header advert
Results 1 - 3 of 3
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 6 - 6
1 Jun 2023
Watts D Bye D Nelson D Chase H Nunney I Marshall T Sanghrajka A
Full Access

Introduction

Derotation osteotomies are commonly performed in paediatric orthopaedic and limb reconstruction practice. The purpose of this study was to determine whether the use of a digital inclinometer significantly improves the accuracy in attaining the desired correction.

Materials & Methods

We designed an electronic survey regarding derotation femoral osteotomy (DFO) including methods of intra-operative angular correction assessment and acceptable margins of error for correction. This was distributed to 28 paediatric orthopaedic surgeons in our region. A DFO model was created, using an anatomic sawbone with foam covering. 8 orthopaedic surgeons each performed two 30-degree DFOs, one using K-wires and visual estimation (VE), and the other using a Digital Inclinometer (DI). Two radiologists reported pre and post procedure rotational profile CT scans to assess the achieved rotational correction.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 5 - 5
1 Aug 2021
Thompson J Corbett J Bye D Jones A Tissingh E Nolan J
Full Access

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

A systematic literature search was performed adhering to PRISMA criteria. In parallel we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003– June 2020.

There are 25 reported cases of such prosthetic fractures confined to small case series and case reports within the literature. We report an additional 19 cases to the literature (mean age 66.3 ± 11.7 years; 12 female [63%]; body mass index 32.9 ± 5.9 kg/m2). The mean time from index procedure to fracture was 7.8 years (2.5–16.3, ±3.6). Exeter V40 stem fracture incidence was 0.27%. Incidence was significantly higher in 125 mm length stems compared to ≥150 mm length stems (1.26% vs 0.13%, respectively, p <0.001) and revision arthroplasty (1.209% vs 0.149%, p <0.001). When comparing different stem length cohorts, 125-mm short-stem were associated with stem body fractures (92% vs 29%, p = 0.0095), earlier time to fracture (6.2 vs 11.0 years, p = 0.0018), younger patient age at time of fracture (62.7 vs 72.6 years old, p = 0.037) and female sex (75% vs 43%, p = 0.33).

This case series in conjunction with the systematic review provides evidence stem morphology plays a role in femoral implant fracture. This complication remains rare, although we report a significantly higher incidence at up to 17 years follow-up than in the literature. As femoral geometries remain the same, increasing BMIs in THR patients should raise concern. Short 125 mm length Exeter V40 stems undoubtedly have a role in restoring anatomy and biomechanics in smaller femoral geometries, although the surgeon has to appreciate the higher risk of stem fracture and the associated predisposing factors which may necessitate meticulous surgical technique and planning.


Bone & Joint Open
Vol. 2, Issue 6 | Pages 443 - 456
28 Jun 2021
Thompson JW Corbett J Bye D Jones A Tissingh EK Nolan J

Aims

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

Methods

A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.