The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1192 - 1198
1 Oct 2019
Sköldenberg OG Rysinska AD Chammout G Salemyr M Mukka SS Bodén H Eisler T

Aims

Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results.

Patients and Methods

We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (sd 5), in a double-blinded, noninferiority, randomized controlled trial. The subjects were randomized in a 1:1 ratio to receive a reverse hybrid THA with a cemented component of either argon-gas gamma-sterilized PE component (controls) or VEPE, with identical geometry. The primary endpoint was proximal implant migration of the component at two years postoperatively measured with RSA. Secondary endpoints included total migration of the component, penetration of the femoral head into the component, and patient-reported outcome measurements.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 36 - 36
1 Aug 2018
Muratoglu O Galea V Connelly J Shareghi B Kärrholm J Sköldenberg O Salemyr M Laursen M Bragdon C Malchau H
Full Access

The primary purpose of the current study was to evaluate and compare the wear properties of vitamin E-doped, highly-crosslinked PE (VEPE) and one formulation of moderately cross-linked and mechanically-annealed ultra-high molecular weight PE (ModXLPE) in patients five years after primary THA. We also sought to understand whether polyethylene wear is associated with radiographic evidence of bone resorption or with deterioration in patient-reported outcome measures (PROMs).

A total of 221 patients from four international centers were recruited into a prospective RSA and clinical outcomes study. Seventy percent (76%) of patients received VEPE (vs. ModXLPE) liners, and 36% received ceramic (vs. metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA radiographs were collected to measure PE wear.

We observed similar bedding in through the one-year interval and wear through the two-year interval between the two liner types. However, there was significantly more femoral head penetration in the ModXLPE cohort compared to the VEPE cohort at the five-year follow-up (p<0.001). The only variables independently predictive of increased wear were ModXLPE (vs VEPE) liner type (β=0.22, p=0.010) and metal (vs. ceramic) femoral head type (β=0.21, p=0.013). There was no association between increased wear and radiolucency development (p=0.866) or PROMs. No patients were found to have evidence of osteolysis.

At five-years postoperatively, patients treated with VEPE (vs. ModXLPE) and ceramic (vs. metal) femoral heads demonstrated decreased wear. At the longest follow-up (five years postoperatively), the wear rates for both liner groups were very low and have not led to any osteolysis or implant failures via aseptic loosening.