Zampelis V Ornstein E Franzén H Atroshi I

Previously, radiostereometric analysis following hip revision performed using impacted morsellised allograft bone and a cemented Exeter stem has shown continuous subsidence of the stem for up to five years. It is not known whether the subsidence continues thereafter. In our study, 17 of 25 consecutive osteo-arthritic patients with aseptically loose stems who underwent first-time revision using impacted morsellised allograft bone and a cemented Exeter stem were followed by yearly radiostereometric examinations for nine years. The mean subsidence at six weeks was 1.1 mm (0.1 to 2.3), from six weeks to one year 1.3 mm (0 to 2.6), from one to five years 0.7 mm (0 to 2.0), and from five to nine years 0.7 mm (0.1 to 3.1). That from six weeks to nine years was 2.7 mm (0 to 6.4) (95% confidence interval 2.0 to 3.5). The Charnley pain score significantly improved after revision, and was maintained at nine years, but walking ability deteriorated slightly as follow-up extended. Of the eight patients who were not followed for nine years, two had early subsidence exceeding 11 mm.

Our findings show that in osteo-arthritic patients who undergo revision for aseptic loosening of the stem using impacted morsellised allograft bone and a cemented Exeter stem, migration of the stem continues over nine years at a slower rate after the first year, but without clinical deterioration or radiological loosening.