Weale AE Murray DW Baines J Newman JH

Failure of a unicompartmental knee replacement (UKR) may be caused by progressive osteoarthritis of the knee and/or failure of the prosthesis. Limb alignment can influence both of these factors. We have examined the fate of the other compartments and measured changes in leg alignment after UKR.

A total of 50 UKRs was carried out on 45 carefully selected patients between 1989 and 1992. At operation, deliberate attempts were made to avoid overcorrection of the deformity. Four patients died, one patient was lost to follow-up and two knees were revised before review which was at a minimum of five years. Standard long-leg weight-bearing anteroposterior views of the knee and skyline views of the patellofemoral joint were taken before and at eight months and five years after operation. The radiographs of the remaining 43 knees were reviewed twice by blind and randomised assessment to measure the progression of osteoarthritis within the joints. Overcorrection of the deformity in the coronal plane was avoided in all but two knees. Only one showed evidence of progression of osteoarthritis within the patellofemoral joint, and this was only identified in one of the four assessments. Deterioration in the state of the opposite tibiofemoral compartment was not seen. Varus deformity tended to recur. Recurrent varus of 2° was observed between eight months and five years after operation. There was no correlation between the postoperative tibiofemoral angle and the extent of recurrent varus recorded at five years. Changes in alignment may be indicative of minor polyethylene wear or of subsidence of the tibial component.

The incidence of progressive osteoarthritis within the knee was very low after UKR. Patients should be carefully selected and overcorrection of the deformity be avoided.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 5 | Pages 790 - 795
1 Sep 1999
Weale AE Murray DW Newman JH Ackroyd CE

Patella infera may occur after reconstruction of the anterior cruciate ligament (ACL), high tibial osteotomy and total knee replacement (TKR). Restriction of movement of the knee and pain may result. Our aim was to compare the incidence and to assess the effects of patella infera after TKR and unicompartmental knee replacement (UKR).

We reviewed radiographs of the knees of 84 patients who had had either TKR or UKR as part of a randomised, controlled trial. The length of the patellar tendon was measured on serial radiographs taken before, at eight months and at five years after operation.

There was no significant change in the length of the patellar tendon after UKR, but a significant reduction was observed after TKR. Five years after the operation, the shortening of the tendon had increased to a mean of 3.5 mm. Of the knees with TKR reviewed at five years, 34% developed patella infera, defined as 10% or more of shortening, compared with 5% of those with UKR. Shortening was greatest in those knees which had required a lateral release; in this subgroup the mean shortening was 7.2 mm. Shortening correlated with restriction of movement and pain in the knee.

Our study has shown that patella infera develops in most patients after TKR with lateral release, and in approximately 25% of patients after TKR without this additional procedure. Patella infera rarely occurs after UKR. It is associated with restriction of movement and pain in the knee. It may be an effect of the more extensive exposure required to perform TKR and may, in part, explain the better clinical results of UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 5 | Pages 783 - 789
1 Sep 1999
Weale AE Murray DW Crawford R Psychoyios V Bonomo A Howell G O’Connor J Goodfellow JW

We determined the outcome of 56 ‘Oxford’ unicompartmental replacements performed for anteromedial osteoarthritis of the knee between 1982 and 1987. Of these, 24 were in patients who had died without revision, one was lost to follow-up and two had been revised. Of the remaining 29 knees, 26 were examined clinically and radiologically, two were only examined clinically and one patient was contacted by telephone. The mean age of the patients was 80.3 years.

At a mean follow-up of 11.4 years (10 to 14) the measurements of the knee score, range of movement and degree of deformity were not significantly different from those made one to two years after operation, except that the range of flexion had improved. Comparison of fluoroscopically-controlled radiographs at a similar interval of time showed no change in the appearance of the lateral compartments. The retained articular cartilage continued to function for ten or more years which suggests that anteromedial osteoarthritis may be considered as a focal disorder of the knee. This justifies continued efforts to develop methods of treatment which preserve intact joint structures.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 990 - 993
1 Nov 1998
Crawford R Sabokbar A Wulke A Murray DW Athanasou NA

We present a case in which the growth of an intraosseous cyst arising from the proximal tibiofibular joint appeared to have been increased by polyethylene wear particles from a medial unicompartmental knee replacement. Histological examination of the cyst wall showed a histiocytic response associated with numerous polyethylene wear particles. This case demonstrates that there is a direct communication between the joint cavity and the cyst. Such communication is probably through openings in the articular cartilage large enough to allow the passage of these particles.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 983 - 989
1 Nov 1998
Murray DW Goodfellow JW O’Connor JJ

Retrieval studies have shown that the use of fully congruent meniscal bearings reduces wear in knee replacements. We report the outcome of 143 knees with anteromedial osteoarthritis and normal anterior cruciate ligaments treated by unicompartmental arthroplasty using fully congruous mobile polyethylene bearings. At review, 34 knees were in patients who had died and 109 were in those who were still living. The mean elapsed time since operation was 7.6 years (maximum 13.8). We established the status of all but one knee.

There had been five revision operations giving a cumulative prosthetic survival rate at ten years (33 knees at risk) of 98% (95% CI 93% to 100%). Considering the knee lost to follow-up as a failure, the ‘worst-case’ survival rate was 97%. No failures were due to polyethylene wear or aseptic loosening of the tibial component. One bearing which dislocated at four years was reduced by closed manipulation.

The ten-year survival rate is the best of those reported for unicompartmental arthroplasty and not significantly different from the best rates for total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 976 - 982
1 Nov 1998
Psychoyios V Crawford RW Murray DW O’Connor JJ

Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements.

We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm per year.

In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 426 - 431
1 May 1998
Murray DW Frost SJD

The results of total knee replacement (TKR) are commonly assessed by survival analysis using revision as the endpoint. We have used the assessment of pain by a patient-based questionnaire as an alternative.

In one hospital, 1429 TKRs were inserted by 66 surgeons between 1987 and 1993. The survival at seven years, with revision as the endpoint, was 97.5% (CI 94 to 100). There were no significant differences between the three different types of implant used, the AGC, the IB2 and the Nuffield Knee. When the endpoint was the development of moderate pain, the survival at seven years for the AGC knee was 72% and that for the IB2 was similar. Significantly more patients (p = 0.007) with the Nuffield Knee, however, had developed moderate pain.

Using revision as the endpoint, it is difficult to discriminate between the various types of TKR, but this can be achieved using pain. In this investigation 30% of the patients reported moderate pain at some stage by seven years from operation.