Sewell MD Spiegelberg BGI Hanna SA Aston WJS Meswania JM Blunn GW Henry C Cannon SR Briggs TWR

We describe the application of a non-invasive extendible endoprosthetic replacement in skeletally-mature patients undergoing revision for failed joint replacement with resultant limb-length inequality after malignant or non-malignant disease. This prosthesis was developed for tumour surgery in skeletally-immature patients but has now been adapted for use in revision procedures to reconstruct the joint or facilitate an arthrodesis, replace bony defects and allow limb length to be restored gradually in the post-operative period.

We record the short-term results in nine patients who have had this procedure after multiple previous reconstructive operations. In six, the initial reconstruction had been performed with either allograft or endoprosthetic replacement for neoplastic disease and in three for non-neoplastic disease. The essential components of the prosthesis are a magnetic disc, a gearbox and a drive screw which allows painless lengthening of the prosthesis using the principle of electromagnetic induction. The mean age of the patients was 37 years (18 to 68) with a mean follow-up of 34 months (12 to 62). They had previously undergone a mean of six (2 to 14) open procedures on the affected limb before revision with the non-invasive extendible endoprosthesis.

The mean length gained was 56 mm (19 to 107) requiring a mean of nine (3 to 20) lengthening episodes performed in the outpatient department. There was one case of recurrent infection after revision of a previously infected implant and one fracture of the prosthesis after a fall. No amputations were performed. Planned exchange of the prosthesis was required in three patients after attainment of the maximum lengthening capacity of the implant. There was no failure of the lengthening mechanism. The Mean Musculoskeletal Tumour Society rating score was 22 of 30 available points (18 to 28).

The use of a non-invasive extendible endoprosthesis in this manner provided patients with good functional results and restoration of leg-length equality, without the need for multiple open lengthening procedures.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1373 - 1377
1 Oct 2009
Spiegelberg BGI Sewell MD Aston WJS Blunn GW Pollock R Cannon SR Briggs TWR

This paper describes the preliminary results of a proximal tibial endoprosthesis which spares the knee joint and enables retention of the natural articulation by replacing part of the tibial metaphysis and diaphysis. In eight patients who had a primary malignant bone tumour of the proximal tibia, the distal stem, which had a hydroxyapatite-coated collar to improve fixation, was cemented into the medullary canal. The proximal end had hydroxyapatite-coated extracortical plates which were secured to the remaining proximal tibial metaphysis using cortical screws. The mean age of the patients at operation was 28.9 years (8 to 43) and the mean follow-up was for 35 months (4 to 48). The mean Musculoskeletal Tumour Society score was 79% (57% to 90%), the mean Oxford Knee score was 40 points of 48 (36 to 46) and the mean knee flexion was 112° (100° to 120°). In one patient, revision to a below-knee amputation through the prosthesis was required because of recurrence of the tumour. Another patient sustained a periprosthetic fracture which healed with a painful malunion. This was revised to a further endoprosthesis which replaced the knee.

In the remaining six patients the prosthesis allowed preservation of the knee joint with good function and no early evidence of loosening. Further follow-up is required to assess the longevity of these prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1257 - 1262
1 Sep 2009
Sundar S Pendegrass CJ Oddy MJ Blunn GW

We used demineralised bone matrix (DBM) to augment re-attachment of tendon to a metal prosthesis in an in vivo ovine model of reconstruction of the extensor mechanism at the knee. We hypothesised that augmentation of the tendon-implant interface with DBM would enhance the functional and histological outcomes as compared with previously reported control reconstructions without DBM. Function was assessed at six and 12 weeks postoperatively, and histological examination was undertaken at 12 weeks.

A significant increase of 23.5% was observed in functional weight-bearing at six weeks in the DBM-augmented group compared with non-augmented controls (p = 0.004). By 12 weeks augmentation with DBM resulted in regeneration of a more direct-type enthesis, with regions of fibrocartilage, mineralised fibrocartilage and bone. In the controls the interface was predominantly indirect, with the tendon attached to the bone graft-hydroxyapatite base plate by perforating collagen fibres.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 815 - 819
1 Jun 2009
Spiegelberg BGI Sewell MD Coltman T Blunn GW Flanagan AM Cannon SR Briggs TWR

We report a case which highlights the progression of osteofibrous dysplasia to adamantinoma and questions whether intralesional curettage is the appropriate treatment. The role of a joint-sparing massive endoprosthesis using cortical fixation is demonstrated and we describe a unique biomedical design which resulted in the manufacture of an end cap to allow amputation through a custom-made proximal tibial replacement, rather than an above-knee amputation following recurrence.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_II | Pages 262 - 262
1 May 2009
Sundar S Pendegrass CJ Blunn GW
Full Access

Research in to tendon-bone healing techniques focus on increasing bone growth at the interface such as cell or growth factor (e.g. BMP-2) augmentation. Demineralised bone matrix (DBM) is osseoinductive and is in use clinically.

Hypothesis: DBM augmentation of a healing tendon-bone interface will result in improved function at 3, 6, 9 and 12 weeks, and a morphology that more closely resembles that of a normal enthesis at 12 weeks.

Materials and methods: An ovine patellar tendon model was used. 19 skeletally mature ewes were allocated to the control group or DBM group. In both groups the patellar tendon was detached, and following tibial tubercle osteotomy, was re-attached using 3 suture anchors. In the DBM group a piece of DBM was placed between the tendon and bone. 2 animals were sacrificed at 6 weeks and 6 animals at 12 weeks. Animals underwent force plate analysis at 3, 6, 9 and 12 weeks. The tendon-bone interface length which was fibrous or fibrocartilaginous, and the area of fibrocartilage, mineralised fibrocartilage and new bone was quantified.

Results: 3 control group animals (33%) failed within 6 weeks. None failed in the DBM group. The DBM group was significantly better than the control at all time points (p< 0.05). DBM produced a significantly more fibrocartilaginous enthesis than the control group (p< 0.05). Controls were significantly more fibrous than the DBM group (p< 0.05). DBM produced significantly more fibrocartilage (p< 0.05), and mineralised fibrocartilage (p< 0.05).

Discussion: 33% of the control group failed within 6 weeks, whilst no failures were observed in the DBM group. DBM animals mobilised earlier and had significantly better function at all time points. Histologically, the DBM group showed a more mature direct type enthesis at earlier time points.

Conclusion: DBM augmentation of a healing tendon-bone interface enhances functional and morphological recovery at earlier time points.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 535 - 541
1 Apr 2008
Pendegrass CJ Sundar S Oddy MJ Cannon SR Briggs T Blunn GW

We used an in vivo model to assess the use of an autogenous cancellous bone block and marrow graft for augmenting tendon reattachment to metallic implants. We hypothesised that augmentation of the tendon-implant interface with a bone block would enable retention of the graft on the implant surface, enhance biological integration, and result in more consistent functional outcomes compared with previously reported morcellised graft augmentation techniques.

A significant improvement in functional weight-bearing was observed between six and 12 weeks. The significant increase in ground reaction force through the operated limb between six and 12 weeks was greater than that reported previously with morcellised graft augmented reconstructions. Histological appearance and collagen fibre orientation with bone block augmentation more closely resembled that of an intact enthesis compared with the morcellised grafting technique. Bone block augmentation of tendon-implant interfaces results in more reliable functional and histological outcomes, with a return to pre-operative levels of weight-bearing by 24 weeks.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 114 - 121
1 Jan 2008
Pendegrass CJ Gordon D Middleton CA Sun SNM Blunn GW

Conventional amputation prostheses rely on the attachment of the socket to the stump, which may lead to soft-tissue complications. Intraosseous transcutaneous amputation prostheses (ITAPs) allow direct loading of the skeleton, but their success is limited by infection resulting from breaching of the skin at the interface with the implant. Keratinocytes provide the skin’s primary barrier function, while hemidesmosomes mediate their attachment to natural ITAP analogues. Keratinocytes must attach directly to the surface of the implant. We have assessed the proliferation, morphology and attachment of keratinocytes to four titaniumalloy surfaces in order to determine the optimal topography in vitro. We used immunolocalisation of adhesion complex components, scanning electron microscopy and transmission electron microscopy to assess cell parameters.

We have shown that the proliferation, morphology and attachment of keratinocytes are affected by the surface topography of the biomaterials used to support their growth. Smoother surfaces improved adhesion. We postulate that a smooth topography at the point of epithelium-ITAP contact could increase attachment in vivo, producing an effective barrier of infection.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1498 - 1503
1 Nov 2007
Hanna SA David LA Aston WJS Gikas PD Blunn GW Cannon SR Briggs TWR

Between 1988 and 2006, 18 patients had a custom-made endoprosthetic replacement of the distal humerus for bone tumours at our institution. There were 11 primary malignant neoplasms, six secondary deposits, and one benign aggressive tumour. The mean follow-up was for 4.4 years (1 to 18.2). Complications occurred in nine patients and included aseptic loosening in three (16.6%), local recurrence in two (11%), infection in two (11%), neuropraxia of the radial nerve in one (5.5%) and a peri-prosthetic fracture in one (5.5%). Excision was inadequate in four patients (22%), all of which developed local recurrence and/or metastases. There were seven deaths from the primary disease after a mean of 2.3 years (1 to 5), one of whom had an above-elbow amputation for local recurrence seven months before death. The remaining six had satisfactory elbow function at their last follow-up. The 11 living patients were evaluated using the Musculoskeletal Tumour Society and Toronto Extremity Salvage scoring systems. The mean scores achieved were 76% (67% to 87%) and 73% (59% to 79%), respectively. Overall, 17 of 18 patients had significant improvement in the degree of their pain following operation.

Custom-made endoprosthetic reconstruction of the elbow for bone tumours is a viable treatment in carefully selected patients. It maintains satisfactory function and provides good pain relief.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_III | Pages 416 - 417
1 Oct 2006
Coathup MJ Smith N Kingsley C Collins L Dattani R Blunn GW
Full Access

Introduction Bone graft supply for impaction grafting can be problematic due to the supply of graft, sterilisation, which alters the biological properties of the graft, and the immunogencity of the graft which may lead to graft rejection. Reducing the amount of graft can be accomplished by using increased amounts of synthetic materials such as hydroxyapatite (HA). This study evaluated the effect of using mixtures of porous HA (Apapore™) with allograft for cemented impaction allografting of the femoral stem in an ovine model. The aim was to test the hypothesis that increased quantities of Apapore™ will be stable and induce similar bone remodelling to that where a 50:50 mixture with allograft was used.

Method Twelve hemi-arthroplasty femoral components were inserted into the right hip of skeletally mature female commercially cross-bred sheep weighing between 65 and 80kg. Femoral components were manufactured from Cobalt Chromium alloy and cemented in place following impaction of the femoral canal. Animals were randomly placed into one of two groups according to the allograft-apapore mixture used. Group 1: Apapore:allograft mixed 50:50. Group 2: Apapore: allograft mixed 90:10. Six animals were investigated in each group. Implants remained in vivo for 6 months. In order to quantify bone formation rates, oxytetracycline injections were given 2 months post-surgery and 3 weeks later, followed by a third administration in the fifth month post-surgery and 3 weeks later. Animals were walked over a force plate pre-operatively and at 8, 16 and 24 weeks post-operatively. Twelve readings of maximum force (Fmax, N/m2) were taken and average values of right over left were calculated as a percentage (%AR/AL) and represented how well the animal used its operated leg where 100% represents full weight-bearing. Thin sections (~70μm thick) were prepared through four regions of the femur. The proximal, mid and tip of the femoral component region and one distal to the implant tip were analysed where bone area, Apapore™ area, Apapore™-bone contact and cement mantle thickness were quantified and compared using image analysis techniques.

Results In both groups, the use of graft resulted in the formation of a cancellous network of bone on the endosteal surface which incorporated the Apapore™ granules. When all regions were compared, femoral bone turnover results demonstrated significantly increased rates in group 1 (0.0021mm day-1) when compared with group 2 (0.0015mm day-1) (p< 0.05). No significant differences were identified when the proximal, mid and tip regions in the two groups were compared however, significantly increased turnover was identified in the distal region in group 1 (0.0027 mm day-1) when compared with group 2 (0.0013mm day-1) (p< 0.05). In both groups increased turnover was observed in the proximal, tip and distal regions with least in the mid region of the stem. Ground Reaction Force (GRF) results demonstrated no significant differences between the two experimental groups at 8, 16 and 24 weeks postoperatively. In both groups, a significant decline in function was demonstrated 8 weeks post-op when compared with pre-operative values and in both groups function gradually increased over time. Results for new bone area demonstrated significantly increased new bone in the proximal and distal regions in both groups (proximal =7.94mm2 and 7.13mm2; distal =7.03mm2 and 8.17mm2, group 1 and 2 respectively) with least new bone in the mid region of the stem (4.53mm2 and 4.79mm2). No significant differences in any of the regions were demonstrated when group 1 and 2 were compared. In both groups, significantly increased amounts of Apapore™ was observed in the proximal and distal regions of the femoral stem with least in the mid and tip region. No significant difference in cement mantle thickness was identified between the two groups.

Discussion Results demonstrated that hips maintained functional stability when a higher amount of Apapore™ mixture was used. Results for bone turnover rates and the amount of new bone formation in the 90:10 mixture demonstrated Apapore™ to be a comparable and suitable alternative to replace allograft in impaction grafting of a femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 9 | Pages 1245 - 1251
1 Sep 2006
Pendegrass CJ Oddy MJ Sundar S Cannon SR Goodship AE Blunn GW

We examined the mechanical properties of Vicryl (polyglactin 910) mesh in vitro and assessed its use in vivo as a novel biomaterial to attach tendon to a hydroxyapatite-coated metal implant, the interface of which was augmented with autogenous bone and marrow graft. This was compared with tendon re-attachment using a compressive clamp device in an identical animal model. Two- and four-ply sleeves of Vicryl mesh tested to failure under tension reached 5.13% and 28.35% of the normal ovine patellar tendon, respectively. Four-ply sleeves supported gait in an ovine model with 67.05% weight-bearing through the operated limb at 12 weeks, without evidence of mechanical failure.

Mesh fibres were visible at six weeks but had been completely resorbed by 12 weeks, with no evidence of chronic inflammation. The tendon-implant neoenthesis was predominantly an indirect type, with tendon attached to the bone-hydroxyapatite surface by perforating collagen fibres.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_II | Pages 257 - 257
1 May 2006
Wardle NS Lee JKL Haddad FS Blunn GW
Full Access

Background And Aims: Total knee replacements provide a cost effective treatment for painful joint conditions such as osteoarthritis. Their long term performance is governed by ultra-high molecular weight polyethylene (UHMWPE) wear which produces wear debris and leads to osteolysis and aseptic loosening of the implant. Using a new material which is more scratch resistant than cobalt chrome it is hoped to reduce wear of UHMWPE and its subsequent complications.

Methods: Two total knee replacements made from cobalt chrome and two of oxidised zirconium were wear tested in a knee simulator. Surface roughness data from the femoral components was collected. Gravimetric and volumetric wear of the polyethylene inserts was measured. SEM analysis of the surfaces was also performed.

Results: Oxidised zirconium is significantly more scratch resistant than CoCr; Ra (mean average roughness) of 0.7μm compared to 0.43μm (p< < 0.01) at end of test, and confirmed by SEM analysis. This was accompanied by a 4 fold reduction in wear of UHMWPE 49.60mg to 12.48mg (p=0.02). Barium sulphate particles were found to be embedded in the surface of the CoCr implants.

Summary And Conclusions: Oxidised zirconium leads to a reduction in wear of UHMWPE due to its increased resistance to third body wear. It has the potential to increase the longevity of total knee replacements by reducing UHMWPE related complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 873 - 878
1 Jun 2005
Oddy MJ Pendegrass CJ Goodship AE Cannon SR Briggs TWR Blunn GW

We developed an in vivo model of the attachment of a patellar tendon to a metal implant to simulate the reconstruction of an extensor mechanism after replacement of the proximal tibia. In 24 ewes, the patellar tendon was attached to a hydroxyapatite (HA)-coated titanium prosthesis. In 12, the interface was augmented with autograft containing cancellous bone and marrow. In the remaining ewes, the interface was not grafted.

Kinematic gait analysis showed nearly normal function of the joint by 12 weeks. Force-plate assessment showed a significant increase in functional weight-bearing in the grafted animals (p = 0.043). The tendon-implant interface showed that without graft, encapsulation of fibrous tissue occurred. With autograft, a developing tendon-bone-HA-implant interface was observed at six weeks and by 12 weeks a layered tendon-fibrocartilage-bone interface was seen which was similar to a direct-type enthesis.

With stable mechanical fixation, an appropriate bioactive surface and biological augmentation the development of a functional tendon-implant interface can be achieved.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 1 | Pages 118 - 123
1 Jan 2001
Coathup MJ Blunn GW Flynn N Williams C Thomas NP

We investigated the implant-bone interface around one design of femoral stem, proximally coated with either a plasma-sprayed porous coating (plain porous) or a hydroxyapatite porous coating (porous HA), or which had been grit-blasted (Interlok). Of 165 patients implanted with a Bimetric hip hemiarthroplasty (Biomet, Bridgend, UK) specimens were retrieved from 58 at post-mortem.

We estimated ingrowth and attachment of bone to the surface of the implant in 21 of these, eight plain porous, seven porous HA and six Interlok, using image analysis and light morphometric techniques. The amount of HA coating was also quantified.

There was significantly more ingrowth (p = 0.012) and attachment of bone (p > 0.05) to the porous HA surface (mean bone ingrowth 29.093 ± 2.019%; mean bone attachment 37.287 ± 2.489%) than to the plain porous surface (mean bone ingrowth 21.762 ± 2.068%; mean bone attachment 18.9411 ± 1.971%). There was no significant difference in attachment between the plain porous and Interlok surfaces. Bone grew more evenly over the surface of the HA coating whereas on the porous surface, bone ingrowth and attachment occurred more on the distal and medial parts of the coated surface. No significant differences in the volume of HA were found with the passage of time.

This study shows that HA coating increases the amount of ingrowth and attachment of bone and leads to a more even distribution of bone over the surface of the implant. This may have implications in reducing stress shielding and limiting osteolysis induced by wear particles.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 927 - 937
1 Nov 1997
Schindler OS Cannon SR Briggs TWR Blunn GW

The use of extendible distal femoral replacements is a relatively new treatment alternative for malignant bone tumours in growing individuals. Although their appearance was widely appreciated, questions about functional practicality and longevity remain unclear. With longer follow-up, advantages of immediate functional restoration and beneficial psychological aspects seem to be overshadowed by an increase in complications such as aseptic loosening, infection or prosthetic failure.

We have reviewed 18 children with such tumours who were treated between 1983 and 1990 by custom-made Stanmore extendible distal femoral replacements. Four died from metastatic disease within 2.5 years of operation and two required amputation for local recurrence or chronic infection.

The remaining 12 patients were followed for a mean of 8.7 years (6 to 13.2). A mean total lengthening of 5.2 cm was achieved, requiring, on average, 4.3 operations. Using the Musculoskeletal Tumor Society rating score the functional result at review was, on average, 77% of the expected normal function, with seven patients achieving ≥ 80%. Revision of the prosthesis was required in ten patients, in six for aseptic loosening, at a mean of 6.2 years after the initial procedure.