header advert
Results 1 - 10 of 10
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 6 | Pages 747 - 757
1 Jun 2013
Jameson SS Baker PN Mason J Rymaszewska M Gregg PJ Deehan DJ Reed MR

The popularity of cementless total hip replacement (THR) has surpassed cemented THR in England and Wales. This retrospective cohort study records survival time to revision following primary cementless THR with the most common combination (accounting for almost a third of all cementless THRs), and explores risk factors independently associated with failure, using data from the National Joint Registry for England and Wales. Patients with osteoarthritis who had a DePuy Corail/Pinnacle THR implanted between the establishment of the registry in 2003 and 31 December 2010 were included within analyses. There were 35 386 procedures. Cox proportional hazard models were used to analyse the extent to which the risk of revision was related to patient, surgeon and implant covariates. The overall rate of revision at five years was 2.4% (99% confidence interval 2.02 to 2.79). In the final adjusted model, we found that the risk of revision was significantly higher in patients receiving metal-on-metal (MoM: hazard ratio (HR) 1.93, p < 0.001) and ceramic-on-ceramic bearings (CoC: HR 1.55, p = 0.003) compared with the best performing bearing (metal-on-polyethylene). The risk of revision was also greater for smaller femoral stems (sizes 8 to 10: HR 1.82, p < 0.001) compared with mid-range sizes. In a secondary analysis of only patients where body mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m2 significantly increased the risk of revision (HR 1.55, p = 0.002). The influence of the bearing on the risk of revision remained significant (MoM: HR 2.19, p < 0.001; CoC: HR 2.09, p = 0.001). The risk of revision was independent of age, gender, head size and offset, shell, liner and stem type, and surgeon characteristics.

We found significant differences in failure between bearing surfaces and femoral stem size after adjustment for a range of covariates in a large cohort of single-brand cementless THRs. In this study of procedures performed since 2003, hard bearings had significantly higher rates of revision, but we found no evidence that head size had an effect. Patient characteristics, such as BMI and American Society of Anesthesiologists grade, also influence the survival of cementless components.

Cite this article: Bone Joint J 2013;95-B:747–57.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1611 - 1617
1 Dec 2012
Jameson SS Baker PN Mason J Gregg PJ Brewster N Deehan DJ Reed MR

Despite excellent results, the use of cemented total hip replacement (THR) is declining. This retrospective cohort study records survival time to revision following primary cemented THR using the most common combination of components that accounted for almost a quarter of all cemented THRs, exploring risk factors independently associated with failure. All patients with osteoarthritis who had an Exeter V40/Contemporary THR (Stryker) implanted before 31 December 2010 and recorded in the National Joint Registry for England and Wales were included in the analysis. Cox’s proportional hazard models were used to analyse the extent to which risk of revision was related to patient, surgeon and implant covariates, with a significance threshold of p < 0.01. A total of 34 721 THRs were included in the study. The overall seven-year rate of revision for any reason was 1.70% (99% confidence interval (CI) 1.28 to 2.12). In the final adjusted model the risk of revision was significantly higher in THRs with the Contemporary hooded component (hazard ratio (HR) 1.88, p < 0.001) than with the flanged version, and in smaller head sizes (< 28 mm) compared with 28 mm diameter heads (HR 1.50, p = 0.005). The seven-year revision rate was 1.16% (99% CI 0.69 to 1.63) with a 28 mm diameter head and flanged component. The overall risk of revision was independent of age, gender, American Society of Anesthesiologists grade, body mass index, surgeon volume, surgical approach, brand of cement/presence of antibiotic, femoral head material (stainless steel/alumina) and stem taper size/offset. However, the risk of revision for dislocation was significantly higher with a ‘plus’ offset head (HR 2.05, p = 0.003) and a hooded acetabular component (HR 2.34, p < 0.001).

In summary, we found that there were significant differences in failure between different designs of acetabular component and sizes of femoral head after adjustment for a range of covariates.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1293 - 1298
1 Oct 2006
Allami MK Fender D Khaw FM Sandher DR Esler C Harper WM Gregg PJ

Using the Trent regional arthroplasty register, we analysed the survival at ten years of 1198 consecutive Charnley total hip replacements carried out across a single health region of the United Kingdom in 1990. At ten years, information regarding outcome was available for 1001 hips (83.6%). The crude revision rate was 6.2% (62 of 1001) and the cumulative survival rate with revision of the components as an end-point was 93.1%. At five years, a review of this series of patients identified gross radiological failure in 25 total hip replacements which had previously been unrecognised. At ten years the outcome was known for 18 of these 25 patients (72%), of whom 13 had not undergone revision.

This is the first study to assess the survival at ten years for the primary Charnley total hip replacement performed in a broad cross-section of hospitals in the United Kingdom, as opposed to specialist centres. Our results highlight the importance of the arthroplasty register in identifying the long-term outcome of hip prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 716 - 720
1 Jun 2006
Hardoon SL Lewsey JD Gregg PJ Reeves BC van der Meulen JHP

New brands of joint prosthesis are released for general implantation with limited evidence of their long-term performance in patients. The CUSUM continuous monitoring method is a statistical testing procedure which could be used to provide prospective evaluation of brands as soon as implantation in patients begins and give early warning of poor performance. We describe the CUSUM and illustrate the potential value of this monitoring tool by applying it retrospectively to the 3M Capital Hip experience.

The results show that if the clinical data and methodology had been available, the CUSUM would have given an alert to the underperformance of this prosthesis almost four years before the issue of a Hazard Notice by the Medical Devices Agency. This indicates that the CUSUM can be a valuable tool in monitoring joint prostheses, subject to timely and complete collection of data. Regional or national joint registries provide an opportunity for future centralised, continuous monitoring of all hip and knee prostheses using these techniques.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 2 | Pages 187 - 190
1 Mar 2003
Fender D van der Meulen JHP Gregg PJ

Using a regional arthroplasty register, we assessed the outcome, at five years, of 1198 primary Charnley total hip replacements (THRs) undertaken across a single health region in England in 1990. An independent clinical and radiological assessment was completed for 497 operations, carried out in 18 different hospitals, under the care of 56 consultants and by differing grades of surgeon. The overall number of failures in this group was 44 (8.9%). We found that the risk of failure in patients operated on by a consultant whose firm carried out 60 or more THRs in 1990 was 25% of that of patients under the care of a consultant whose firm undertook less than 30, adjusting for a number of patient, surgeon and hospital characteristics (16% v 4%; p < 0.001 for linear trend).

Our study shows that the early outcome of hip replacement surgery varies with the number of replacements undertaken by the consultant firm. A national arthroplasty register would be a convenient source for such data.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 6 | Pages 838 - 842
1 Aug 2001
Chambers IR Fender D McCaskie AW Reeves BC Gregg PJ

The radiological features of the cement mantle around total hip replacements (THRs) have been used to assess aseptic loosening. In this case-control study we investigated the risk of failure of THR as predictable by a range of such features using data from patients recruited to the Trent Regional Arthroplasty Study (TRAS).

An independent radiological assessment was undertaken on Charnley THRs with aseptic loosening within five years of surgery and on a control group from the TRAS database. Chi-squared tests were used to test the probability of obtaining the observed data by chance, and odds ratios were calculated to estimate the strength of association for different features.

Several features were associated with a clinically important increase (> twofold) in the risk of loosening, which was statistically significant for four features (p < 0.01). Inadequate cementation (Barrack C and D grades) was the most significant feature, with an estimated odds ratio of 9.5 (95% confidence interval 3.2 to 28.4, p < 0.0001) for failure.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 7 | Pages 944 - 947
1 Sep 2000
Fender D Harper WM Gregg PJ

We have assessed the usefulness of a regional hip register in the assessment of the outcome of primary total hip replacement (THR). Over 97% of THRs performed in the Trent region in 1990 were captured onto the register and the inaccuracies recorded were less than 1.8%. In an independent assessment of 2111 patients five years after THR, 85.9% of those available for assessment responded, and 66.8% agreed to an assessment.

The cost of this independent assessment at five years, utilising a regional hip register, was approximately £50 per implant. This is a reasonable outlay to identify problems early. Some form of registration and outcome assessment should be performed on a national level.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 4 | Pages 577 - 581
1 Jul 1999
Fender D Harper WM Gregg PJ

Using a regional arthroplasty register we assessed the outcome at five years of 1198 primary Charnley total hip replacements (THRs) carried out in 1152 patients across a single UK health region in 1990. Information regarding outcome was available for 1080 hips (90%) and 499 had an independent clinical and radiological assessment.

By five years the known rate of aseptic loosening was 2.3%, of deep infection 1.4%, of dislocation 5.0% and of revision 3.2%. The radiological assessment of 499 THRs revealed gross failure in a further 5.2%, which had been previously unrecognised. The combined rate of failure of nearly 9% is higher than those published from specialist centres and surgeons, but is probably more representative of the norm.

Our study supports the need for a national register and surveillance of THRs. It emphasises that all implants should be followed, and suggests that the results of such surgery, when performed in the general setting, may not be as good as expected.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 4 | Pages 611 - 613
1 Jul 1998
Clarke MT Green JS Harper WM Gregg PJ

A total of 110 total knee replacements (TKRs) was randomised to receive either a cemented or an uncemented prosthesis. Postoperative venography at five to seven days was used to compare the prevalence, site and size of deep-vein thrombosis (DVT). We also compared the findings with those of postoperative venography in a group of patients with cemented total hip replacements (THRs).

The total prevalence of DVT was significantly greater after uncemented (81%) than after cemented TKR (55%). Both knee groups had a significantly higher prevalence of DVT than in cemented hip replacements (32%). We found no difference in the proportion with proximal DVT in the three groups (14%, 15% and 16%). The median length of the thrombi was significantly greater after cemented (26.5 cm) than after uncemented TKR (11 cm) or after cemented THR (7 cm). This difference was mainly due to greater lengths of distal rather than proximal thrombi.

We conclude that the use of cement may affect the formation of DVT after joint replacement, but does not appear to lead to an increased incidence.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 2 | Pages 191 - 194
1 Mar 1996
McCaskie AW Brown AR Thompson JR Gregg PJ

Three radiological methods are commonly used to assess the outcome of total hip replacement (THR). They aim to record the appearance of lucent areas and migration of the prosthesis in a reproducible manner. Two of them were designed to monitor the implant through time and one to grade the quality of cementing. We have measured the level of inter- and intraobserver agreement in all three systems.

We randomised 30 patients to receive either finger packing or retrograde gun cementing during Charnley hip replacements. The postoperative departmental radiographs were evaluated in a blinded study by two orthopaedic trainees, two consultants and two experts in THR. The trainees and consultants repeated the exercise at least two weeks later. We used the unweighted kappa statistic to establish the levels of agreement.

In general, intraobserver agreement was moderate but interobserver agreement was poor, with levels similar to or less than those expected by chance. Our results indicate that such systems cannot provide reliable data from centres in different parts of the world, with various levels of surgeon evaluating radiographs at differing time intervals. We discuss the problem and suggest some methods of improvement.