header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 78 - 78
1 Dec 2013
Hothi H Cro S Bills P Blunt L Racasan R Blunn GW Skinner JA Hart A
Full Access

Introduction

Metal-on-metal (MOM) total hip arthroplasty using large diameter femoral heads offer clinical advantages however the failure rates of these hips is unacceptably high. Retrieved hips have a wide range of wear rates of their bearing and taper surfaces and there is no agreement regarding the cause of failure.

Detailed visual inspection is the first step in the forensic examination of failed hip components and may help explain the mechanisms of failure. The aim of this study was to determine if there was a correlation between the results of detailed inspections and the volumetric wear of the bearing and taper surfaces of retrieved hips.

Method

Detailed, non-destructive macroscopic and stereomicroscopic examinations of 89 retrieved MOM hip components were performed by a single experienced examiner using quantitative assessment to document the severity of 10 established damage features:

Light scratches, Moderate scratches, Heavy scratches, Embedded particles, Discolouration, Haziness, Pitting, Visible wear zone, Corrosion, Fretting

Each surface was considered in terms of zones comprising of quadrants (cup, head, and taper) and subquadrants (cup and head), Figure 1. Each zone was scored on a scale of 0 to 3 by determining the percentage of the surface area of the zone that exhibited the feature in question: a score of 0=0%, 1<25%, 25%<2<75%, 3>75%. The sum of the scores of each zone was used for the assessment of each damage feature.

The volume of wear at the surfaces of each hip was measured with a Zeiss Prismo coordinate measuring machine (cup and head) and a Talyrond 365 roundness measurement instrument (taper), using previously reported methods1, 2.

Simple linear regression models were used to asses the univariable associations between the inspection scores and wear volumes. Multiple linear regression models were subsequently used to asses the simultaneous contribution of the inspection scores, found significant in univariable analyses, on the wear outcome variables.

All statistical analysis was performed using Stata/IC version 12.1 (StataCorp, USA) and throughout a p value < 0.05 was considered statistically significant.