Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 59 - 59
1 May 2016
Buente D Huber G Morlock M
Full Access

Introduction

Failure of the neck-stem taper in one particular bi-modular primary hip stem due to corrosion and wear of the neck piece has been reported frequently1, and stems were recalled. A specific pattern of material loss on the CoCr neck-piece taper in the areas of highest stresses on the proximal medial male taper was observed in a retrieval study of 27 revised Rejuvenate implants revised after 3 to 38 month time in situ (Stryker, Kalamazoo, MI, USA) (Figure 1). One neck piece exhibited additionally wear marks at the distal end of the flat male neck taper indicating contact with the female taper of the stem. The purpose of this study was to understand the observed failure scenario of bottoming-out by investigating the stem taper morphologies.

Materials and Methods

The geometry of taper contact surfaces was determined using a Coordinate Measurement Machine (BHN 805, Mitutoyo, Japan). An algorithm based on the individual unworn areas of the respective taper surfaces was applied to all retrievals. One retrieval is additionally investigated by infinite focus microscopy (G4, Alicona, Austria) in the main wear areas on the neck piece taper, and the bottom, facing each other inside the junction (surfaces of the distal end of the male and the bottom of the female taper).


The early failure and revision of bimodular primary total hip arthroplasty prostheses requires the identification of the risk factors for material loss and wear at the taper junctions through taper wear analysis. Deviations in taper geometries between revised and pristine modular neck tapers were determined using high resolution tactile measurements. A new algorithm was developed and validated to allow the quantitative analysis of material loss, complementing the standard visual inspection currently used.

The algorithm was applied to a sample of 27 retrievals (in situ from 2.9 to 38.1 months) of the withdrawn Rejuvenate modular prosthesis. The mean wear volumes on the flat distal neck piece taper was 3.35 mm3 (0.55 to 7.57), mainly occurring in a characteristic pattern in areas with high mechanical loading. Wear volume tended to increase with time to revision (r² = 0.423, p = 0.001). Implant and patient specific data (offset, stem size, patient’s mass, age and body mass index) did not correlate with the amount of material loss observed (p >  0.078). Bilaterally revised implants showed higher amounts of combined total material loss and similar wear patterns on both sides. The consistent wear pattern found in this study has not been reported previously, suggesting that the device design and materials are associated with the failure of this prosthesis.

Cite this article: Bone Joint J 2015;97-B:1350–7.