Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 31 - 31
1 Mar 2021
Hopwood J Redmond A Chapman G Richards L Collins S Brockett C
Full Access

Abstract

Objectives

Implant loosening remains a common cause of total ankle replacement (TAR) revision, and has been associated with wear-mediated osteolysis. Limited pre-clinical studies for TARs have been reported and the variety of experiment settings make it difficult to compare wear rates. Factors such as simulator control mechanism; whether pneumatic or electromechanical, may influence the integrity of the simulator outputs with respect to input profiles. This study compares the wear of a TAR, tested in electromechanical and pneumatic experimental simulators under identical input conditions.

Methods

Twelve medium BOX® (MatOrtho Ltd) TARs (n=6 for each simulator) were tested in an electromechanical and pneumatic knee simulator (Simulation Solutions, UK) for 3 million cycles (Mc). Standard ‘Leeds’ displacement-controlled inputs were used. Kinematic performance was investigated by comparing the output profiles against the maximum demanded input values. The lubricant used was 25% new-born calf serum and wear was determined gravimetrically.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 16 - 16
1 Feb 2021
Wade A Beadling A Neville A De Villiers D Collins S Bryant M
Full Access

The vast majority of total hip replacements (THR) implanted today enable modularity by means of a tapered junction; based on the Morse taper design introduced for cutting tools in the 19th Century 1. Morse-type tapers at the head-stem junction provide many benefits, key for a successful surgical outcome such as wider component selection and restoration of better biomechanics 2. However, moving from mono-block to modular designs has not been without its issues. Fluid ingress and motion at the interface has led to a complex multifactorial degradation mechanism better known as fretting-corrosion 3. Fretting-corrosion products created at the junction are commonly associated with adverse local tissue reactions 4.

There is a wide variation in the taper junction of THR differing quite significantly from Morse's original design. Performance of the taper junction has been found to vary with different designs 5,6. However, there is still a lack of common understanding of what design inputs makes a ‘good’ modular taper interface. The aim of this study was to better understand the links between implant design and fretting-corrosion initially focussing on the role of angular mismatch between male and female taper. A combination of experimental approaches with the aid of computational models to assist understanding has been adopted. A more descriptive understanding between taper design, engagement, motion and fretting-corrosion will be developed.

Three different sample designs were created to represent the maximum range of possible angular mismatches seen in clinically available THR modular tapers (Matched: 0.020 ±0.002 °, Proximal: 0.127 ±0.016 °, Distal: −0.090 ±0.002 °). Head-stem components were assembled at 2 kN. Motion and fretting-corrosion at the interface was simulated under incremental uniaxial sinusoidal loading between 0.5–4 kN at 8 intervals of 600 cycles. The different types of motions at the interface was measured using a developed inductance circuit composed of four sensing coils, digital inductance converter chip (LDC1614, Texas Instruments, US) and microcontroller (myRIO, National Instruments, US). Fretting-corrosion was measured using potentiostatic electrochemical techniques with an over potential of +100 mV vs OCP (Ivium, NL). Complimentary finite element (FE) models were created in Ansys (Ansys 19.2, US).

Under uniaxial loading, the ‘matched’ modular taper assemblies corroded most and allowed the greatest pistoning motion due to a seating action. ‘Distal’ and ‘proximal’ engaged modular tapers showed reduced corrosion and seating when compare to the ‘matched’ components. However the kinetics of corrosion and motion were interface dependent. It is hypothesized, and complimented by FEA analysis, that lower initial contact stress in the ‘matched’ modular tapers allows for greater subsidence and depassivation of the oxide layer and higher corrosion. ‘Matched’ modular tapers allowed less rotational and toggling motions compared to mismatched tapers, suggesting a reduced mismatch might perform better once the heads have seated over time. Future work involves tests conducted under a surgically relevant impaction force and physiological loading kinematics to develop this descriptive link between taper design, engagement and performance.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 64 - 64
1 Feb 2020
Hopwood J Redmond A Chapman G Richards L Collins S Brockett C
Full Access

Background

Total ankle arthroplasty (TAA) is an alternative to ankle arthrodesis, replacing the degenerated joint with a mechanical motion-preserving alternative. Implant loosening remains a primary cause of TAA revision, and has been associated with wear-mediated osteolysis. Differing implant designs have a major influence on the wear performance of joint replacements. Providing a range of implant sizes allows surgeons a greater intra-operative choice for varying patient anatomy and potential to minimise wear. Minimal pre-clinical testing exists in the literature that investigates the effect of implant size on the wear behaviour. The aim of this study therefore was to investigate the effect of two different implant sizes on the wear performance of a TAA.

Materials & Methods

Six ‘medium’ and six ‘extra small’ BOX® (MatOrtho Ltd, UK) TAA implants, of the same conceptual design and polyethylene insert thickness, were tested in a modified 6 station pneumatic knee simulator. 5 million cycles (Mc) of wear simulation were completed for each implant size, under kinematics aiming to replicate an ankle gait cycle (Figure 1) [1]. The simulator used had six degrees of freedom, of which four were controlled. The maximum axial load was 3150N, equivalent to 4.5 times body weight of a 70kg individual. The flexion profile ranged from −15° plantarflexion to 15° dorsiflexion. Rotation about the tibial component ranged from −2.3° of internal rotation to 8° external rotation, and anterior/posterior (AP) displacement ranged from 3.1 mm anterior to −0.9 mm posterior displacement. The lubricant used was 25% bovine serum supplemented with 0.04% sodium azide to prevent bacterial degradation. The wear of the TAA polyethylene inserts were determined gravimetrically after each Mc, with unloaded soak controls used to compensate for the uptake of moisture by the polyethylene.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 24 - 24
1 Feb 2020
De Villiers D Collins S
Full Access

INTRODUCTION

Ceramic-on-ceramic hip resurfacing offers a bone conserving treatment for more active patients without the potential metal ion risks associated with resurfacing devices. The Biolox Delta ceramic material has over 15 years of clinical history with low wear and good biocompatibility but has been limited previously in total hip replacement to 48mm diameter bearings [1]. Further increasing the diameter for resurfacing bearings and removing the metal shell to allow for direct fixation of the ceramic cup may increase the wear of this material and increase the risk of fracture.

METHODS

Eighteen implants (ReCerf™, MatOrtho, UK; Figure1) were wear tested; six were ⊘40mm (small) and twelve ⊘64mm (large). All small and six large implants were tested under ISO 14242 standard conditions for 5 million cycles (mc) at 30° inclination (45° clinically). The six remaining large implants were tested under microseparation conditions in which rim contact was initiated during heel strike of the gait cycle for 5mc. Cups were orientated at 45° inclination (60° clinically) to allow for separation of the head and cup with a reduced 50N swing phase load and a spring load applied to induce a 0.5mm medial-superior translation of the cup. Wear was determined gravimetrically at 0.5mc, 1mc and every mc after.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 25 - 25
1 Feb 2020
De Villiers D Collins S Taylor A Dickinson A
Full Access

INTRODUCTION

Hip resurfacing offers a more bone conserving solution than total hip replacement (THR) but currently has limited clinical indications related to some poor design concepts and metal ion related issues. Other materials are currently being investigated based on their successful clinical history in THR such as Zirconia Toughened Alumina (ZTA, Biolox Delta, CeramTec, Germany) which has shown low wear rates and good biocompatibility but has previously only been used as a bearing surface in THR. A newly developed direct cementless fixation all-ceramic (ZTA) resurfacing cup offers a new solution for resurfacing however ZTA has a Young's modulus approximately 1.6 times greater than CoCr - such may affect the acetabular bone remodelling. This modelling study investigates whether increased stress shielding may occur when compared to a CoCr resurfacing implant with successful known clinical survivorship.

METHODS

A finite element model of a hemipelvis constructed from CT scans was used and virtually reamed to a diameter of 58mm. Simulations were conducted and comparisons made of the ‘intact’ acetabulum and ‘as implanted’ with monobloc cups made from CoCr (Adept®, MatOrtho Ltd, UK) and ZTA (ReCerf ™, MatOrtho Ltd. UK) orientated at 35° inclination and 20° anteversion. The cups were loaded with 3.97kN representing a walking load of 280% for an upper bound height patient with a BMI of 35. The cup-bone interface was assigned a coulomb slip-stick function with a coefficient of friction of 0.5. The percentage change in strain energy density between the intact and implanted states was used to indicate hypertrophy (increase in density) or stress shielding (decrease in density).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 110 - 110
1 Apr 2019
Farrier A Manning W Moore L Avila C Collins S Holland J
Full Access

INTRODUCTION

Experience with Metal on Metal (MoM) hip resurfacing devices has shown adequate cementation of the femoral head is critical for implant survival. Bone necrosis can be caused by the temperature change in the peri-prosthetic bone whilst the cement cures during implantation. This can lead to implant loosening, head/neck fracture and implant failure. During the implantation it is known that implants change shape potentially altering joint clearance and causing loosening.

Given the history of Metal on Metal implant failure due adverse tissue reactions from Cobalt and Chromium particles we sought to test a novel Ceramic on Ceramic (CoC) bearing which may mitigate such problems.

AIM

We set out to compare the behaviour of a novel ceramic femoral head component to a standard metal component in a hip resurfacing system after cemented implantation in a physiological warmed cadaveric model.

Our first aim was to perform heat transfer analysis: To document time to, and extent of, maximum temperature change on the metal/ceramic surface and inside the resurfaced femoral head bone.

Our second aim was to perform a dimensional analysis: To document any resulting deformation in the metal/ceramic femoral head bearing diameter during cementation.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 112 - 112
1 Apr 2019
Farrier A Manning W Moore L Avila C Collins S Holland J
Full Access

INTRODUCTION

The cup component of modern resurfacing systems are often coated creating a cementless press-fit fixation in the acetabulum based on surgical under-reaming, also enabling osseoconduction/integration. Due to the higher density of cortical bone along the antero-superior and postero-inferior regions of the acetabulum, the greatest forces occur between the anterior and posterior columns of the pelvis. This produces pinching of the implant that can result in deformation of the cup. Metal shell/modularpress-fit acetabular cups are susceptible to substantial deformation immediately after implantation. This deformation may affect the lubrication, producing point loading and high friction torques between the head and the cup that increase wear and may lead to head clamping and subsequent cup loosening. We sought to test a novel ceramic on ceramic (CoC) hip resurfacing system that should allay any concerns with the Adverse Reaction to Metal Debris associated with metal on metal (MoM) resurfacing devices.

AIM

We sought to quantify the deformation of a novel CoC hip-resurfacing cup after implantation, using a standard surgical technique in a cadaveric model, and compare to the MoM standard. We also assessed if the design clearances proposed for this CoC hip resurfacing implant are compatible with the measured deformations, allowing for an adequate motion of the joint.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 93 - 93
1 Apr 2019
Avila C Taylor A Collins S
Full Access

INTRODUCTION

Unlike current acetabular cups, this novel ceramic cup has a Ti/HA coating which removes the requirement for assembly into a metal shell which avoiding potential chipping/misalignment and reducing wall thickness [Figure 1]. This study examines the resistance of novel thin-walled, direct to bone fixation ceramic cups to critical impact loads.

METHODS

Samples of the smallest (Ø46mm) and largest (Ø70mm) diameter ReCerfTM acetabular cups and corresponding femoral head implants were implanted into Sawbones foam blocks considered representative of pelvic cancellous bone. Two different positional configurations were tested and were considered worst case and the extremes of surgical compromise; P1 simulates the cup fully supported by the acetabulum with a high inclination angle (70°) and a vertical impaction axis (worst case loading near the cup rim) and. P2 simulates the cup implanted with a lower inclination (55°) but with the superior section unsupported by acetabulum bone [Figure 2]. For each size, three acetabular cups were tested in each position. The impact fixture was positioned within a drop weight rig above a bed of sand and ≈22mm of pork belly representative of soft tissues damping effect and the implant components aligned to achieve the defined impact point on the cup [Figure 2]. Lateral falls were tested on all available samples applying impact energy of 140J [1] and 3m/s impact velocity [2]. After the lateral fall test, each sample was tested under impact conditions equivalent to a frontal car crash considering a peak impact force of 5.7kN occurring 40ms from initial contact (able to produce acetabular fracture)[3].


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 48 - 48
1 Jul 2014
Lowry C Vincent G Traynor A Collins S
Full Access

Summary Statement

Corin has developed bone conserving prosthesis (MiniHip™) to better replicate the physiological load distribution in the femur. This study assessed whether the MiniHip™ prosthesis can better match the pre-osteoarthritic head centre for patient demographics when compared to contemporary long stem devices.

Introduction

Leg length and offset discrepancy resulting from Total Hip Replacement (THR) is a major cause of concern for the orthopaedic community. The inability to substitute the proximal portion of the native femur with a device that suitably mimics the pre-operative offset and head height can lead to loss of abductor power, instability, lower back pain and the need for orthodoses. Contemporary devices are manufactured based on predicate studies to cater for the variations within the patient demographic. Stem variants, modular necks and heads are often provided to meet this requirement. The number of components and instruments that manufacturers are prepared to supply however is limited by cost and an unwillingness to introduce unnecessary complexity. This can restrict the ability to achieve the pre-osteoarthritic head centre for all patient morphologies. Corin has developed MiniHip™ to better replicate the physiological load distribution in the femur. This study assessed whether the MiniHip™ prosthesis can better match the pre-osteoarthritic head centre for patient demographics when compared to contemporary long stem devices.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 81 - 81
1 Jul 2014
Kinbrum A Traynor A Collins S
Full Access

Summary

This work uses a mathematical method to correlate the forces calculated to push-on and pull off a femoral head from a stem and correlate the results of in vitro testing.

Introduction

This work aimed to mathematically model the force needed to disassemble the THR unit for a given assembly load. This work then compared these results with the results of an in vitro experiment. The research presented aimed to determine the assembly forces necessary to prevent movement of the head on the stem through friction. By assessing the forces necessary to push the head onto the stem securely enough to prevent any movement of the head through friction, it is likely that the fretting and corrosion of the head taper interface will be reduced.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 269 - 269
1 Dec 2013
Lowry C Vincent G Traynor A Simpson D Collins S
Full Access

Introduction:

Leg length and offset discrepancy resulting from Total Hip Replacement (THR) is a major cause of concern for the orthopaedic community. The inability to substitute the proximal portion of the native femur with a device that suitably mimics the pre-operative offset and head height can lead to loss of abductor power, instability, lower back pain and the need for orthodoses (1). Contemporary devices are manufactured based on predicate studies (2–4) to cater for the variations within the patient demographic. Stem variants, modular necks and heads are often provided to meet this requirement. The number of components and instruments that manufacturers are prepared to supply however is limited by cost and an unwillingness to introduce unnecessary complexity. This can restrict their ability to achieve the pre-osteoarthritic head centre for all patient morphologies. Corin has developed bone conserving prosthesis (MiniHip™) to better replicate the physiological load distribution in the femur. This study assesses whether the MiniHip™ prosthesis can better match the pre-osteoarthritic head centre for patient demographics when compared to contemporary long stem devices.

Method:

The Dorr classification is a well accepted clinical method for defining femoral endosteal morphology (5). This is often used by the surgeon to select the appropriate type and size of stem for the individual patient. It is accepted that a strong correlation exists between Flare Index (FI), characterising the thinning of cortical walls and development of ‘stove-pipe’ morphology, and age, in particular for females (Table 1) (3). A statistical model of the proximal femur was built from 30 full length femoral scans (Imorphics, UK). Minimum and maximum intramedullary measurements calculated from the statistical model were applied to relationships produced by combining Corins work with that of prior authors (Table 2) (2; 3; 6). This data was then used to generate 2D CAD models into which implants were inserted to compare the head centres achievable with a MiniHip™ device compared to those of a contemporary long stem.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 281 - 281
1 Dec 2013
De Villiers D Kinbrum A Traynor A Collins S Banfield S Housden J Shelton J
Full Access

Introduction

Vitamin-E has been introduced into highly-crosslinked polyethylene liners to reduce the oxidation potential of the material while maintaining low wear rates. However, little has been reported on adverse testing of the material with one test on diffused vitamin-E polyethylene [1] and no adverse tests of vitamin-E blended polyethylene reported. Adverse testing of crosslinked polyethylene has focused on the use of large diameters, the incorporation of third body particles, roughening of the counterface or severe activity [2–4]. This investigation considers the wear of vitamin-E blended highly-crosslinked polyethylene under standard and adverse conditions articulating against uncoated and chromium nitride (CrN) coated metal heads.

Methods

Seven metal heads were tested against prototype ϕ52 mm 0.1 wt% vitamin-E blended highly-crosslinked polyethylene liners (Corin, UK). Three heads remained as cast double heat treated metal (MoP) while four, of similar metallurgy, were coated with CrN via electron beam physical vapour deposition (CrNoP) (Tecvac, UK) and polished to a similar surface finish. Tests were conducted for 5 million cycles (mc) under conditions described in ISO 14242–3: 2009. Alumina particles (mean size 2.4 μm) at concentrations of 0.15 mg/mL were added to the lubricant for 1 mc to consider the effect of severe head damage. Testing continued for a further 1 mc without the presence of the particles and then 3 jogging intervals (14,400 cycles each) were conducted at slow, medium and fast speeds [3]. Wear volume was determined gravimetrically for the heads and liners and fluid collected throughout the testing was analysed for cobalt concentration using graphite furnace atomic absorption spectroscopy.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 34 - 34
1 Dec 2013
Lowry C Traynor A Collins S
Full Access

Introduction

Total Knee Replacement (TKR) is a highly effective treatment providing pain relief and improved function to patients experiencing advanced stage osteoarthritis. Tray fit or bone coverage is a critical design feature for both cemented and cementless designs affecting stability, load transfer and potential for infection. Many authors have attempted to characterise the relationship between the profile of the proximal tibia and gender and ethnicity1–3. As a consequence, a number of manufacturers have commercialised devices designed for specific gender and racial demographics. This study was initiated to compare the effect of the fixed minimum tibial resection depth prescribed by existing surgical instruments with that of a proportionate resection based on the size of the tibia.

Method

A dataset consisting of 30 donor scans from a US cadaver tissue bank (ScienceCare, Memphis, US) was used for this study. The dataset consisted of 12 male and 18 female specimens. Due to the limited view of the diaphysis for most scans, the natural slope of the lateral compartment was used as a guide for orienting the resection. All scans were resected with a 3° posterior slope. For the first part of this study, an equal mediolateral (ML) resection of 9.5 mm, reflecting the minimum resection for the Unity TKR tibia (Corin, UK), was performed on all specimens (Figure 1). Following this, two proportionate resection depths (13.5 mm and 6.7 mm) were calculated based on the ML relationship between the smallest and largest available Unity components (59.5 mm: 84.5 mm). Two further resection depths (11.3 mm and 8.0 mm) were calculated based on a mid size (71.0 mm). Three resection depths (8.0 mm, 9.5 mm & 11.3 mm) were applied to four medium sized specimens. In addition to this two larger specimens were resected at 9.5 mm and 13.5 mm and two smaller specimens at 6.7 mm and 9.5 mm.

A grid was applied to all cut surfaces and oriented using the posterior axis. The cut surface was divided based on lines drawn at 10%, 25%, 50%, 75% and 90% of the overall ML dimension and 10%, 25%, 50%, 75% and 90% of the overall anteroposterior (AP) dimension. Measurements were taken from the medial side and recorded from the points at which lines intersected the external profile of the cut tibia (Figure 2).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 270 - 270
1 Dec 2013
Traynor A Kinbrum A Housden J Collins S
Full Access

Introduction

Bearing surfaces of metal-on-metal (MoM) hip resurfacing devices and total hip replacements (THRs) are a known source of metallic debris. Further, large diameter heads and the high friction of a MoM joint are thought to lead to fretting and corrosion at the taper interface between modular components1. The metal debris generated can cause significant problems on the joint area2. This paper investigated fretting and corrosion of femoral head-neck junctions. Variables of the head-neck junction which may have an effect on fretting and corrosion were identified with the aim of determining the key drivers so that their risk on fretting and corrosion could be reduced through design. Additionally, a Chromium Nitride (CrN) coating was assessed to determine the effect on fretting and corrosion of coating the stem (male), head (female) or both trunnion interfaces. As there is currently no standard specification for a head-neck trunnion interface and trunnion designs vary significantly across the market, this work may lead to a positive change in the design and materials used in head-neck taper interfaces for all THR devices.

Methods

Suitable head and stem combinations were identified to enable individual variables such as; coating, medial-lateral (M-L) offset, head offset and taper angle to be isolated (Figure 1 and Figure 2). For the coated components a 3 μm CrN coating was applied to trunnion using electron beam physical vapour deposition (Tecvac, Cambridge, UK). Fretting and corrosion testing was carried out in accordance with ASTM F1875-98 (2009) method II procedure B3 following assembly of the components under a 2 kN load.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 219 - 219
1 Sep 2012
Wang QQ Wu JJ Unsworth A Simpson D Collins S Jarman-Smith M
Full Access

Introduction

Recent concerns over adverse effects of metal ion release, have led to the development of alternative hip joint replacements. This study reports the performance of new hemispherical MOTIS® (milled pitch-carbon fibre reinforced polyetheretherketone) acetabular cups articulating against Biolox Delta® femoral heads with the aim of producing lower wear and more biologically compatible bearings.

Materials and Methods

The wear performance of 40mm hemispherical MOTIS® cups articulating against Biolox Delta® heads has been investigated. The diametral clearance was 322±15.3nm (mean ± standard deviation). Wear tests were carried out on the Simplified Mark II Durham Hip Wear Simulator to 8 million cycles. New born bovine calf serum was used as the lubricant, diluted to give a protein content of 17g/l. Friction tests were carried out on the unworn joints and worn joints after 7.5 million cycles using lubricants containing protein (bovine serum based carboxymethyl cellulose (CMC) fluids) and without protein (water based CMC fluids). Temperature measured near every hip joint over a continuous wear testing period of 0.5 million cycles was recorded using PICO TC-08 data logger. One K-type thermocouple was placed carefully and consistently in each wear station and two were used to record the ambient room temperature. After stopping the wear test, the data logger continued recording the temperature for a further ten hours to indicate the cooling period. Additionally surface analyses were undertaken before and after wear testing using a non-contacting profilometer and atomic force (AFM) microscope.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 142 - 142
1 Sep 2012
Traynor A Simpson D Ellison P Collins S
Full Access

Introduction

Cobalt chrome on polyethylene remains a widely used bearing combination in total joint replacement. However wear induced osteolysis, bulk material property degradation of highly cross-linked polyethylene (HXLPE) [1], and oxidation after implantation (thought to be as a result of lipid absorption or cyclic loading [2]) remains a concern. ECIMA is a cold-irradiated, mechanically annealed, vitamin E blended next generation HXLPE developed to maintain mechanical properties, minimise wear and to improve the oxidation resistance in the long-term. The aim of this study was to compare the in-vitro wear rate and mechanical properties of three different acetabular liners; conventional UHMWPE, HXLPE and ECIMA.

Methods

Twelve liners (Corin, UK) underwent a 3 million cycle (mc) hip simulation. Three conventional UHMWPE liners (GUR1050, Ø32 mm, 30 kGy sterilised in Nitrogen), three HXLPE liners (GUR1020, Ø40 mm, 75 kGy cross-linking and EtO sterilised) and six ECIMA liners (0.1 wt% vitamin E GUR1020, Ø40 mm, 120 kGy cross-linking, mechanically deformed and annealed, and EtO sterilised) articulated against CoCrMo alloy femoral heads to ASTM F75 (Corin, UK). Wear testing was performed in accordance with ISO 14242 parts 1 and 2, with a maximum force of 3.0 kN and at a frequency of 1 Hz. The test lubricant used was calf serum with a protein content of 30 g/l and 1% (v/v) patricin added as an antibacterial agent. Volumetric wear rate was determined gravimetrically after the first 0.5 mc and every 1 mc thereafter.

ASTM D638 type V specimens (3.2 mm thick) were machined from ECIMA material for uniaxial tension testing to ASTM D638. Ultimate tensile strength (UTS), yield strength and elongation values were measured. These values were compared to mechanical data available for the other material types.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 177 - 177
1 Sep 2012
Yeoman M Lowry C Cizinauskas A Vincent G Simpson D Collins S
Full Access

INTRODUCTION

Bone resorption around hip stems, in particular periprosthetic bone loss, is a common observation post-operatively. A number of factors influence the amount of bone loss over time and the mechanical environment following total hip replacement (THR) is important; conventional long stem prostheses have been shown to transfer loads distally, resulting in bone loss of the proximal femur. More conservative, short stems have been recently introduced to attempt to better replicate the physiological load distribution in the femur. The aim of this study was to evaluate the bone mineral density (BMD) change over time, in a femur implanted with either a short or a long stem.

METHODS

Finite element models of two implants, a short (Minihip, Corin, UK) and long (Metafix, Corin, UK) hip stem were used to simulate bone remodeling under a physiological load condition (stair climbing). The magnitudes and directions of the muscle forces and joint reaction force were obtained from Heller et al (2001, 2005). An unimplanted femur was also simulated.

A strain-adaptive remodelling theory (Scannel & Prendergast 2009) was utilised to simulate remodelling in the bone after virtual implantation. COMSOL Multiphysics software was used for the analysis. The strain component of the remodelling stimulus was strain energy density per unit mass. This was calculated in the continuum model from the strain energy density, and apparent density.

Bone mass was adapted using a site-specific approach in an attempt to return the local remodelling stimulus to the equilibrium stimulus level (calculated from the unimplanted femur). The minimal inhibitory signal proposed by Frost (1964), was included in the model and described by a ‘lazy zone’, where no bone remodelling occurred.

The three dimensional geometry of the femur was constructed from computed tomography data of the donor (female, 44 years old, right side). Elemental bone properties were assigned from the Hounsfield Unit values of the CT scans. The elastic modulus of the bone was assumed to be isotropic and was determined using a relationship to the apparent bone density (Frost 1964, Rho 1995). The Poisson's ratio for the bone regions varied between 0.2 and 0.32 depending on the apparent density of the bone (Stulpner 1997).

The period of implantation analysed was 2 years. The muscle forces and joint contact loads applied were ramped linearly from zero to full load over a period of two weeks, representing the estimated post operative rest period of a patient.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 176 - 176
1 Sep 2012
Simpson D Traynor A Collins S
Full Access

INTRODUCTION

Wear induced osteolysis, material property degradation and oxidation remain a concern in cobalt chrome on polyethylene THR. ECIMA is a cold-irradiated, mechanically annealed, vitamin E blended HXLPE developed to maintain mechanical properties, minimise wear and improve long-term oxidation resistance. This study aimed to compare the in-vitro wear rate and mechanical properties of three different acetabular liners; UHMWPE, HXLPE and ECIMA.

METHODS

Twelve liners (Corin, UK) underwent a 3 million cycle (mc) hip simulation. Three UHMWPE (GUR1050, Ø32 mm, γ sterilised), three HXLPE (GUR1020, Ø40 mm, 75 kGy γ, EtO sterilised) and six ECIMA (0.1 wt% vitamin E GUR1020, Ø40 mm, 120 kGy γ, mechanically annealed, EtO sterilised) liners articulated against CoCrMo femoral heads (Corin, UK). Wear testing was performed in accordance with ISO 14242 parts 1 and 2, in calf serum, with a maximum force of 3.0 kN and at a frequency of 1 Hz. Volumetric wear rate was determined gravimetrically.

ASTM D638 type V specimens were machined from ECIMA material for uniaxial tension testing. Ultimate tensile strength (UTS), yield strength and elongation values were measured. These values were compared to mechanical data available for the other material types.

Following completion of the ECIMA wear testing, three of the tested liners were cut in half. One half of each was subject to accelerated ageing in accordance with ASTM F2003-02, while the other half was tested as received. Each liner half was cross-sectioned and a microtome was used to section 200μm thick slices from each cross-section. Oxidation analysis was performed using a Fourier Transform Infra-red technique in accordance with ASTM F2102-01 throughout the thickness of each liner half. Average oxidation indices for each sample were determined.