header advert
Results 1 - 1 of 1
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1416 - 1421
1 Oct 2010
Qureshi F Draviaraj KP Stanley D

Between September 1993 and September 1996, we performed 34 Kudo 5 total elbow replacements in 31 rheumatoid patients. All 22 surviving patients were reviewed at a mean of 11.9 years (10 to 14). Their mean age was 56 years (37 to 78) at the time of operation. All had Larsen grade IV or V rheumatoid changes on X-ray. Nine (three bilateral replacements and six unilateral) had died from unrelated causes. One who had died before ten years underwent revision for dislocation.

Of the 22 total elbow replacements reviewed six had required revision, four for aseptic loosening (one humeral and three ulnar) and two for infection. Post-operatively, one patient had neuropraxia of the ulnar nerve and one of the radial nerve. Two patients had valgus tilting of the ulnar component.

With revision as the endpoint, the mean survival time for the prosthesis was 11.3 years (95% confidence interval (10 to 13) and the estimated survival of the prosthesis at 12 years according to Kaplan-Meier survival analysis was 74% (95% confidence interval 0.53 to 0.91).

Of the 16 surviving implants, ten were free from pain, four had mild pain and two moderate. The mean arc of flexion/extension of the elbow was 106° (65° to 130°) with pronation/supination of 90° (30° to 150°) with the joint at 90° of flexion. The mean Mayo elbow performance score was 82 (60 to 100) with five excellent, ten good and one fair result.

Good long-term results can be expected using the Kudo 5 total elbow replacement in patients with rheumatoid disease, with a low incidence of loosening of the components.