header advert
Results 1 - 7 of 7
Results per page:
Bone & Joint Open
Vol. 2, Issue 8 | Pages 599 - 610
1 Aug 2021
Hothi H Bergiers S Henckel J Iliadis AD Goodier WD Wright J Skinner J Calder P Hart AJ

Aims

The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices.

Methods

We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions.


Bone & Joint Research
Vol. 10, Issue 6 | Pages 348 - 350
1 Jun 2021
Skinner JA Sabah SA Hart AJ


Bone & Joint Research
Vol. 8, Issue 3 | Pages 136 - 145
1 Mar 2019
Cerquiglini A Henckel J Hothi H Allen P Lewis J Eskelinen A Skinner J Hirschmann MT Hart AJ

Objectives

The Attune total knee arthroplasty (TKA) has been used in over 600 000 patients worldwide. Registry data show good clinical outcome; however, concerns over the cement-tibial interface have been reported. We used retrieval analysis to give further insight into this controversial topic.

Methods

We examined 12 titanium (Ti) PFC Sigma implants, eight cobalt-chromium (CoCr) PFC Sigma implants, eight cobalt-chromium PFC Sigma rotating platform (RP) implants, and 11 Attune implants. We used a peer-reviewed digital imaging method to quantify the amount of cement attached to the backside of each tibial tray. We then measured: 1) the size of tibial tray thickness, tray projections, peripheral lips, and undercuts; and 2) surface roughness (Ra) on the backside and keel of the trays. Statistical analyses were performed to investigate differences between the two designs.


Bone & Joint Research
Vol. 6, Issue 5 | Pages 345 - 350
1 May 2017
Di Laura A Hothi H Henckel J Swiatkowska I Liow MHL Kwon Y Skinner JA Hart AJ

Objectives

The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single stem design.

Methods

This retrieval study involved 48 total hip arthroplasties (THAs) with CoCr V40 trunnions paired with either CoCr (n = 21) or ceramic (n = 27) heads. The taper junction of all hips was evaluated for fretting/corrosion damage and volumetric material loss using a roundness-measuring machine. We used linear regression analysis to investigate taper damage differences after adjusting for potential confounding variables.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 10 - 18
1 Jan 2015
Sabah SA Henckel J Cook E Whittaker R Hothi H Pappas Y Blunn G Skinner JA Hart AJ

Arthroplasty registries are important for the surveillance of joint replacements and the evaluation of outcome. Independent validation of registry data ensures high quality. The ability for orthopaedic implant retrieval centres to validate registry data is not known. We analysed data from the National Joint Registry for England, Wales and Northern Ireland (NJR) for primary metal-on-metal hip arthroplasties performed between 2003 and 2013. Records were linked to the London Implant Retrieval Centre (RC) for validation. A total of 67 045 procedures on the NJR and 782 revised pairs of components from the RC were included. We were able to link 476 procedures (60.9%) recorded with the RC to the NJR successfully. However, 306 procedures (39.1%) could not be linked. The outcome recorded by the NJR (as either revised, unrevised or death) for a primary procedure was incorrect in 79 linked cases (16.6%). The rate of registry-retrieval linkage and correct assignment of outcome code improved over time. The rates of error for component reference numbers on the NJR were as follows: femoral head category number 14/229 (5.0%); femoral head batch number 13/232 (5.3%); acetabular component category number 2/293 (0.7%) and acetabular component batch number 24/347 (6.5%).

Registry-retrieval linkage provided a novel means for the validation of data, particularly for component fields. This study suggests that NJR reports may underestimate rates of revision for many types of metal-on-metal hip replacement. This is topical given the increasing scope for NJR data. We recommend a system for continuous independent evaluation of the quality and validity of NJR data.

Cite this article: Bone Joint J 2015;97-B:10–18.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 400 - 400
1 Jul 2010
Hart AJ Bandi A Maggiore P Skinner JA Underwood R Cann P
Full Access

Data on retrieval analysis of current generation metal on metal hip replacements is scarce. Such analysis may help to reduce the incidence of failure and revision procedures. Our aim was to investigate the wear characteristics of explanted (ie failed) metal on metal (MOM) acetabular components in terms of; 1) wear rate; and 2) distribution of the wear (specifically edge loading).

30 hips were collected from 20 centres. The types of prostheses were: 15 BHR; 10 Cormet and 5 ASR. Wear of the acetabular components of the prostheses was measured using an out of roundness (Rondcom 60A) machine. We recorded the implantation and removal date of each hip.

The median linear wear rate was 7.32μm/year; this is at least 3 times greater than steady state wear rates reported for similar components worn in hip simulator studies. For 24 out of 30 cups, the greatest linear wear was recorded at the cup edge.

Failed metal-on-metal acetabular components were associated with higher than expected wear rates. The highest wear was seen closest to the cup edge in the majority of patients suggesting edge loading had occurred and probably explained the high wear rates. Accurate cup placement (to avoid edge loading) may reduce the failure of MOM hips.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 399 - 400
1 Sep 2009
Brust K Khanduja V Dandachli W Iranpour F Henckel J Hart AJ Cobb J
Full Access

Radiological measurements are an essential component of the assessment of outcome following knee arthroplasty. However, plain radiographic techniques can be associated with significant projectional errors because they are a two-dimensional (2D) representation of a three-dimensional (3D) structure. Angles that are considered within the target zone on one film may be outside that zone on other films. Moreover, these parameters can be subject to significant inter-observer differences when measured. The aim of our study therefore was to quantify the variability between observers evaluating plain radiographs following Unicompartmental knee arthroplasty.

Twenty-three observers, made up of Orthopaedic Consultants and trainees, were asked to measure the coronal and sagittal alignment of the tibial and femoral components from the post-operative long-leg plain radiograph of a Unicompartmental knee arthroplasty. A post-operative CT scan using the low dose Imperial knee protocol was obtained as well and analysed with 3D reconstruction software to measure the true values of these parameters. The accuracy and spread of the pain radiographic measurements were then compared with the values obtained on the CT.

On the femoral side, the mean angle in coronal alignment was 1.5° varus (Range 3.8, SD 1, min 0.1, max 3.9), whereas the mean angle in sagittal alignment was 8.6° of flexion (Range 7.5, SD 1.5, Min 3.7, Max 11.2). The true values measured with CT were 2.4° and 11.0° respectively. As for the tibial component, the mean coronal alignment angle was 89.7° (Range 11.6, SD 3.3, Min 83.8, Max 95.4), and the mean posterior slope was 2.4° (Range 8.7, SD 1.6, Min -2, Max 6.7). The CT values for these were 87.6° and 2.7° respectively.

We conclude that the plain radiographic measurements had a large scatter evidenced by the wide ranges in the values obtained by the different observers. If only the means are compared, the plain radiographic values were comparable with the true values obtained with CT (that is; accuracy was good) with differences ranging from 0.3° to 2.4°. The lack of precision can be avoided with the use of CT, particularly with the advent of low-dose scanning protocols.