Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 18 - 18
1 Jan 2016
Brockett C Smyth A Day M Harris N Partridge S Williams S Fisher J
Full Access

Introduction

Total ankle replacement (TAR) has been used as a surgical intervention for arthritis since the 1970s. However, unlike clinically successful hip and knee replacements, TARs are renowned for extensive contraindications to surgery and high failure rates with an average of 83% survival at 5 years. The majority cite aseptic loosening as the reason for failure. The aim of this study wais to analyse retrieved TARs visually and through interferometry to identify potential the failure mechanisms associated with these devices.

Methods

Retrieved total ankle replacements (n=11) from consecutive revision surgeries carried out at Chapel Allerton Hospital, Leeds between August 2012 and January 2014, were collected for study at the University of Leeds, under an NRES approved procedure (09/H1307/60).

The bearing surfaces of the samples were visually inspected for evidence of damage and wear. The bearing surfaces between the tibial component and the flat surface of the polyethylene insert were then examined using a scanning white light interferometer (NP Flex, Bruker, USA). It was not possible to characterise the talar bearing surface or the inferior polyethylene surface at this stage through interferometry due to the curvature of the surface. The components were aligned and five sections on each of the surfaces measured. These sections represented; anterior-medial, anterior-lateral, posterior-medial, posterior lateral and central regions of the bearing surfaces. 3D roughness values were recorded, and the mean 3D surface roughness compared between implants. Measurements were taken on the medial and lateral aspects of the bearing surfaces to investigate whether damage was location specific. A coefficient of determination was calculated to assess the relationship between implantation time and surface roughness.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 100 - 100
1 Jan 2016
Cowie R Briscoe A Fisher J Jennings L
Full Access

Introduction

UHMWPE articulating against PEEK-OPTIMA® has the potential for use as a novel bearing couple in joint arthroplasty due to its potentially low wear rates and the bioinertness of its wear debris. The aim of this study was to investigate the role of protein in the lubricant on the wear of UHMWPE articulating against PEEK at both room and physiological temperature.

Methods

The wear of GUR1020 UHMWPE pins articulating against PEEK plates (Ra ∼0.06µm) was compared to highly polished cobalt chrome plates (Ra <0.01µm) in a 6-station multi-axial pin-on-plate rig using kinematics to replicate those in total knee arthroplasty. Tests were carried out at either ∼20°C or ∼36°C and wear was investigated under varying concentrations of bovine serum (0, 25 or 90%). Studies were carried out for 1 Million cycles with wear of the UHMWPE pins assessed gravimetrically using unloaded soak controls to compensate for moisture uptake.

Statistical analysis was carried out using ANOVA with significance taken at p<0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 26 - 26
1 Jan 2016
Hammouche S Fisher J Tipper J Williams S
Full Access

Introduction

Hip replacements are falling short of matching the life expectancy of coxarthritis patients, due to implanting THR in younger patients and due to increasingly active patients. The most frequently implanted hip prostheses use cross linked (XL) polyethylene (PE) on metal bearings in the USA and most of the Western world. Concerns remain in the long term around the potential of wear debris-induced aseptic loosening. Thus exploring lower-wearing alternative bearings remains a major research goal.

PEEK (poly-ether-ether-ketone) is a thermoplastic polymer with enhanced mechanical properties. This study compared the wear of PEEK to the wear of cross linked polyethylene, when sliding against cobalt chrome (CoCr) metallic counterfaces, and compared the wear of carbon-fibre reinforced (CFR)-PEEK to cross linked polyethylene when sliding against metallic and ceramic counterfaces under different contact stresses within the hip joint.

Methods

The following materials were studied: unfilled PEEK (OPTIMA, Invibio) and CFR-PEEK (MOTIS, Invibio) against either high carbon (HC) CoCr or Biolox Delta ceramic plates. The comparative control material was a moderately cross-linked PE (Marathon, DePuy Synthes).

A simple geometry wear study was undertaken. A rotational motion of ±30° across a sliding distance of ±28 mm (cross shear of 0.087), and contact pressures of 1.6 or 4 MPa were applied. The lubricant was 25% (v/v) bovine serum and the wear test was conducted for 1 million cycles at 1 Hz. Wear was assessed gravimetrically. A validated soak control method was used to adjust for serum absorption-induced mass changes during the wear test. Surface profilometry was assessed pre and post wear test.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 99 - 99
1 Jan 2016
Cowie R Carbone S Aiken S Cooper J Fisher J Jennings L
Full Access

Introduction

When third body particles originating from bone cement or bone void fillers become trapped between articulating surfaces of joint replacements, contact surfaces may be damaged leading to accelerated wear and premature failure of the implant. In this study, the damage to cobalt chrome counterfaces by third body particles from PMMA bone cement (GMV, DePuy) and various bone void fillers was investigated; then wear tests of UHMWPE were carried out against these surfaces.

Methods

Third body particles of polymerised GMV bone cement and the bone void fillers; OsteoSet (with tobramycin), Stimulan and Stimulan+ (with vancomycin and tobramycin) (provided by Biocomposites Ltd.) were trapped between an UHMWPE pin and a highly polished cobalt chrome plate. A load of 120N was applied to the pin and using an Instron materials testing machine, the plate was pulled beneath the pin to recreate third body damage [1]. The resulting surface topography of the plate was analysed using white light interferometry (Bruker NPFLEX). Pin on plate wear tests of GUR 1020 UHMWPE pins were carried out against the plates perpendicular to the direction of damage for 500,000 cycles in 25% bovine serum using a 6-station multi-axial reciprocating rig under conditions to replicate the kinematics in total knee replacement. Wear of the pins was determined by gravimetric analysis and results were compared to negative (highly polished) control plates and positive controls scratched with a diamond stylus (lip height 2µm). Statistical analysis was carried out using one-way ANOVA with significance taken at p<0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 77 - 77
1 Jan 2016
Al-Hajjar M Williams S Jennings L Thompson J Isaac G Ingham E Fisher J
Full Access

Introduction

Increased wear rates [1, 2] and acetabular rim fracture [3] of hip replacement bearings reported clinically have been associated with edge loading, which could occur due to rotational and/or translational mal-positioning [4]. Surgical mal-positioning can lead to dynamic microseparation mechanisms resulting in edge loading conditions. In vitro microseparation conditions have replicated stripe wear and the bi-modal wear debris distribution observed clinically [5, 6]. The aim of this study was to investigate the effect of steep cup inclination, representing rotational mal-positioning, on the magnitude of dynamic microseparation, severity of edge loading, and the resulting wear rate of a ceramic-on-ceramic bearing, under surgical translational mal-positioning conditions.

Materials and Methods

Ceramic-on-ceramic bearings where the ceramic liner was inserted into a titanium alloy cup (BIOLOX® delta and Pinnacle® respectively, DePuy Synthes, UK) were tested on the six-station Leeds II hip simulator. The first test was run with the cups inclined at an angle equivalent, clinically, to 45° (n=6) and the second test was run with the cups inclined at an angle equivalent, clinically, to 65° (n=6). A standard gait cycle was run. A fixed surgical translational mal-positioning of 4mm between the centres of rotations of the head and the cup in the medial/lateral axis was applied on all stations. Both tests ran for three million cycles each. The lubricant used was 25% new-born calf serum. Wear was assessed gravimetrically using a microbalance (XP205, Mettler Toledo, UK) and geometrically using a coordinate measuring machine (CMM, Legex 322, Mitutoyo, UK). Statistical analysis was done using one way ANOVA with significance taken at p<0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 78 - 78
1 Jan 2016
Brockett C Carbone S Fisher J Jennings L
Full Access

Introduction

Wear debris induced osteolysis and loosening continue to cause clinical failure in total knee replacement (TKR). To improve longevity and reduce wear alternative materials have been examined. Carbon-fibre-reinforced poly–ether-ether-ketone (CFR-PEEK) has shown promising results in wear studies [1–2].

The aim of this study was to explore the use of CFR-PEEK and PEEK as alternative bearing materials for polyethylene in TKR through experimental knee joint wear simulation.

Methods

Two novel materials were studied as an alternative to polyethylene as the tibial bearing surface in a TKR configuration using a Cobalt chrome femoral bearing in current clinical use. Six right Sigma CR fixed bearing TKRs (DePuy Synthes, UK) were paired with either PEEK or CFR-PEEK custom-made flat inserts (Invibio, UK) in a Prosim knee simulator (Simulator Solutions, UK). The tibial inserts were 14mm thick, to give an equivalent thickness to existing insert designs. A flat geometry was selected as this has previously been shown to yield low wear in polyethylene bearings [3]. The tests were conducted under High Kinematics, with anterior-posterior and internal-external displacement control [4]. Tests were conducted for three million cycles, lubricated with 25% bovine serum, with wear assessed gravimetrically at 1 and 3Mc.

Images of the wear scars were recorded at completion of the study. Visual inspection of the tibial inserts was used to identify regions of wear damage for SEM analysis (EVO MA15-Smart SEM, Zeiss, Germany)


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 7 - 7
1 Jul 2014
Brockett C Carbone S Jennings L Fisher J
Full Access

Summary Statement

Wear of total knee replacement (TKR) is a clinical concern. This study demonstrated low-conformity moderately cross-linked-polyethylene fixed bearing TKRs showed lower volumetric wear than conventional-polyethylene curved fixed bearing TKRs highlighting potential improvement in TKR performance through design and material selection.

Introduction

Wear of total knee replacement (TKR) continues to be a significant factor in the clinical performance of the implants. Historically, failure due to delamination and fatigue directed implant design towards more conforming implants to reduce contact stress. However, the new generations of more oxidatively-stable polyethylene have improved the long-term mechanical properties of the material, and therefore allowed more flexibility in the bearing design. The purpose of this study was to investigate the effect of insert conformity and material on the wear performance of a fixed bearing total knee replacement through experimental simulation.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 144 - 144
1 Jul 2014
Al-Hajjar M Fisher J Hardaker C Kurring G Isaac G Williams S
Full Access

Summary Statement

The frictional torque of ceramic-on-ceramic bearings tended to increase with increasing the bearings size (32, 48, 56mm). However, the frictional torque was significantly lower than that measured on metal-on-metal bearings under well positioned and well lubricated conditions.

Introduction

Larger head size in total hip replacement theoretically provides increased range of motion and enhanced stability. However, there are potential clinical concerns regarding increased frictional torques with large diameter metal-on-metal bearings causing loosening of the acetabular cups and corrosion at the taper. The aim of this study was to determine the frictional torques of large diameter BIOLOX® delta ceramic-on-ceramic bearings.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 214 - 214
1 Dec 2013
Abdelgaied A Brockett C Liu F Jennings L Jin Z Fisher J
Full Access

Introduction:

Backside wear has been previously reported through in-vitro and in-vivo to have a significant contribution to the total wear in rotating bearing TKRs.

The present study investigated the contribution of backside wear to the total wear in the PFC Sigma rotating platform mobile bearing TKR. In addition, the wear results were compared to the computed wear rates of the PFC Sigma fixed bearing TKR, with two different bearing materials.

Materials and Methods:

The commercially available PFC Sigma rotating platform mobile bearing and PFC Sigma fixed bearing total knee replacements, size 3 (DePuy, UK) were tested, with either conventional or moderately cross-linked (5 MRad) GUR1020 UHMWPE bearing materials. The computational wear model for the knee implants was based on the contact area and an independent experimentally determined non-dimensional wear coefficient [1,2,3].

The experimental wear test for the mobile bearing was force controlled using the ISO anterior-posterior force (ISO14243-1-2009). However, due to time limitation of the explicit simulation required to run the force controlled model, the simulation was run using the AP displacements taken from the experimental knee simulator which was run under the ISO AP force. The Sigma fixed bearing TKR was run under high level of anterior-posterior displacements (maximum of 10 mm).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 2 - 2
1 Dec 2013
Al-Hajjar M Clarkson P Williams S Jennings L Thompson J Fisher J
Full Access

Introduction

Stripe wear, observed on retrieved ceramic hip replacements, has only been replicated in vitro under translational mal-positioning conditions where the centres of rotation of the head and the cup are mismatched1,2; an in vitro condition termed “microseparation”.

The aim of this study was to compare the edge loading mechanisms observed under microseparation conditions due to translational mal-positioning conditions simulated on two different hip joint simulators.

Materials and Methods

The components used in this study were zirconia-toughened-alumina ceramic-on-ceramic bearings (36 mm) inserted into titanium alloy acetabular cups (BIOLOX® delta and Pinnacle® respectively, DePuy Synthes Joint Reconstruction, Leeds, UK). Six couples were tested for two million cycles under 0.5 mm dynamic microseparation conditions on the Leeds II hip joint simulator as described by Nevelos et al2 and Stewart et al3 (Figure 1). Ten bearing couples were tested for two million cycles under microseparation conditions achieved in two different ways on the ProSim pneumatic hip joint simulator (SimSol, Stockport, UK). Two conditions were tested; condition (1)- the femoral head was left to completely separate (the vertical motion was controlled at 1 mm) causing it to contact the inferior rim of the acetabular cup before edge loading on the superior rim at heel strike (n = 5) and condition (2)- springs were placed below the plate holding the femoral head to control the tilt of the head laterally towards the rim of the acetabular cup as the negative pressure was applied (n = 5; Figure 1). Wear was assessed gravimetrically every million cycles using a microbalance (Mettler AT201, UK). Three-dimensional reconstructions of the wear area on the heads were obtained using a coordinate measuring machine (Legex 322, Mitutoyo, UK) and SR3D software (Tribosol, UK).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 90 - 90
1 Dec 2013
Brockett C Abdelgaied A Hardaker C Fisher J Jennings L
Full Access

Introduction

Wear debris induced osteolysis and loosening continue to be causes of clinical failure in total knee replacement (TKR). Laboratory simulation aims to predict the wear of TKR bearings under specific loading and motion conditions. However, the conditions applied may have significant influence on the study outcomes (1)

The aim of this study was to examine the influence of femoral setup and kinematic inputs on the wear of a conventional polyethylene fixed bearing TKR through experimental and computational models.

Methods

Six right Sigma CR fixed bearing TKRs (DePuy Synthes, Leeds, UK) with curved polyethylene inserts (GVF, GUR1020 UHMWPE) were tested in Prosim knee simulator (Simulator Solutions, UK). The femoral bearing was set up with the centre of rotation (CoR) on either on the distal radius of the implant (Distal CoR), as indicated by the device design, or according to the ISO specification (ISO CoR; ISO14243-3). The tests were conducted under ‘High Kinematics’ (2). It was necessary to reverse the direction of the anterior-posterior displacement for the tests conducted with the ISO centre of rotation to maintain the contact region within the insert surface (Reverse High Kinematics). Tests were conducted for three million cycles, lubricated with 25% bovine serum, with wear assessed gravimetrically.

The computational wear model for the TKR was based on the contact area and an independent experimentally determined non-dimensional wear coefficient, previously validated against the experimental data (3).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 249 - 249
1 Mar 2013
Maiti R Fisher J Jin Z Rowley L Jennings L
Full Access

Introduction

Patella femoral joint bearings in total knee replacements have shown low wear (3.1 mm3/MC) under standard gait simulator conditions1. However, the wear in retrieval studies have shown large variations between 1.3 to 45.2 mm3/year2. Previous in vitro studies on the tibial femoral joint have shown wear is dependent on design, materials and kinematics3.

The aim of this study was to investigate the influence of the design (geometry) and shape on the wear rate of patella femoral joints in total knee replacements.

Materials and Methods

The Leeds/Prosim knee simulator was used to investigate the wear of two types of commercially available patellae. The PFC Sigma cobalt chrome femoral component was coupled with 2 types of patellae buttons: round and oval dome. The UHMWPE was the same for the both types – GUR1020 GVF (gamma irradiated in vacuum and foiled packed). 25% bovine serum was used as the lubricant. The test were carried out at three conditions – high medial lateral (ML) rotations (<4°) and uncontrolled ML displacement (<4 mm), low ML rotation (<1°) and uncontrolled ML displacement (<4 mm); the physiological gait cycle; and low ML rotation and controlled ML displacement (<1.5 mm). In this abstract the two designs were tested in physiological gait condition (Figure 1). Patella ML displacement and tilt were passively controlled and measured after every 300,000 cycles. A ligament resisting force equivalent to 10 N4 was applied on the lateral side of the patella to avoid patella slip.

Five samples of each design were tested for 3 million cycles at a cycle rate of 1 Hz. The wear volume was obtained gravimetrically every million cycles and presented with 95% confidence limits. Statistical significance was taken at p<0.05.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 139 - 139
1 Mar 2013
Brockett C Hardaker C Fisher J Jennings L
Full Access

Introduction

Wear of polyethylene continues to be a significant factor in the longevity of total knee replacement (TKR). Moderately cross-linked polyethylene has been employed to reduce the wear of knee prostheses, and more recently anti-oxidants have been introduced to improve the long-term stability of the polyethylene material. This is the initial study of the wear of a new anti-oxidant polyethylene and a new TKR design, which has modified femoral condylar geometry.

Materials and Methods

The wear of a new TKR the Attune knee was investigated using a physiological six station Prosim knee wear simulator (Simulator Solutions, UK). Six mid-size Attune fixed bearing cruciate retaining TKRs (DePuy Inc, Warsaw, USA) were tested for a period of 6 million cycles. The inserts were manufactured from AOX™, a compression moulded GUR1020 polyethylene incorporating Covernox™ solid anti-oxidant. The AOX polymer was irradiated to 8M Rad, to give a moderately cross-linked material.

High and intermediate kinematics, under anterior-posterior displacement control were used for this study (McEwen et al 2005). The maximum femoral axial loading was 2600N, with flexion-extension of 0 to 58°, an anterior-posterior displacement of 0–10 mm for high kinematics and 0–5 mm for intermediate, and an internal-external rotation of ±5°. The lubricant was 25% (v/v) calf serum supplemented with 0.03% (v/v) sodium azide solution in deionised water, as an antibacterial agent, and was changed approximately every 0.33 Mc. Wear was assessed gravimetrically and moisture uptake was assessed using unloaded soak controls. Volumetric wear was calculated using a density of 0.94 mg/mm3, and compared with a previous study examining the Sigma XLK TKR design which uses moderately crosslinked polyethylene which is in current clinical use (Brockett et al 2012).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 200 - 200
1 Sep 2012
Williams S Isaac G Fisher J
Full Access

INTRODUCTION

Ceramic-on-metal hip replacements (COM, where the head is a Biolox Delta ceramic and liner is Co Cr alloy), have demonstrated reduced wear under standard conditions in vitro compared to metal-on-metal (MOM) [1]. Early clinical results are also encouraging [2]. Recently concerns have been raised regarding the poor clinical performance of MOM hip resurfacings [3], particularly when cups are steeply inclined. Laboratory hip simulator testing has been used to replicate edge loading, also demonstrating elevated wear [4]. Therefore, a range of conditions to replicate sub-optimal use clinically to better predict in vivo performance should be used. The aim of this study was to compare the wear rates of MOM and COM under adverse edge loading conditions in an in vitro hip simulator test.

METHODS

Ceramic-on-metal (n=3) and metal-on-metal (n=3) 36mm hip prostheses (supplied by DePuy International Ltd, UK) were tested in the Leeds Physiological Anatomical Hip Joint Simulator. Liners were mounted to provide a clinical angle of 45o, and stems positioned anatomically. A simplified gait cycle and microseparation was applied as previously described [5] for two million cycles in 25% new born calf serum. Gravimetric analysis was completed every million cycles and wear volumes calculated.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 22 - 22
1 Sep 2012
Brockett C John G Williams S Isaac G Fisher J
Full Access

Introduction

Concerns regarding UHMWPE wear particle induced osteolysis in total hip replacement (THR, [1]) have led to alternative materials to be sought. Carbon-fibre reinforced poly-ether-ether-ketone (CFR-PEEK) has shown reduced wear in hip and knee configurations compared with conventional polyethylene [2-4]. The aim of this study was to investigate the wear performance of a ceramic-on-CFR PEEK THR through a simulator study.

Methodology

Five 36mm diameter Biolox Delta heads were paired with extruded CFR-PEEK cups and tested in a hip wear simulator (Simulator Solutions, UK) for 10 million cycles (Mc). Tests were performed in a Prosim hip simulator, which applied a twin peak loading cycle, with a peak load of 3kN. Flexion-extension of −15 to +30 degrees was applied to the head and internal-external rotation of +/− 10 degrees was applied to the cup, components were mounted anatomically. The lubricant was 25% (v/v) calf serum supplemented with 0.03% (w/v) sodium azide. Wear was assessed gravimetrically at several intervals adjusted for moisture uptake using loaded and unloaded soak controls.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 1 - 1
1 Sep 2012
Al-Hajjar M Fisher J Tipper J Williams S Jennings L
Full Access

INTRODUCTION

Ceramic-on-ceramic hip replacements have generated great interest in recent years due to substantial improvements in manufacturing techniques and material properties1. Microseparation conditions that could occur due to several clinical factors such as head offset deficiency, medialised cup combined with laxity of soft tissue resulting in a translation malalignment, have been shown to cause edge loading, replicate clinically relevant wear mechanisms2,3 and increase the wear of ceramic-on-ceramic bearings3,4. The aim of this study was to investigate the influence of increasing the femoral head size on the wear of ceramic-on-ceramic bearings under several clinically relevant simulator conditions.

MATERIALS AND METHODS

The wear of size 28mm and 36mm ceramic-on-ceramic bearings (BIOLOX® Delta, CeramTec, Germany) was determined under different in vitro conditions using the Leeds II hip simulator. For each size bearing, two clinical cup inclination angles were considered, 55° (n=3) and 65° (n=3) for the 28mm bearing and 45° (n=3) and 65° (n=3) for the 36mm bearing. The first two (28mm study) or three (36mm study) million cycles ran under standard gait conditions and a subsequent three million cycles ran under microseparation conditions. A standard gait cycle included a twin peak load (300N–3000N), extension/flexion (−15°/+30°) and internal/external rotation (±10°). Microseparation3 was achieved by applying a 0.4–0.5mm medial displacement to the cup relative to the head during the swing phase of the standard gait cycle resulting in edge loading at heel strike. The lubricant was 25% (v/v) new-born calf serum which was changed approximately every 333,000 cycles. The wear volume was ascertained through gravimetric analysis every million cycles. One way ANOVA was performed (significance: p<0.05), and 95% confidence limits were calculated.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 107 - 107
1 Sep 2012
Hadley M Hadfield F Hardaker C Isaac G Fisher J Wye J Barnett J
Full Access

Introduction

Hip wear simulation is a widely used technique for the pre-clinical evaluation of new bearing designs. However, wear rates seen in vitro can often be significantly different to those seen clinically. This can be attributed to the difference between the optimal conditions in a simulator and wide ranging conditions in real patients.

This study aimed to develop more clinically relevant simulator tests, looking specifically at the effects of cup inclination angle (in vivo) and stop-dwell-start (SDS) protocols on a clinically available product.

Method

Five tests using a Paul type walking cycle (ISO 14242) were carried out on two ProSim hip simulators:

28mm MoM, standard walking, cup inclination 45°, (n = 5)

36mm MoM, standard walking, cup 45°, (n = 4)

36mm MOM, SDS: 10 walking cycles and pause of 5s with stance load of 1250N cup 45°, (n = 5)

36mm MOM, SDS: 10 walking cycles and pause of 30s with stance load of 1250N, (n = 5) cup 45°

36mm MOM, standard walking, cup 55°(n = 5), and 65°(n = 5).

All samples had matched clearances, measured using a CMM (Prismo Navigator, Zeiss, Germany). Wear was measured gravimetrically (Sartorius ME235S: 0.01mg).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 2 - 2
1 Sep 2012
Al-Hajjar M Fisher J Williams S Tipper J Jennings L
Full Access

INTRODUCTION

Retrieval and clinical studies of metal-on-metal (MoM) bearings have associated increased wear1 and elevated patient ion levels2 with steep cup inclination angles and edge loading conditions. The University of Leeds have previously developed a hip simulator method that has been validated against retrievals and shown to replicate clinically relevant wear rates and wear mechanisms3,4. This method involves introducing lateral microseparation to represent adverse joint laxity and offset deficiency. This study aimed to investigate the effect of microseparation representing translational malpostion, and increased cup inclination angle, representing rotational malposition, in isolation and combined on the wear of different sizes (28 and 36mm) MoM bearing in total hip replacement (THRs).

MATERIALS AND METHODS

The wear of size 28mm and 36mm MoM THRs bearings was determined under different in vitro conditions using the Leeds II hip simulator. For each size bearing, two clinical cup inclination angles were considered, 45° (n=3) and 65° (n=3). The first three million cycles were run under standard gait conditions and subsequently three million cycles were run under microseparation conditions. Standard gait cycles included a twin peak load (300N–3000N), extension/flexion (−15°/+30°) and internal/external rotation (±10°). Microseparation4 was achieved by applying a 0.4–0.5mm medial displacement to the cup relative to the head during the swing phase of the standard gait cycle resulting in edge loading at heel strike. The lubricant was 25% (v/v) new-born calf serum. The wear volume was determined through gravimetric analysis every million cycles. One way ANOVA was performed (significance: p<0.05), and 95% confidence limits were calculated.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 25 - 25
1 Aug 2012
AL-Hajjar M Fisher J Williams S Tipper J Jennings L
Full Access

In vitro the introduction of microseparation and edge loading to hip simulator gait cycle has replicated clinically relevant wear rates and wear mechanisms in ceramic-on-ceramic bearings[1], and elevated the wear rates of MoM surface replacements (SR) to levels similar to those observed in retrievals[2]. The aim was to assess the wear of two different sized MoM total hip replacement bearings under steep cup inclination angles and adverse microseparation and edge loading conditions.

Two tests were performed on the Leeds II hip joint simulator using two different size bearings (28mm and 36mm). Cups were mounted to provide inclination angles of 45 degrees (n=3) and 65 degrees (n=3). The first three million cycles were under standard gait conditions. Microseparation and edge loading conditions as described by Nevelos et al[1] were introduced to the gait cycle for the subsequent three million cycles. The lubricant was 25% new born calf serum. The mean wear rates and 95% confidence limits were determined and statistical analysis was performed using One Way ANOVA.

Under standard gait conditions, when the cup inclination angle increased from 45 degrees to 65 degrees, the wear of size 28mm bearing significantly (p=0.004) increased by 2.7-fold, however, the larger bearings did not show any increase in wear (p=0.9). The introduction of microseparation conditions resulted in a significant (p=0.0001) increase in wear rates for both bearing sizes under both cup inclination angle conditions. Under microseparation conditions, the increase in cup inclination angle had no influence on the wear rate for both bearing sizes (Figure 1).

With larger bearings, head-rim contact occurs at a steeper cup inclination angle providing an advantage over smaller bearings. The introduction of edge loading and microseparation conditions resulted in a significant increase in wear rates for both bearing sizes. The wear rates obtained in this study under combined increased cup inclination angle and microseparation were half of those obtained when SR MoM bearings were tested under similar adverse conditions[2]. This study shows the importance of prosthesis design and accurate surgical positioning of the head and acetabular cup in MoM THRs.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 46 - 46
1 Aug 2012
Taylor S Tsiridis E Ingham E Jin Z Fisher J Williams S
Full Access

Tribology and wear of articular cartilage is associated with the mechanical properties, which are governed by the extracellular matrix (ECM). The ECM adapts to resist the loads and motions applied to the tissue. Most investigations take cartilage samples from quadrupeds, where the loading and motions are different to human. However, very few studies have investigated the differences between human and animal femoral head geometry and the mechanical properties of cartilage.

This study assessed the differences between human, porcine, ovine and bovine cartilage from the femoral head; in terms of anatomical geometry, thickness, equilibrium elastic modulus and permeability.

Diameter of porcine (3-6 months old), bovine (18-24 months old), ovine (4 years old) and human femoral heads were measured (n=6). Plugs taken out of the superior region of each femoral head and creep indentation was performed. The human femoral heads were obtained from surgery due to femoral neck fracture. Cartilage thickness was measured by monitoring the resistive force change as a needle traversed the cartilage and bone at a constant feed rate using a mechanical testing machine. The percentage deformation over time was determined by dividing deformation by thickness. A biphasic finite element model was used to obtain the intrinsic material properties of each plug. Data is presented as the mean ± 95% confidence limits. One-way ANOVA was used to test for significant differences (p < or = 0.05).

Significant differences in average femoral head diameter were observed between all animals, where bovine showed the largest femoral head. Human cartilage was found to be significantly thicker than cartilage from all quadrupedal hips. Human cartilage had a significantly larger equilibrium elastic modulus compared to porcine and bovine cartilage. Porcine articular cartilage was measured to be the most permeable which was significantly larger than all the other species. No significant difference in permeability was observed between human and the other two animals: bovine and ovine (Table 1).

The current study has shown that articular cartilage mechanical properties, thickness and geometry of the femoral heads differ significantly between different species. Therefore, it is necessary to consider these variations when choosing animal tissue to represent human.