header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 9 - 9
1 Oct 2019
Wood T Ndoja S Howard JL McCalden RW Lanting BA Naudie DDR
Full Access

Introduction

Trunnion wear has been reported as a cause for failure of modern total hip replacement implants. While there are several reports on the prevalence of trunnion corrosion with specific stem designs, little is known about the prevalence of this problem across other femoral stem designs. The purpose of this study is to review three commonly used uncemented femoral stems and to correlate any established risk factors with rates of revision.

Methods

We reviewed all patients followed prospectively over a 15-year period (March 2000 to September 2015) who underwent total hip arthroplasty with one of three modern uncemented femoral stem designs. All of the stems were made of titanium (Ti6Al4V alloy) with a 12/14 taper design. We included only metal (cobalt chromium) on highly cross-linked polyethylene articulations. We evaluated age, gender, body mass index (BMI), femoral head size, head length, neck angle and offset as potential risk factors for all cause revision and revision excluding infection by univariate analysis. We performed a logistic regression analysis with odds ratios (OR) for each parameter for both all cause revision as well as with infection excluded. We used a p value of <0.05 as a level of significance.