Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 539 - 539
1 Sep 2012
Bragdon C Doerner M Callanan M Zurakowski D Kwon Y Rubash H Malchau H
Full Access

Introduction

Acetabular cup positioning has been linked to dislocation and increased bearing surface wear. A previous study found correlations between patient and surgical factors and acetabular component position. The purpose of this study was to determine if acetabular cup positioning improves when surgeons receive feedback on their performance.

Methods

Post-op anteroposterior (AP) pelvis and cross-table lateral radiographs were previously obtained for 2061 patients who received a total hip arthroplasty (THA) or hip resurfacing from 2004–2008. The surgeries were performed by 7 surgeons. AP radiographs were measured using Hip Analysis Suite to calculate the cup inclination and version angles. Acceptable ranges were defined for abduction (30–45 °) and version (5–25 °). The same surgeons performed a THA or hip resurfacing on 385 patients from January 2009 through June 2010. Cup inclination and version angles for this set of surgeries were compared to surgeries from 2004–2008 to determine if cup inclination and version angles improved in response to previous acetabular cup positioning studies. Improvement in accuracy was assessed by the chi-square test.