header advert
Results 1 - 2 of 2
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 446 - 450
1 Apr 2008
Bailie AG Lewis PL Brumby SA Roy S Paterson RS Campbell DG

The Unispacer knee system is a cobalt-chrome self-centring tibial hemiarthroplasty device for use in the treatment of isolated medial compartment osteoarthritis of the knee. The indications for use are similar to those for high tibial osteotomy, but insertion does not require bone cuts or component fixation, and does not compromise future knee replacement surgery. A prospective study of a consecutive series of 18 patients treated with the Unispacer between June 2003 and August 2004 was carried out to determine the early clinical results of this device. The mean age of the patients was 49 years (40 to 57). A total of eight patients (44%) required revision within two years. In two patients revision to a larger spacer was required, and in six conversion to either a unicompartmental or total knee replacement was needed. At the most recent review 12 patients (66.7%) had a Unispacer remaining in situ. The mean modified visual analogue score for these patients at a mean follow-up of 19 months (12 to 26) was 3.0 (0 to 11.5). The mean pain level was 30% that of the mean pre-operative level of 10. The early clinical results using this device have been disappointing.

This study demonstrates that use of the Unispacer in isolated medial compartment osteoarthritis is associated with a high rate of revision surgery and provides unpredictable relief of pain.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 734 - 739
1 Jun 2006
Campbell DG Duncan WW Ashworth M Mintz A Stirling J Wakefield L Stevenson TM

A series of 100 consecutive osteoarthritic patients was randomised to undergo total knee replacement using a Miller-Galante II prosthesis, with or without a cemented polyethylene patellar component. Knee function was evaluated using the American Knee Society score, Western Ontario and McMaster University Osteoarthritis index, specific patellofemoral-related questions and radiographic evaluation until the fourth post-operative year, then via questionnaire until ten years post-operatively. A ten-point difference in the American Knee Society score between the two groups was considered a significant change in knee performance, with α and β levels of 0.05.

The mean age of the patients in the resurfaced group was 71 years (53 to 88) and in the non-resurfaced group was 73 years (54 to 86).

After ten years 22 patients had died, seven were suffering from dementia, three declined further participation and ten were lost to follow-up. Two patients in the non-resurfaced group subsequently had their patellae resurfaced. In the resurfaced group one patient had an arthroscopic lateral release. There was no significant difference between the two treatment groups: both had a similar deterioration of scores with time, and no further patellofemoral complications were observed in either group.

We are unable to recommend routine patellar resurfacing in osteoarthritic patients undergoing total knee replacement on the basis of our findings.