header advert
Results 1 - 20 of 35
Results per page:
Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 53 - 53
1 Apr 2019
Van Onsem S Verstraete M Verrewaere D Van Der Straeten C Victor J
Full Access

Background

Under- or oversizing of either component of a total knee implant can lead to early component loosening, instability, soft tissue irritation or overstuffing of joint gaps. All of these complications may cause postoperative persistent pain or stiffness. While survival of primary TKA's is excellent, recent studies show that patient satisfaction is worse. Up to 20% of the patients are not satisfied with the outcome as and residual pain is still a frequent occurrence.

The goal of this study was therefore to evaluate if the sizing of the femoral component, as measured on a 3D-reconstructed projection, is related to patient reported outcome measures.

From our prospectively collected TKA outcome database, all patients with a preoperative CT and a postoperative X-ray of their operated knee were included in this study. Of these 43 patients, 26 (60,5%) were women and 17 (39,5%) were men. The mean age (+/−SD) was 74,6 +/− 9 years.

Methods

CT scans were acquired. All patients underwent TKA surgery in a single institution by one surgical team using the same bi- cruciate substituting total knee (Journey II BCS, Smith&Nephew, Memphis, USA). Using a recently released X-ray module in Mimics (Materialise NV, Leuven, Belgium), this module allows to align the post-operative bi-planar x-rays with the 3D- reconstructed pre-operative distal femur and to determine the 3D position of the bone and implant models using the CAD- file of the implant. This new technique was validated at our department and was found to have a sub-degree, sub-millimeter accuracy. Eleven zones of interest were defined. On the medial and the lateral condyle, the extension, mid-flexion and deep flexion facet were determined. Corresponding trochlear zones were defined and two zones were defined to evaluate the mediolateral width. In order to compare different sizes, elastic deforming mesh matching algorithms were implemented to transfer the selected surfaces from one implant to another. The orthogonal distances from the implant to the nearest bone were calculated. Positive values represent a protruding (oversized) femoral component, negative values an undersized femoral component. The figure shows the marked zones on the femoral implant. The KOOS subscores and KSS Satisfaction subscore were evaluated.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 55 - 55
1 Apr 2019
Van Onsem S Verstraete M Van Der Straeten C Victor J
Full Access

Background

Kinematic patterns in total knee arthroplasty (TKA) can vary considerably from the native knee. No study has shown a relation between a given kinematic pattern and patient satisfaction yet.

Questions

The purpose of this study was to test whether the kinematical pattern, and more specifically the anteroposterior translation during (1) open kinetic chain flexion-extension, (2) closed kinetic chain chair rising and (3) squatting, is related to the level of patient satisfaction after TKA.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 2 - 2
1 Apr 2019
Chappell K Van Der Straeten C McRobbie D Gedroyc W Brujic D Meeson R
Full Access

Introduction

Cruciate retaining knee replacements are only implanted into patients with “healthy” ligaments. However, partial anterior cruciate ligament (ACL) tears are difficult to diagnose with conventional MRI. Variations of signal intensity within the ligament are suggestive of injury but it is not possible to confirm damage or assess the collagen alignment within the ligaments. The potential use of Magic Angle Directional Imaging (MADI) as a collagen contrast mechanism is not new, but has remained a challenge. In theory, ligament tearing or joint degeneration would decrease tissue anisotropy and reduce the magic angle effect. Spontaneous cruciate ligament rupture is relatively common in dogs. This study presents results from ten canine knees.

Methods

Ethical approval was obtained to collect knees from euthanized dogs requiring a postmortem (PM). A Siemens Verio 3T MRI scanner was used to scan a sphere containing the canine knees in 9 directions to the main magnetic field (B0) with an isotropic 3D-T1-FLASH sequence. After imaging, the knees were dissected and photographed. The images were registered and aligned to compare signal intensity variations. Segmentation using a thresholding technique identified voxels containing collagen. For each collagen-rich voxel the orientation vector was computed using Szeverenyi and Bydder's method. Each orientation vector reflects the net effect of all fibers comprised within a voxel. The assembly of all unit vectors represents the fiber orientation map and was visualised in ParaView using streamlines. The Alignment Index (AI) is defined as a ratio of the fraction of orientations within 20° (solid angle) centred in that direction to the same fraction in a random (flat) case. By computing AI for a regular gridded orientation space we can visualise differences in AI on a hemisphere. AI was normalised so that AI=0 indicates isotropic collagen alignment. Increasing AI values indicate increasingly aligned structures: AI=1 indicates that all collagen fibers are orientated within the cone of 20° centred at the selected direction.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 57 - 57
1 Apr 2019
Van Onsem S Van Damme E Dedecker D Van Der Straeten C Sande I Wefula E
Full Access

Introduction

Today, Uganda has the second highest rate of road accidents in Africa and the world after Ethiopia. According to the World Health Organization's Global Status Report on Road Safety 2013, Uganda is named among countries with alarmingly high road accident rates. If such trend of traffic accidents continues to increase, the health losses from traffic injuries may be ranked as the second to HIV/AIDS by 2020. These road traffic accidents often result in terrible open injuries. Open fractures are complex injuries of bone and soft tissue. They are orthopedic emergencies due to risk of infection secondary to contamination and compromised soft tissues and sometimes vascular supply and associated healing problems. Any wound occurring on the same limb should be suspected as result of open fracture until proven otherwise. The principles of management of open fracture are initial evaluation and exclusion of life threatening injuries, prevention of infection, healing of fracture and restoration of function to injured extremity. Because of the poor hygienic circumstances and the high rate of cross-infection due to the crowded patient-wards, the risk of getting a post-operative infection is relatively high.

Osteoset-T® (Wright Medical) is a medical grade calcium sulfate bone graft substitute which is enhanced for use in infected sites by incorporating 4% tobramycin sulfate. The tobramycin is released locally, allowing therapeutic antibiotic levels at the graft site, while maintaining low systemic antibiotic levels. This local treatment of infection allows new bone formation in the defect site, while decreasing potential systemic effects.

Purpose/aim

Prevention and treatment of postoperative osteomyelitis by introducing alcoholic hand-sanitizers and the use of wound debridement and implantation of a medicated bone graft substitute.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 32 - 32
1 Apr 2018
Van Der Straeten C Abdulhussein D Brevadt MJ Cobb J
Full Access

Background

Hip resurfacing arthroplasty (HRA) and total hip arthroplasty (THA) are treatments of end-stage hip disease. Gait analysis studies comparing HRA and THA have demonstrated that HRA results in a more normal gait than THA. The reasons may include the larger, more anatomic head diameter or the preservation of the neck of the femur with restoration of the anatomical position of the hip centre and normal proprioception. This study investigated (1) whether femoral head size diameter affects gait; (2) whether gait still differs between THA and HRA patients even with comparable head diameters.

Methods

We retrospectively analysed the gait of 33 controls and 50 patients with a unilateral hip replacement, operated by the same surgeon. Follow-up ranged from 9–68 months. In 27 hips a small femoral head size was used (≤ 36mm); in 23 hips a large head size (>36mm). The small size group consisted of 11 long femoral stem THA and 16 short-stem THA and the large group of 5 long-stem THA, 8 short-stem THA and 10 HRA patients. There were 14 females/19 males in the control group; 22 females/5 males in the small size group; 13 females/10 males in the large size group.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 33 - 33
1 Apr 2018
Van Der Straeten C Cameron-Blackie A Auvinet E
Full Access

INTRODUCTION

Osteoarthritis (OA) is a growing societal burden, due to the ageing population. Less invasive, less damaging, and cheaper methods for diagnosis are needed, and sound technology is an emerging tool in this field. Some studies investigate ultrasound signals, while others look at acoustic signals in the audible range.

AIMS

The aim of the current research was to: 1) investigate the potential of visual scalogram analysis of Acoustic Emission (AE) frequencies within the human audible range (20–20000 Hz) to diagnose knee OA, 2) correlate the qualitative visual scalogram analysis of the AE with OA symptoms, and 3) to do this based on information gathered during gait.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 116 - 116
1 Mar 2017
Riviere C Lazennec J Muirhead-Allwood S Auvinet E Van Der Straeten C Cobb J
Full Access

The current, most popular recommendation for cup orientation, namely the Lewinnek box, dates back to the 70's, that is to say at the stone age of hip arthroplasty. Although Lewinnek's recommendations have been associated with a reduction of dislocation, some complications, either impingement or edge loading related, have not been eliminated. Early dislocations are becoming very rare and most of them probably occur in “outlier” patients with atypical pelvic/hip kinematics. Because singular problems usually need singular treatments, those patients need a more specific personalised planning of the treatment rather than a basic systematic application of Lewinnek recommendations. We aim in this review to define the potential impacts that the spine-hip relations (SHRs) have on hip arthroplasty. We highlight how recent improvements in hip implants technology and knowledge about SHRs can substantially modify the planning of a THR, and make the «Lewinnek recommendations» not relevant anymore. We propose a new classification of the SHRs with specific treatment recommendations for hip arthroplasty whose goal is to help at establishing a personalized planning of a THR. This new classification (figures 1 and 2) gives a rationale to optimize the short and long-term patient's outcomes by improving stability and reducing edge loading. We believe this new concept could be beneficial for clinical and research purposes.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 52 - 52
1 Mar 2017
Navruzov T Riviere C Van Der Straeten C Harris S Aframian A Iranpour F Cobb J Auvinet E
Full Access

Background

The accurate positioning of the total knee arthroplasty affects the survival of the implants(1). Alignment of the femoral component in relation to the native knee is best determined using pre- and post-operative 3D-CT reconstruction(2). Currently, the scans are visualised on separate displays. There is a high inter- and intra-observer variability in measurements of implant rotation and translation(3). Correct alignment is required to allow a direct comparison of the pre- and post-operative surfaces. This is prevented by the presence of the prostheses, the bone shape alteration around the implant, associated metal artefacts, and possibly a segmentation noise.

Aim

Create a novel method to automatically register pre- and post-operative femora for the direct comparison of the implant and the native bone.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 64 - 64
1 Mar 2017
Van Onsem S Van Der Straeten C Arnout N Deprez P Van Damme G Victor J
Full Access

Background

Total knee arthroplasty (TKA) is a proven and cost-effective treatment for osteoarthritis. Despite the good to excellent long-term results, some patients remain dissatisfied. Our study aimed at establishing a predictive model to aid patient selection and decision-making in TKA.

Methods

Using data from our prospective arthroplasty outcome database, 113 patients were included. Pre- and postoperatively, the patients completed 107 questions in 5 questionnaires: KOOS, OKS, PCS, EQ-5D and KSS. First, outcome parameters were compared between the satisfied and dissatisfied group. Secondly, we developed a new prediction tool using regression analysis. Each outcome score was analysed with simple regression. Subsequently, the predictive weight of individual questions was evaluated applying multiple linear regression. Finally, 10 questions were retained to construct a new prediction tool.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 53 - 53
1 Mar 2017
Navruzov T Van Der Straeten C Riviere C Jones G Cobb J Auvinet E
Full Access

Introduction

Hip resurfacing arthroplasty (HRA) is currently regaining positive attention as a treatment of osteoarthritis in young, active individuals[1]. The procedure is complex and has low tolerance for implant malpositioning [2]. ‘Precision tools', such as imageless navigation and patient specific instruments, have been developed to assist with implant positioning but have not been shown to be fully reliable [3]. The aim of this study is to present and validate the first step of novel quality control tool to verify implant position intra-operatively. We propose that, before reaming of the femoral head, a handheld structured light 3D scanner can be used to assess the orientation and insertion point of femoral guide wire.

Methods

Guide wires were placed into the heads of 29 solid foam synthetic femora. A specially designed marker (two orthogonal parallelepipeds attached to a shaft) was inserted into the guide wire holes. Each bone (head, neck and marker) was 3D scanned twice (fig 1). The insertion point and guide wire neck angle were calculated from the marker's parameters. Reference data was acquired with an optical tracking system. The measurements calculated with the 3D scans were compared to the reference ones to evaluate the precision. The comparison of the test retest measurements done with the new method are used to evaluate intra-rater variability.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 63 - 63
1 Mar 2017
Van Der Straeten C Banica T De Smet A Van Onsem S Sys G
Full Access

Introduction

Systemic metal ion monitoring (Co;Cr) has proven to be a useful screening tool for implant performance to detect failure at an early stage in metal-on-metal hip arthroplasty. Several clinical studies have reported elevated metal ion levels after total knee arthroplasty (TKA), with fairly high levels associated with rotating hinge knees (RHK) and megaprostheses1. In a knee simulator study, Kretzer2, demonstrated volumetric wear and corrosion of metallic surfaces. However, prospective in vivo data are scarce, resulting in a lack of knowledge of how levels evolve over time. The goal of this study was to measure serum Co and Cr levels in several types TKA patients prospectively, evaluate the evolution in time and investigate whether elevated levels could be used as an indicator for implant failure.

Patients and Methods

The study was conducted at Ghent University hospital. 130 patients undergoing knee arthroplasty were included in the study, 35 patients were lost due to logistic problems. 95 patients with 124 knee prostheses had received either a TKA (primary or revision) (69 in 55 patients), a unicompartimental knee arthroplasty (7 UKA), a RHK (revision −7 in 6 patients) or a megaprosthesis (malignant bone tumours − 28 in 27 patients) (Fig 1). The TKA, UKA and RHK groups were followed prospectively, with serum Co and Cr ions measured preoperatively, at 3,6 and 12 months postoperatively. In patients with a megaprosthesis, metal ions were measured at follow-up (cross-sectional study design).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 76 - 76
1 Feb 2017
Cobb J Wiik A Brevadt MJ Auvinet E Van Der Straeten C
Full Access

Intro

Across much of medicine, activity levels predict life expectancy, with low levels of activity being associated with increased mortality, and higher levels of activity being associated with longer healthier lives. Resurfacing is a technically demanding procedure that has suffered significant fallout from the failure of a couple of poorly performing designs. However strong evidence associates resurfacing with improved life expectancy in both the short and longer term following surgery.

We wondered if there was any relationship between the function of hips following surgery and the extent of that surgery. Could a longer stem inside the femur be the reason for a slightly reduced step length? We proposed the nul hypothesis that there was no clinically relevant difference between stem length and gait.

Method

After informed consent each subject was allowed a 5 minute acclimatisation period at 4km/hr on the instrumented treadmill (Kistler Gaitway, Amherst, NY). Their gait performance on an increasing incline at 5, 10 and 15%. At all 0.5km incremental intervals of speed, the vertical component of the ground reaction forces, center of pressure and temporal measurements were collected for both limbs with a sampling frequency of 100Hz over 10sec.

They were also asked to log onto our JointPRO website and report their function using Oxford, EQ5D, and Imperial scores.

Owing to current restrictions in indications, the patient groups selected were not comparable. However, from our database of over 800 patients who have been through the gait lab. 82 subjects were tested from 2 diagnostic groups (29 conventional THR, 27 hip resurfacing) and compared with a slightly younger group of 26 healthy controls. Patients were excluded if less than 12 months postop, or with any other documented joint disease or medical comorbidities which might affect gait performance.

Body weight scaling was also applied to the outputted mechanical data to correct for mass differences. All variables for each subject group were compared to each other using an analysis of variance (ANOVA) with Tukey post hoc test with significance set at α=0.05.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 34 - 34
1 Feb 2017
Brevadt MJ Wiik A Aqil A Auvinet E Loh C Johal H Van Der Straeten C Cobb J
Full Access

Introduction

Financial and human cost effectiveness is an increasing evident outcome measure of surgical innovation. Considering the human element, the aim is to restore the individual to their “normal” state by sparing anatomy without compromising implant performance. Gait lab studies have shown differences between different implants at top walking speed, but none to our knowledge have analysed differing total hip replacement patients through the entire range of gait speed and incline to show differences. The purpose of this gait study was to 1) determine if a new short stem femoral implant would return patients back to normal 2) compare its performance to established hip resurfacing and long stem total hip replacement (THR) implants.

Method

110 subjects were tested on an instrumented treadmill (Kistler Gaitway), 4 groups (short-stem THR, long-stem THR, hip resurfacing and healthy controls) of 28, 29, 27, and 26 respectively. The new short femoral stem patients (Furlong Evolution, JRI) were taken from the ongoing Evolution Hip trial that have been tested on the treadmill minimum 12months postop. The long stem total hip replacements and hip resurfacing groups were identified from our 800+ patient treadmill database, and only included with tests minimum 12 months postop and had no other joint disease or medical comorbidities which would affect gait performance.

All subjects were tested through their entire range of gait speeds and incline after having a 5 minute habituation period. Speed were increased 0.5kmh until maximum walking speed achieved and inclines at 4kmh for 5,10,15%. At all incremental intervals of speed 10seconds ere collected, including vertical ground reaction forces (normalized to body mass), center of pressure and temporal measurements were for both limbs (fs=100Hz). Symmetry Index(SI) were calculated on a range of features comparing leg with implanted hip to the contralateral normal hip. Group means for each feature for each subject group were compared using an analysis of variance (ANOVA) with Tukey post-hoc test with significance set at α=0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_23 | Pages 80 - 80
1 Dec 2016
Malizos K Capuano N Mezzoprete R D'Arienzo M Van Der Straeten C Drago L Romanò C
Full Access

Aim

Infection remains among the first reasons of failure of joint prosthesis. According to various preclinical reports, antibacterial coatings of implants may prevent bacterial adhesion and biofilm formation. Aim of this study is to present the first clinical trial on an antibiotic-loaded fast-resorbable hydrogel coating*, in patients undergoing hip or knee prosthesis.

Method

In this multi-center, randomized, prospective, study, a total of 380 patients, scheduled to undergo primary or revision total hip or knee joint replacement, using a cementless or a hybrid implant, were randomly assigned, in six European orthopedic centers, to receive the antibiotic-loaded DAC coating or to a control group, without coating. Pre- and post-operative assessment of clinical scores, wound healing, laboratory tests and x-ray were performed at fixed time intervals.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 103 - 103
1 May 2016
Van Onsem S Lambrecht D Verstraete M Van Der Straeten C Victor J
Full Access

Introduction

Better functional outcomes, lower pain and better stability have been reported with knee designs which restore physiological knee kinematics. Also the ability of the TKA design to properly restore the physiological femoral rollback during knee flexion, has shown to be correlated with better restoration of the flexor/extensor mechanism (appropriate flexor/extensor muscle lever arm, sufficient quadriceps force to extend the knee under load and limited patello-femoral force), which is fundamental to the function of the human knee. The purpose of the study is to compare the kinematics of three different TKA designs, by evaluating knee motion during Activities of Daily Living. The second goal is to see if there is a correlation between the TKA kinematics and the patient reported outcomes.

Methods

Ten patients who are at least 6 months after their Total Knee Replacement are included in this study. Seven satisfied and 3 dissatisfied patients are selected for this design. In this study 5 different movements are being analysed: flexion/extension; Sitting on and rising from a chair, Stair climbing, descending stairs, Flexion and extension open chain and squatting. These movements will be captured with a fluoroscope. The 2D images that are obtained, are matched with the 3D implants. (see figure 1 and 2.) This 3D image is processed with custom-made software to be able to analyse the movement (figure 3.). Tibio-femoral contactpoints of the medial and lateral condyles, tibio-femoral axial rotation, determination of the pivot-point are analysed and described. After this analysis, a correlation between the kinematics and the KOOS and KSS is investigated.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 100 - 100
1 May 2016
Van Der Straeten C De Smet K
Full Access

Background

Reasons for revision of metal-on-metal hip resurfacing arthroplasty (MoMHRA) have evolved with improving surgical experience and techniques. Early revisions were often due to fracture of the femoral neck while later revisions are associated with loosening and/or adverse local tissue reactions (ALTR) to wear debris. In some studies, revisions of MoMHRA with ALTR have been complicated by an increased risk of rerevision and poor outcome. The purpose of this study was to investigate the causes of failure and to identify factors that improve outcome following revision of a failed HRA.

Methods

From 2001 to May 2015, 180 consecutive HRA revisions were performed in 172 patients. Ninety-nine primary surgeries were done at a HRA specialist centre (99/4211, revision rate: 2.4%), 81 elsewhere. Eight different HRA designs were revised mainly in females (60%). Components’ orientation was measured from radiographs using EBRA. Ion levels were used as a diagnostic tool since 2006 (n=153). Harris-Hip-Score (HHS) was obtained prerevision and at latest follow-up. The initial experience of the first 42 cases (Initial Group) was compared to cases 43–180 (Later Group). Patients of the Later group were noted to have less soft tissue damage, had significantly bigger THA heads implanted at surgery, were educated of the increased complication risk and some wore an abduction brace for 6 weeks.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 107 - 107
1 May 2016
Van Der Straeten C De Smet K
Full Access

Background and aim

Arthroplasty registries and consecutive series indicate significantly worse results of conventional metal-on-polyethylene total hip arthroplasty (THA) in patients younger than 50 years compared to older patients, with inferior clinical outcomes and 10-year survivorship ranging between 70 and 90%. At our institution, patients under 50 needing a THA receive either a metal-on-metal hip resurfacing (MoMHRA) or a ceramic-on-ceramic (CoC)THA. In order to evaluate the outcome of these options at minimum 10 years, we conducted a retrospective review of all MoMHRA and CoCTHA with more than 10 years follow-up implanted in patients under 50.

Methods

From a single surgeon patients’ prospective database, we identified all consecutive THA performed before May 2005 in patients under 50. All patients are contacted by phone and asked to present for a clinical exam and patient reported outcome questionnaires, standard radiographs and metal ion measurements unless the hip arthroplasty has been revised. Complications and reasons for revision are noted. Kaplan-Meier survivorship is analysed for the whole cohort and sub-analysis is performed by type hip arthroplasty, gender, diagnosis and component size.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 101 - 101
1 May 2016
Van Der Straeten C De Smet K
Full Access

Background and aim

Since the market withdrawal of the ASR hip resurfacing in August 2010 because of a higher than expected revision rate as reported in the Australian Joint Replacement Registry (AOAJRR), metal-on-metal hip resurfacing arthroplasty (MoMHRA) has become a controversial procedure for hip replacement. Failures related to destructive adverse local tissue reactions to metal wear debris have further discredited MoMHRA. Longer term series from experienced resurfacing specialists however, demonstrated good outcomes with excellent 10-to-15-year survivorship in young and active men. These results have recently been confirmed for some MoMHRA designs in the AOAJRR. Besides, all hip replacement registries report significantly worse survivorship of total hip arthroplasty (THA) in patients under 50 compared to older ages. The aim of this study was to review MoMHRA survivorship from the national registries reporting on hip resurfacing and determine the risk factors for revision in the different registries.

Methods

The latest annual reports from the AOAJRR, the National Joint Registry of England and Wales (NJR), the Swedish Hip Registry (SHR), the Finnish Arthroplasty Registry, the New Zealand Joint Registry and the Arthroplasty Registry of the Emilia-Romagna Region in Italy (RIPO) were reviewed for 10-year survivorship of MoMHRA in general and specific designs in particular. Other registries did not have enough hip resurfacing data or long term data yet. The survivorship data were compared to conventional THA in comparable age groups and determinants for success/failure such as gender, age, diagnosis, implant design and size and surgical experience were reviewed.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 65 - 65
1 May 2016
Campbell P Kung M Ebramzadeh E Van Der Straeten C DeSmet K
Full Access

Bone ingrowth fixation of large diameter, beaded cobalt chromium cups is generally considered to be reliable but this is typically judged radiographically. To date, implant retrieval data of attached bone has been limited. This study evaluated correlations between the pre-revision radiographic appearance and the measured amount of bone attachment on one design of porous coated cup.

Methods

Twenty-six monoblock, CoCr Birmingham Hip Resurfacing (BHR, Smith and Nephew, TN, USA) cups with macroscopic beads and hydroxyapatite coating were studied. Seventeen were revised for acetabular malposition with the remainder revised for femoral loosening (4), pain (1), infection (1), dislocation (1) or lysis (2). Median time to revision was 35 months (10 – 70 months). Ten patients were female; the median age of all patients was 54 years. The pre-revision radiographs were visually ranked for cup-bone integration as follows: 0 = none, 1 = < 50%, 2 = 50 – 75%, 3 = 76 – 95%, 4 = > 96% integration. Rankings were made for the superior and inferior aspects, without knowledge of the appearance of bone on the retrievals. The revised cups were photographed at an angle so the dome and the cup periphery were visualized. The area of bone in four equal segments in each of the superior and inferior aspects was measured with image analysis software. A probe was used to differentiate bone from soft tissue. Only bone that covered the beads was counted. Correlation coefficients were calculated for the radiographic and image analysis data.

Results

Radiographically, most cups were assessed as having more than 50% of bone attachment and 7 cups were ranked as having almost total integration with bone. Only 2 cups were assessed radiographically as fully loose. Measured total bone attachment ranged from none to 55%. Superior and inferior percent ingrowth were highly correlated (corr=0.68, p<0.001) but there was no correlation between percent bone and x-ray rank (inferior corr=0.01, p=0.96; superior corr=0.23, p=0.26). There was no correlation between cup malpositioning as a reason for revision and x-ray integration ranking (superior p=0.34; inferior p=0.80).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 47 - 47
1 May 2016
Bonnin M De Kok A Verstraete M Van Hoof T Van Der Straeten C Victor J
Full Access

Goals of the study

(1) to investigate the relationships between the bony contours of the knee and the Popliteus Tendon (PT) in the healthy knee and after implantation of a TKA and (2) to analyze the influence of implant sizing.

Hypothesis

With an apparently well-sized TKA, the position of the PT during knee flexion is modified compared with the preoperative situation.