header advert
Results 1 - 2 of 2
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 129 - 137
1 Jun 2020
Knowlton CB Lundberg HJ Wimmer MA Jacobs JJ

Aims

A retrospective longitudinal study was conducted to compare directly volumetric wear of retrieved polyethylene inserts to predicted volumetric wear modelled from individual gait mechanics of total knee arthroplasty (TKA) patients.

Methods

In total, 11 retrieved polyethylene tibial inserts were matched with gait analysis testing performed on those patients. Volumetric wear on the articular surfaces was measured using a laser coordinate measure machine and autonomous reconstruction. Knee kinematics and kinetics from individual gait trials drove computational models to calculate medial and lateral tibiofemoral contact paths and forces. Sliding distance along the contact path, normal forces and implantation time were used as inputs to Archard’s equation of wear to predict volumetric wear from gait mechanics. Measured and modelled wear were compared for each component.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 48 - 48
1 Oct 2019
Jacobs JJ Knowlton CB Lundberg HJ Wimmer MA
Full Access

Introduction

Studies of retrieved total knee replacement (TKR) components demonstrate that in vivo wear on the articular surface of polyethylene liners exhibits a much higher variability than their in vitro counterparts tested on simulators.1 Only one study has attempted to validate a patient-specific model of wear with a clinically retrieved component.2 The purpose of this study is to investigate the relationship between observed TKR contact conditions during gait and measured volume loss on retrieved tibial components.

Methods

Eleven retrieved ultra-high molecular weight polyethylene (UHMWPE) cruciate-retaining tibial liner components from ten separate patients (implantation time = 8.6±5.6 years) had matching gait trials of normal level walking for each knee. Volume loss on retrieved components was calculated using a coordinate measuring machine and autonomous reconstruction.3 Motion analysis of normal level walking gait had been conducted between 1986 and 2005 for various previous studies and stored in a consented Human Mechanics Repository, ranging from pre-operative to long-term post-operative testing. Contact location between the femoral component and the tibial component on the medial and lateral plateaus were calculated throughout stance.4 A previously validated and fine-tuned parametric numerical model was used to calculate TKR contact forces for each gait trial.5 Vertical contact forces and contact paths on the medial and lateral plateaus were input as normal force and sliding distance to a simplified Archard equation for wear with material wear constant = 2.42 × 10−7 mm3/Nm2,6 to compute average wear per gait cycle. Wear rates were calculated using linear regression, and Pearson correlation examined correlations between modeled and measured wear.