Vol. 93-B, Issue SUPP_IV | Pages 549 - 549
1 Nov 2011
Joyce T Langton D Lord J Grigg H Jameson S Cooke N Tulloch C Logishetty R Meek D Nargol A
Full Access

Background: There is a paucity of published data with regard to the wear of failed metal on metal (MoM) resurfacing devices.

Materials and Methods: MoM components retrieved from patients from two independent centres experiencing failure secondary to ARMD were analysed using a Mitutoyo Legex 322 coordinate measuring machine (CMM) which has an accuracy of 0.8 microns. Between 4000 – 6000 points were taken on each explant, dependent on the size of the bearing surface. Maximum wear depths and total volumetric wear were calculated. These values were compared to those from control samples retrieved following uncomplicated fractures/femoral collapse secondary to avascular necrosis (after calculating equivalent yearly wear rates).

Results: 58 ARMD components were analysed. This included 22 36mm MoM THRs (DePuy Pinnacle), 28 DePuy ASRs and 8 Zimmer Duroms. There were 30 resurfacing fracture/avascular necrosis controls. Volumetric wear rates and maximum wear depths of ARMD resurfacing components were significantly greater than the resurfacing control group for both the ASR and Duroms (p< 0.05) however 2 ARMD components exhibited similar amounts of wear compared to controls. Wear rates of the ARMD THR group were significantly lower than the ARMD resurfacing group (p< 0.05).

Conclusions: Increased articular wear is associated with an increased incidence of local adverse effects including tissue necrosis, joint effusions and fractures. However, there are a minority of patients who can develop tissue necrosis in the absence of accelerated wear, implying a spectrum of sensitivity This is reflected in the incidence of ARMD in the patient groups at the main study centre: > 5% in the ASR group and approx 1% in the THR group. We believe this indicates a failure of adequate lubrication and the resultant negative effects in larger bearing devices.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 95 - 95
1 May 2011
Alshryda S Sharda P Shetty A Vaghela M Logishetty R Tulloch C Antoni N Mason J
Full Access

Introduction: Today’s aging population has resulted in an increase in the number of major orthopaedic surgical interventions in the elderly. Total knee replacement (TKR) is one of the commonest operations in orthopaedic practice. The fourth annual report of the National Joint Registry showed that there were 60 986 TKR performed in England and Wales in 2006. The true figure is probably much higher. Literature showed that 20–70% of patients who had TKR needed 1–3 units of blood.

Although safer than ever, allogeneic transfusion is still associated with risks for the recipient (haemolysis, infection, immunosuppression, transfusion-related acute lung injury and even death).

Tranexamic acid (TA) is a synthetic antifibrinolytic agent that has been successfully used to stop bleeding after dental operation, removal of tonsils, prostate surgery, heavy menstrual bleeding, eye injuries and in patients with Haemophilia.

In this study Tranexamic acid was applied topically to the exposed tissue around the knee joint prior to the wound closure and tourniquet release. It is anticipated that this method of administration is quick, easy, associated with less systemic side effect. Also, it provides a higher concentration of the Tranexamic acid at the bleeding site.

Objectives: To find out whether Tranexamic acid can reduce blood loss and subsequent blood transfusion significantly after total knee replacement when applied topically without extra side effects.

Design: A double blind randomised controlled trial of 150 patients who underwent unilateral primary cemented total knee replacement. This number gives a 90% power to detect a 50% reduction in blood loss and 80% power to detect a reduction in blood transfusion from current local standard 30% to 10%.

Outcome Measures: Blood loss, transfusion, Length of stay, complications, Euroqol and Oxford Knee Score.

Results: The two groups were comparable in age, weight, height, BMI, Tourniquet time, and type of anaesthesia. There has been significant differences in the amount of blood loss and blood transfusion in favour of tranexamic acid (p-values are 0.001 and 0.007 respectively). Fourteen patients needed blood transfusion ranged from 2–6 units. Thirteen were in the Placebo group and only one in the Tranexamic acid. There has been no significant difference among other outcomes in particular complications rates such as DVT and pulmonary embolism.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 390 - 391
1 Jul 2010
Langton D Jameson S Joyce T Natu S Logishetty R Tulloch C Nargol A
Full Access

In our independent centre, in the period from January 2003 to august 2008, over 1100 36mm MoM THRs have been implanted as well as 155 Birmingham Hip Resurfacing procedures, 402 ASR resurfacings and 75 THRs using ASR XL heads on SROM stems.

During this period we have experienced a number of failures with patients complaining of worsening groin pain at varying lengths of time post operatively. Aspiration of the hip joints yielded a large sterile effusion on each occasion. At revision, there were copious amounts of green grey fluid with varying degrees of necrosis. There were 11 failures of this nature in patients with ASR implants (10 females) and 2 in the 36 MoM THR group (one male one female).

Tissue specimens from revision surgery showed varying degrees of ‘ALVAL’ as well as consistently high numbers of histiocytes. Metal debris was also a common finding.

A fuller examination of our ASR cohort as a whole has shown that smaller components placed with inclinations > 45° and anteversions < 10 or > 20° are associated with increased metal ion levels. The 11 ASR failed joints were all sub optimally positioned (by the above definition), small components.

Explant analysis using a coordinate measuring machine and out of roundness device confirmed greater than expected wear of each component. The lower number of failures in the 36mm MoM group, as well as the equal sex incidence, suggests that the majority of these failures are due to the instigation of an immune reaction by large amounts of wear debris rather than adverse reactions to well functioning joints. It is likely that small malpositioned ASRs function in mixed to boundary lubrication, and this, combined with the larger radius of these joints compared to the 36mm MoM joints, results in more rapid wear.