementless OUKR tibial components relies on the interference-fit of the ‘keel’ and a slot in the proximal tibia. Clinically used double blade keel saws (DKS) create slots with two grooves, generating stress concentrations where fractures may initiate. This study aimed to investigate slot factors that may influence incidence of tibial periprosthetic fractures.

Slots were made in PCF20 polyurethane foam using the DKS plus/minus adjuvant rasping, single blade keel saw (SKS), and rasp-only. Round and square slots were machined with milling cutters. Compact tensile tests were conducted per ASTM E399 to determine tensile load to fracture (TLTF) and results were validated using bovine tibia. Cementless OUKR components were implanted into slots in custom polyurethane blocks and compressed to failure to determine anatomical load to fracture (ALTF). A custom MATLAB program calculated slot roundness from cross-sectional images.

Round slots had higher TLTF (29.5N, SD=2.7) than square (25.2N, SD=1.7, p<0.05) and DKS slots (23.3N, SD=2.7, p<0.0001). Fractures occurred at the round slot apices, square slot corners, and deepest DKS slot grooves. ALTF was not significantly different between square and round slots. Adjuvant rasping made DKS slots significantly rounder, resulting in significantly higher TLTF, but rasping did not increase ALTF. ALTF was significantly higher for SKS (850N, SD=133, p<0.01) and rasp-only (912N, SD=100, p<0.001) slots compared to standard DKS slots (703N, SD=81).

Round keel slots minimise stress concentrations and increase TLTF but do not increase ALTF. The SKS and rasp-only slots retain material at slot ends and have significantly higher ALTF. Future studies should assess saw blades that retain material and round slot ends to evaluate if their use may significantly reduce the incidence of tibial periprosthetic fracture.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 16 - 16
7 Aug 2023
Arthur L Amin A Rahman A Tu SJ Mellon S Murray D
Full Access

Abstract

Introduction

Anecdotal reports suggest some cementless Oxford Unicompartmental Knee Replacements (OUKRs) have painful early subsidence of the tibial component with valgus rotation and/or posterior tilting. The incidence of subsidence and its association with pain is poorly understood. This radiographic study aimed to evaluate the incidence of tibial subsidence and five-year patient reported outcome measures (PROMs) of a cementless OUKR cohort.

Methodology

Cementless OUKRs from a high-volume centre with acceptable post-operative and five-year radiographs were included. Subsidence was determined by measuring the angle of the tibial tray/tibial axis angle and distance between the tibial tray axis and fibula head on anteroposterior and lateral radiographs using a custom MATLAB program. Analysis of 5-year PROMs assessed the relationship between subsidence and pain. Radiographs indicating tibial subsidence were validated by two observers.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 46 - 46
7 Aug 2023
Rahman A Heath D Mellon S Murray D
Full Access

Abstract

Introduction

In cementless UKR, early post-operative tibial fractures are 7x more common in very small tibias. A smaller keel has been shown to reduce this fracture risk, but its effect on fixation is unassessed. This mechanical study assesses the effect of keel interference and size on sagittal micromotion of the tibial component in physiological loading positions.

Method

A high-resolution Digital Image Correlation setup was developed and validated to an accuracy of 50 micrometres. Variants of tibial components were 3D-printed: standard, no-interference, no-keel, and a new small keel. Components were implanted into bone-analogue foam which was machined to a CT-reconstructed small tibia, using surgical technique. Tibias were loaded to 200N in physiological loading positions: 8mm (step-up) and 15mm (lunge) posterior to midpoint, and micromotion was assessed.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 17 - 17
7 Aug 2023
Arthur L Ghosh P Mohammad H Campi S Murray D Mellon S
Full Access

Abstract

Introduction

The Oxford Unicompartmental Knee Replacement's (OUKR's) fully-congruent design minimises polyethylene wear. Consequently, wear is a rare failure mechanism. Phase-3 OUKR linear wear at 5 years was higher than previous OUKR phases, but very low compared to fixed-bearing UKRs. This study aimed to measure OUKR bearing wear at 10 years and investigate factors that may affect wear.

Methodology

Bearing thickness for 39 OUKRs from a randomised study was calculated using radiostereometric analysis at regular intervals up to 10 years. Data for 39 and 29 OUKRs was available at 5 and 10 years, respectively. As creep occurs early, wear rate was calculated using linear regression between 6 months and 10 years. Relationships between wear and patient factors, fixation method, Oxford Knee Score (OKS), bearing position, and component position were analysed.


Abstract

Introduction

Cementless fixation of Oxford Unicompartmental Knee Replacements (UKRs) is an alternative to cemented fixation, however, it is unknown whether cementless fixation is as good long-term. This study aimed to compare primary and long-term fixation of cemented and cementless Oxford UKRs using radiostereometric analysis (RSA).

Methodology

Twenty-nine patients were randomised to receive cemented or cementless Oxford UKRs and followed for ten years. Differences in primary fixation and long-term fixation of the tibial components (inferred from 0/3/6-month and 6-month/1-year/2-year/5-year/10-year migration, respectively) were analysed using RSA and radiolucencies were assessed on radiographs. Migration rates were determined by linear regression and clinical outcomes measured using the Oxford Knee Score (OKS).


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_7 | Pages 85 - 85
1 Jul 2022
Rahman A Heath D Mellon S Murray D
Full Access

Abstract

Introduction

In cementless UKR, primary fixation of the tibial component is achieved by press-fitting a keel (i.e. with interference) into a vertical slot cut into the proximal tibia. This may adversely affect the structural integrity of surrounding bone. Early post-operative peri-prosthetic tibial fractures are 7x more common in very small knees, but the aetiology of these fractures is unknown - such sizes are rarely used in the UK but more common in Asian populations. This study explores the effect of keel-related features in fracture risk of these very small tibias.

Method

This in vitro study compares the effect of keel and slot depth (standard vs 33% shallower vs nil) and loading position (anterior/posterior gait range limits: mid-tibia vs 8mm posterior) on fracture load and path. 3D-printed titanium components were implanted using surgical instrumentation/technique, in bone-analogue foam machined to a CT-reconstructed very small tibia which subsequently experienced a peri-prosthetic fracture.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_7 | Pages 84 - 84
1 Jul 2022
Rahman A Dangas K Mellon S Murray D
Full Access

Abstract

Introduction

After remodelling, loss of bone density beside the keel of cementless UKR tibial components has been observed as a potential cause of concern. How this affects patient-reported outcomes, and further clinical implications, is unclear. This study aims to assess the effect of cementless UKR implantation on tibial bone density, and to explore its relationship to patient demographics and outcomes.

Method

This prospective study assesses 115 anterior-posterior radiographs from cementless UKR postoperatively and five years after surgery. Grey values from nine regions around each keel were collected and standardised to enable inter-radiograph comparison. Change between the post-operative and 5-year radiographs (indicating bone density) was calculated, and effect on 5-year patient demographics and pain and functional outcomes was assessed. Repeat measurements were performed by two operators to assess reliability.


Bone & Joint Research
Vol. 7, Issue 8 | Pages 511 - 516
1 Aug 2018
Beverly M Mellon S Kennedy JA Murray DW

Objectives

We studied subchondral intraosseous pressure (IOP) in an animal model during loading, and with vascular occlusion. We explored bone compartmentalization by saline injection.

Materials and Methods

Needles were placed in the femoral condyle and proximal tibia of five anaesthetized rabbits and connected to pressure recorders. The limb was loaded with and without proximal vascular occlusion. An additional subject had simultaneous triple recordings at the femoral head, femoral condyle and proximal tibia. In a further subject, saline injections at three sites were carried out in turn.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1347 - 1354
1 Oct 2016
Palan J Smith MC Gregg P Mellon S Kulkarni A Tucker K Blom AW Murray DW Pandit H

Aims

Periprosthetic fracture (PF) after primary total hip arthroplasty (THA) is an uncommon but potentially devastating complication. This study aims to investigate the influence of cemented stem designs on the risk of needing a revision for a PF.

Patients and Methods

We analysed data on 257 202 primary THAs with cemented stems and 390 linked first revisions for PF recorded in the National Joint Registry (NJR) of England, Wales and Northern Ireland to determine if a cemented femoral stem brand was associated with the risk of having revision for a PF after primary THA. All cemented femoral stem brands with more than 10 000 primary operations recorded in the NJR were identified. The four most commonly used cemented femoral stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem (n = 15 113) and Charnley (n = 20 182).

We compared the revision risk ratios due to PF amongst the stems using a Poisson regression model adjusting for patient factors. Compared with the Exeter V40, the age, gender and ASA grade adjusted revision rate ratio was 3.89 for the cemented CPT stem (95% confidence interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to 1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 49 - 49
1 Jan 2016
Monk A Mellon S Chen M Beard D Gill H Murray D
Full Access

Introduction

Knee arthroplasty is an effective intervention for painful arthritis when conservative measures have failed. Despite recent advances in component design and implantation techniques, a significant proportion of patients experience problems relating to the patella-femoral joint (PFJ).

Detailed knowledge of the shape and orientation of the normal and replaced femoral trochlea groove is critical when considering potential causes of anterior knee pain. Furthermore, to date it has proved difficult to establish a diagnosis due to shortcomings in current imaging techniques for obtaining satisfactory coronal plane motion data of the patella in the replaced knee.

The aim of this study was to correlate the trochlea shape of normal and replaced knees with corresponding coronal plane PFJ kinematic data.

Method

Bony and cartilagenous trochlea geometries from 3T MRI scans of 20 normal healthy volunteers were compared with both anatomical and standard total knee replacements (TKR) and patellofemoral joint replacement (PFJR) geometries. Following segmentation and standardized alignment, the path of the apex of the trochlea groove was measured using customized Matlab software. (Fig1).

Next, kinematic data of the 20 normal healthy volunteers (Normal) was compared with that of 20 TKR, and 20 PFJR patients using the validated MAUSTM system (Motion Analysis and UltraSound) comprising a 12-camera, motion capture system used to capture images of reflective markers mounted on subjects lower limbs and an ultrasound probe. A mapping between the ultrasound image and the motion capture system allows the ultrasound probe to be used to determine the locations of the patella relative to bony landmarks on the femur during a squat exercise.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 24 - 24
1 Nov 2015
Matharu G Mellon S Murray D Pandit H
Full Access

Introduction

This study aimed to: (1) compare published follow-up guidelines for metal-on-metal (MoM) hip patients and analyse protocols in relation to current evidence, and (2) assess the financial implications of these guidelines.

Methods

Follow-up guidance for MoM hips from five national authorities (MHRA in the UK; EFORT; United States FDA; Therapeutic Goods Administration of Australia; Health Canada) were contrasted and critically appraised. Using National Joint Registry (NJR) data (67,363 MoM hips implanted) the cost of annual surveillance for all MoM hips recorded in the NJR was calculated for each protocol.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_26 | Pages 10 - 10
1 Jun 2013
Monk A Chen M Mellon S Gibbons M Beard D Murray D Gill H
Full Access

Previous attempts to measure coronal plane patellofemoral kinematics following knee replacement have suffered from methodological drawbacks; the patella being obscured by the components, metal artefact and technical inaccuracies. The aim of this study was to assess whether there was any significant difference in the patellofemoral kinematics between normal, TKR and PFJR patients using the validated MAUS™ technique (combining motion analysis with ultrasound).

60 patients were recruited into three groups; normal healthy volunteers (Normal), TKR, and PFJR patients. The MAUS technique incorporates a 12 camera analysis system (providing gross alignment data for tibial and femoral segments) and an ultrasound probe (providing coordinates of bony landmarks on patella femur and tibia) during a squat exercise. 6 DOF kinematics were described between 0 and 90° flexion. The validated accuracy of the MAUS technique registering the ultrasound images within the motion capture system is 1.84 mm (2 × SD).

Movements of the Normal group were significantly different from the TKR group (p=0.03) and the PFJR group (p<0.01), whilst there was no significant difference between the TKR and PFJR groups (p=0.27).

Our data suggest that many aspects of patellofemoral kinematics are absent following TKR and PFJR, which could be addressed in future designs of knee TKR and PFJR.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 456 - 456
1 Sep 2012
Grammatopoulos G Pandit H Mellon S Glyn-Jones S Gundle R Mclardy-Smith P Murray D Gill H
Full Access

INTRODUCTION

Studies have suggested that there is a reduction in head-neck-ratio (HNR) associated with MoMHRA. A reduction in HNR at operation would decrease range of movement and increase impingement risk. Impingement could lead to 20 edge loading, increasing wear. Serum ion levels of Chromium (Cr) and Cobalt (Co) are surrogate markers of wear. Although acetabular component orientation has been shown to contribute to wear and PT development, the role of a decrease in HNR has only been highlighted in PT development. This study aimed to measure changes in HNR that occur at resurfacing and determine any gender- and component size-specific differences. In addition it aimed to determine whether changes in HNR could be associated with increased wear.

METHODS

84 patients (56M: 28F) with unilateral MoMHRA were included. The mean age at surgery was 57 years. The mean femoral component was 49mm. Components were considered small if <45mm, average if between 45–50mm and large if >50mm. Three designs were implanted; BHR, C+ and Recap. The average follow up was 4 years. All patients had Cr/Co levels measured at follow up. Patients were considered to have high ions if Cr and Co levels were 5.1ppb and 4.4ppb respectively.

Pre-operative HNR (HNRpre) and the post-operative HNR (HNRpost) were made from the respective pelvic radiographs. Assuming a 2mm thick cartilage layer, the HNR based on the diameter of the articular cartilage pre-operatively (HNRart) was calculated.

The immediate changes in HNR as a result of the operation were expressed relative to articular HNR pre-op:

HNRartpost=HNRpost–HNRart


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 171 - 171
1 May 2011
Kwon Y Mellon S Murray D Gill H
Full Access

Introduction: Edge-loading, a phenomenon whereby the femoral component comes into contact with the edge of the acetabular component, has been suggested to increase wear in metal-on-metal hip resurfacing arthroplasty (MoMHRA). Pseudotumours (soft-tissue mass relating to the hip joint) have been associated with elevated serum and hip aspirate metal ion levels. This study aimed to investigate in vivo edge-loading in MoMHRA patients with pseudotumours by quantifying dynamic loci of the hip joint segment force relative to the acetabular component during functional activities.

Materials and Methods: A total of 21 MoMHRA patients (30 hips) in two groups were investigated in this Ethics approved study:

6 patients with pseudo-tumours detected using ultrasound/MRI;

15 patients without pseudotumours.

Three-dimensional lower limb motion analysis (12 camera Vicon System) was performed to estimate hip joint segment force during walking, chair-rising and stair-climbing. CT scans were used to determine each patient’s specific hip joint centre and acetabular component orientation. Edge-loading was defined to occur when a hip joint segment force vector/ cup intersection was located within 10% of the cup radius from the edge of the cup. Serum cobalt and chromium levels were analysed using Inductively-Coupled Plasma Spectrometer.

Results: Edge-loading in the pseudotumour group occurred with significantly (p=0.02) longer (4-fold increase) duration as well as greater magnitude (7-fold increase) of force, compared to the non-pseudotumour group. The duration and force of the edge-loading were activity-dependent, with proportionally greater difference observed during stair climbing. The acetabular cup orientation values in the pseudotumour group were found within the safe zone of Lewinnek in one third of the hips with the remaining two thirds outside the safe zone. The presence of pseudotumour was associated with:

significantly higher median serum cobalt levels: 14.3ug/l (range 10.6–64.1) vs. 1.9ug/l (range 1.2–5.0), p< 0.001;

significantly higher median serum chromium levels: 21.2ug/l (range 13.8–45.2) vs. 1.8ug/l (range 0.7–7.6), p< 0.001.

Discussion: Edge-loading in MoMHRA patients with pseudotumours occurred in vivo with significantly longer duration and greater magnitude of force impulse compared to the patients with a well functioning MoMHRA during activities of daily living. This suggests that edge-loading may be an important mechanism that leads to localised high wear, with subsequent elevation of metal ion levels in MoMHRA patients with pseudotumours. Although the acetabular component malposition, such as increase in both inclination and anteversion angles, appears to be an important factor in edge-loading, the aetiology of edge-loading is likely to be multi-factorial.