Johnston WD Razii N Banger MS Rowe PJ Jones BG MacLean AD Blyth MJG
Full Access

The objective of this study was to compare differences in alignment following robotic arm-assisted bi-unicompartmental knee arthroplasty (Bi-UKA) and conventional total knee arthroplasty (TKA).

This was a prospective, randomised controlled trial of 70 patients. 39 TKAs were implanted manually, as per standard protocol at our institution, and 31 Bi-UKA patients simultaneously received fixed-bearing medial and lateral UKAs, implanted using robotic arm-assistance. Preoperative and 3-month postoperative CT scans were analysed to determine hip knee ankle angle (HKAA), medial distal femoral angle (MDFA), and medial proximal tibial angle (MPTA). Analysis was repeated for 10 patients by a second rater to validate measurement reliability by calculating the intra-class correlation coefficient (ICC).

Mean change in HKAA towards neutral was 2.7° in TKA patients and 2.3° in Bi-UKA patients (P=0.6). Mean change in MDFA was 2.5° for TKA and 1.0° for Bi-UKA (P<0.01). Mean change in MPTA was 3.7° for TKA and 0.8° for Bi-UKA (P<0.01). Mean postoperative MDFA and MPTA for TKAs were 89.8° and 89.6° respectively, indicating orientation of femoral and tibial components perpendicular to the mechanical axis. Mean postoperative MDFA and MPTA for Bi-UKAs were 91.0° and 86.9° respectively, indicating a more oblique joint line orientation. Inter-rater agreement was excellent (ICC>0.99). Early functional activities, according to the new Knee Society Scoring System, favoured Bi-UKAs (P<0.05).

Robotic arm-assisted, cruciate-sparing Bi-UKA better maintains the natural anatomy of the knee in the coronal plane and may therefore preserve normal joint kinematics, compared to a mechanically aligned TKA. This has been achieved without significantly altering overall HKAA.