Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 76 - 76
1 Apr 2018
Cristofolini L Morellato K Soffiatti R Rotini R Guerra E
Full Access

Introduction

The standard treatment of proximal humerus fractures includes pre-contoured metal plates and up to nine cortical and trabecular screws. Frequent failures are reported, especially in case of poor bone quality. The scope of this study was to assess the strength of an innovative reconstruction technique (Cement-and-screws) based on a commercial plate, with a reduced number of screws compared to the standard, and with the injection of a beta-TCP additivated acrylic bone cement (Cal-Cemex, Tecres, Italy). The focus was on a four-fragment fracture of the proximal humerus, in combination with a bone defect. For comparison, also a standard technique, based on a commercial system of plate and screws was tested (Screws-only).

Materials and Methods

Six pairs of cadaveric humeri were obtained through an ethically-approved donation program. The humeri were osteotomized to simulate a reproducible four-fragment fracture with the aid of a dedicated jig. Preparation included the simulation of a bone defect in the humeral head. One humerus of each pair was randomly assigned to one of two reconstruction techniques: (i) cement-and-screws humeri were repaired with a commercial fixation plate, 2 cortical and 3 trabecular screws (Philos, DePuy Synthes), and with injection of an acrylic cement additivated with beta-TCP (Cal-Cemex, Tecres); (ii) for comparison, screws-only humeri were prepared with the same commercial plate, 2 cortical and 6 trabecular screws. The reconstructed humeri underwent a biomechanical test. An axial force was cyclically applied, where the load magnitude started at 140 N and increased by 1% at each cycle. Failure was defined as fragment motion exceeding 8 mm.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 107 - 107
1 Jan 2017
Danesi V Tozzi G Soffiatti R Cristofolini L
Full Access

Prophylactic augmentation is meant to reinforce the vertebral body (VB), but in some cases it is suspected to actually weaken it. To elucidate the biomechanical efficacy of prophylactic augmentation, the full-field three-dimensional strain distributions were measured for the first time inside prophylactic-augmented vertebrae.

Twelve thoracic porcine vertebrae were assigned to three groups: 4 were augmented with bone cement for vertebroplasty (Mendec-Spine, Tecres), 4 were treated with another bone cement for vertebroplasty (Calcemex-Spine, Tecres) while the other 4 were tested untreated as a control. Destructive tests were carried out under axial compression, in a step-wise fashion (unloaded, 5%, 10% and 15% compression). At each loading step, μCT-images were acquired. The internal strain distribution was investigated by means of DVC analysis.

Some augmented specimens were stronger than the respective control, while others were weaker. In most of the specimens, the strain distribution in the elastic regime (5% compression) seemed to predict the location of the micro-damage initiation before it actually became identifiable (at 10% and 15% compression). The measured strain had the same order of magnitude for all groups. However, in the control vertebrae, the highest strain would unpredictably appear at any location inside the VB. Conversely, for both augmentation groups, the highest strains were measured in the regions adjacent to the injected cement mass, whereas the cement-interdigitated-bone was less strained. Localization of high strains and failure was consistent between specimens, but different between the two cement types: with Mendec-Spine failure the highest strains were mainly localized at mid-height and at the same level where the cement mass was localized; with Calcemex-Spine failure the highest strains were mainly cranial and caudal to the cement mass.

Both the micro-CT images, and the DVC strain analysis highlighted that:

The cement mass was less strained than any other regions in the vertebra. Failure never started inside the cement mass. This can be explained with the additional stiffening and reinforcement associated with the infiltration of the cement inside the trabecular bone.

The highest strains and failure were localized in the bone adjacent to the cement-bone interdigitated region. This can be explained by the strain concentration between the cement-interdigitated bone (stiffer and stronger), and the adjacent non-augmented trabecular bone

The strain maps in the elastic regime and the localization of failure was different in the augmented vertebrae, when compared to the natural controls. This suggests an alteration of the load sharing in the augmented structure where the load is mostly carried by the cement region.

The different localization of failure initiation between the two augmented groups could be explained by the different mechanical properties of the two cements.

This study has demonstrated the potential of DVC in measuring the internal strain and failure in prophylactic-augmented vertebrae. It has been shown that failure starts inside the augmented VB, next to the injected cement mass. This can help establishing better criteria (in terms of localization of the cement mass) in order to improve clinical protocols for vertebroplasty surgical procedures.